Batch-type detection cell using a peroxyoxalate chemiluminescence system for capillary electrophoresis

被引:19
|
作者
Tsukagoshi, K [1 ]
Otsuka, M [1 ]
Hashimoto, M [1 ]
Nakajima, R [1 ]
Kondo, K [1 ]
机构
[1] Doshisha Univ, Fac Engn, Dept Chem Engn & Mat Sci, Kyoto 6100321, Japan
关键词
capillary electrophoresis; chemiluminescence; peroxyoxalate; optical fiber; batch-type; alpha-amino acid;
D O I
10.2116/analsci.15.1257
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A batch-type chemiluminescence detection cell using a peroxyoxalate system was easily combined with capillary electrophoresis equipment; the device was operated without any tedious procedures. Peroxyoxalate reagents and hydrogen peroxide were examined in detail concerning their usage and concentration with the cell. Fluorescence-labeled alpha-amino acids showed the detection limits of 10 - 100 nM order (S/N=3), which were approximately 2 or 3 orders of magnitude lower than those obtained with the absorbance. A mixture sample of the labeled alpha-amino acids was successfully separated and detected.
引用
收藏
页码:1257 / 1260
页数:4
相关论文
共 50 条