Determination of norepinephrine in small volume plasma samples by stable-isotope dilution gas chromatography-tandem mass spectrometry with negative ion chemical ionization

被引:39
|
作者
Kuhlenbeck, DL [1 ]
O'Neill, TP [1 ]
Mack, CE [1 ]
Hoke, SH [1 ]
Wehmeyer, KR [1 ]
机构
[1] Procter & Gamble Co, Pharmaceut, Hlth Care Res Ctr, Mason, OH 45040 USA
来源
JOURNAL OF CHROMATOGRAPHY B | 2000年 / 738卷 / 02期
关键词
norepinephrine;
D O I
10.1016/S0378-4347(99)00541-1
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A stable-isotope based gas chromatography-tandem mass spectrometry-negative ion chemical ionization method was developed for the determination of norepinephrine (NE) levels in small volumes (25-100 mu l) of plasma. NE was stabilized in plasma by the addition of semicarbazide and spiked with deuterium-labeled norepinephrine internal standard. The analytes were isolated from the plasma by solid-phase extraction using phenylborofiic acid columns and derivatized using pentafluoropropionic anhydride. The derivatized analytes were chromatographed on a capillary column and detected by tandem mass spectrometry with negative ion chemical ionization. Unparalleled sensitivity and selectivity were obtained using this detection scheme, allowing the unambiguous analysis of trace levels of NE in small-volume plasma samples. Linear standard curves were obtained for NE over a mass range from 1 to 200 pg per sample. The method had a limit of quantitation of 10 pg NE/ml plasma when using a 100-mu 1*1 sample aliquot (1 pg/sample). Accuracy for the analysis of plasma samples spiked with 10 to 200 pg NE/ml typically ranged from 100+/-10%, with RSD values of less than 108. The methodology was applied to determine the effect of clonidine on plasma NE levels in conscious spontaneously hypertensive rats. Administration of clonidine (30 mu g/kg) produced an similar to 80% reduction in plasma NE accompanied by a 30% reduction in heart and mean arterial pressure that persisted >90 min after drug administration. The ability to take multiple samples from individual rats allowed the time course for the effect of clonidine to be mapped out using only one group of animals. O 7000 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:319 / 330
页数:12
相关论文
共 50 条
  • [31] Simultaneous determination of estrogenic and androgenic hormones in water by isotope dilution gas chromatography-tandem mass spectrometry
    Trang Trinh
    Harden, Nick B.
    Coleman, Heather M.
    Khan, Stuart J.
    JOURNAL OF CHROMATOGRAPHY A, 2011, 1218 (12) : 1668 - 1676
  • [32] Determination of 17β-estradiol in bovine plasma:: development of a highly sensitive technique by ion trap gas chromatography-tandem mass spectrometry using negative chemical ionization
    Biancotto, G
    Angeletti, R
    Traldi, P
    Silvestri, M
    Saccon, M
    Guidugli, F
    JOURNAL OF MASS SPECTROMETRY, 2002, 37 (12): : 1266 - 1271
  • [33] Quantitative determination of clonazepam in plasma by gas chromatography-negative ion chemical ionization mass spectrometry
    Song, DM
    Zhang, S
    Kohlhof, K
    JOURNAL OF CHROMATOGRAPHY B-BIOMEDICAL APPLICATIONS, 1996, 686 (02): : 199 - 204
  • [34] Stable-Isotope Dilution HPLC-Electrospray Ionization Tandem Mass Spectrometry Method for Quantifying Hydroxyurea in Dried Blood Samples
    Marahatta, Anu
    Megaraj, Vandana
    McGann, Patrick T.
    Ware, Russell E.
    Setchell, Kenneth D. R.
    CLINICAL CHEMISTRY, 2016, 62 (12) : 1593 - 1601
  • [35] Quantification of key folate forms in serum using stable-isotope dilution ultra performance liquid chromatography-tandem mass spectrometry
    Kirsch, Susanne H.
    Knapp, Jean-Pierre
    Herrmann, Wolfgang
    Obeid, Rima
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2010, 878 (01): : 68 - 75
  • [36] Methylmalonic acid measured in plasma and urine by stable-isotope dilution and electrospray tandem mass spectrometry
    Magera, MJ
    Helgeson, JK
    Matern, D
    Rinaldo, P
    CLINICAL CHEMISTRY, 2000, 46 (11) : 1804 - 1810
  • [37] Determination of cyclophosphamide and ifosfamide in sewage effluent by stable isotope-dilution liquid chromatography-tandem mass spectrometry
    Llewellyn, N.
    Lloyd, P.
    Juergens, M. D.
    Johnson, A. C.
    JOURNAL OF CHROMATOGRAPHY A, 2011, 1218 (47) : 8519 - 8528
  • [38] Determination of Pentachlorophenol in Milk Powder by Isotope Dilution Liquid Chromatography-Tandem Mass Spectrometry
    Qi Y.
    Li J.
    Li C.
    Xu Q.
    Shipin Kexue/Food Science, 2023, 44 (16): : 324 - 330
  • [39] DETERMINATION OF REDUCED AND OXIDIZED GLUTATHIONE BY ISOTOPE DILUTION LIQUID CHROMATOGRAPHY-TANDEM MASS SPECTROMETRY
    Yu, W. T.
    Kao, C. Y.
    Shiesh, S. C.
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2011, 49 : S783 - S783
  • [40] Quantification of acetylcholine, choline, betaine, and dimethylglycine in human plasma and urine using stable-isotope dilution ultra performance liquid chromatography-tandem mass spectrometry
    Kirsch, Susanne H.
    Herrmann, Wolfgang
    Rabagny, Yannick
    Obeid, Rima
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2010, 878 (32): : 3338 - 3344