Absolute Quantitation of Intact Recombinant Antibody Product Variants Using Mass Spectrometry

被引:18
|
作者
Macchi, Frank D. [1 ]
Yang, Feng [1 ]
Li, Charlene [1 ]
Wang, Chenchen [2 ]
Anh Nguyen Dang [1 ]
Marhoul, Joseph C. [1 ]
Zhang, Hui-min [1 ]
Tully, Timothy [3 ]
Liu, Hongbin [1 ]
Yu, X. Christopher [1 ]
Michels, David A. [1 ]
机构
[1] Genentech Inc, Prot Analyt Chem, San Francisco, CA 94080 USA
[2] Eurofins Lancaster Labs Inc, Lancaster, PA 17601 USA
[3] Genentech Inc, Purificat Dev, San Francisco, CA 94080 USA
关键词
STANDARD ADDITION METHOD; MONOCLONAL-ANTIBODY; HYDROPHOBIC INTERACTION; BISPECIFIC ANTIBODIES; LC-MS/MS; QUANTIFICATION; PROTEINS; CHROMATOGRAPHY; MS; CALIBRATION;
D O I
10.1021/acs.analchem.5b02627
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Accurate and precise quantitative measurement of product-related variants of a therapeutic antibody is essential for product development and testing. Bispecific antibodies (bsAbs) are Abs composed of two different half antibody arms, each of which recognizes a distinct target, and recently they have attracted substantial therapeutic interest. Because of the increased complexity of its structure and its production process, as compared to a conventional monoclonal antibody, additional product-related variants, including covalent and noncovalent homodimers of half antibodies (hAbs), may be present in the bsAb product. Sufficient separation and reliable quantitation of these bsAb homodimers using liquid chromatography (LC) or capillary electrophoresis-based methods is challenging because these homodimer species and the bsAb often have similar physicochemical properties. Formation of noncovalent homodimers and heterodimers can also occur. In addition, since homodimers share common sequences with their corresponding halves and bsAb, it is not suitable to use peptides as surrogates for their quantitation. To tackle these analytical challenges, we developed a mass spectrometry-based quantitation method. Chip-based nanoflow LC time-of-flight mass spectrometry coupled with a standard addition approach provided unbiased absolute quantitation of these drug-product-related variants. Two methods for the addition of known levels of standard (multi- or single-addition) were evaluated. Both methods demonstrated accurate and reproducible quantitation of homodimers at the 0.2% (w/w) level, with the single-addition method having the promise of higher analytical throughput.
引用
收藏
页码:10475 / 10482
页数:8
相关论文
共 50 条
  • [41] Two-dimensional capillary zone electrophoresis–mass spectrometry for the characterization of intact monoclonal antibody charge variants, including deamidation products
    Kevin Jooß
    Jens Hühner
    Steffen Kiessig
    Bernd Moritz
    Christian Neusüß
    Analytical and Bioanalytical Chemistry, 2017, 409 : 6057 - 6067
  • [42] Quantitation of DNA adducts using high resolution mass spectrometry
    Villalta, Peter W.
    Balbo, Silvia
    Hecht, Stephen S.
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2011, 242
  • [43] Quantitation of endogenous peptides using mass spectrometry based methods
    Romanova, Elena V.
    Dowd, Sarah E.
    Sweedler, Jonathan V.
    CURRENT OPINION IN CHEMICAL BIOLOGY, 2013, 17 (05) : 801 - 808
  • [44] Selectivity for quantitation of biomarkers using liquid chromatography and mass spectrometry
    Weng, Naidong
    Jian, Wenying
    BIOANALYSIS, 2018, 10 (18) : 1461 - 1465
  • [45] Quantitation of quaternary ammonium compounds using electrospray mass spectrometry
    Hind, AR
    Bhargava, SK
    Cullis, PG
    ANALYTICA CHIMICA ACTA, 1998, 377 (01) : 39 - 45
  • [46] Quantitation of the Ribosomal Protein Autoregulatory Network Using Mass Spectrometry
    Sykes, Michael T.
    Sperling, Edit
    Chen, Stephen S.
    Williamson, James R.
    ANALYTICAL CHEMISTRY, 2010, 82 (12) : 5038 - 5045
  • [47] Protein Quantitation Using Isotope-Assisted Mass Spectrometry
    Kline, Kelli G.
    Sussman, Michael R.
    ANNUAL REVIEW OF BIOPHYSICS, VOL 39, 2010, 39 : 291 - 308
  • [48] Identification of hemoglobin variants by top-down mass spectrometry using selected diagnostic product ions
    Graca, Didia Coelho
    Hartmer, Ralf
    Jabs, Wolfgang
    Beris, Photis
    Clerici, Lorella
    Stoermer, Carsten
    Samii, Kaveh
    Hochstrasser, Denis
    Tsybin, Yury O.
    Scherl, Alexander
    Lescuyer, Pierre
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2015, 407 (10) : 2837 - 2845
  • [49] Identification of hemoglobin variants by top-down mass spectrometry using selected diagnostic product ions
    Didia Coelho Graça
    Ralf Hartmer
    Wolfgang Jabs
    Photis Beris
    Lorella Clerici
    Carsten Stoermer
    Kaveh Samii
    Denis Hochstrasser
    Yury O. Tsybin
    Alexander Scherl
    Pierre Lescuyer
    Analytical and Bioanalytical Chemistry, 2015, 407 : 2837 - 2845
  • [50] Absolute Quantitation of Tryptophan-Containing Peptides and Amyloid β-Peptide Fragments by Coulometric Mass Spectrometry