Validation of a smartphone-based Rinne test to detect an air-bone gap

被引:5
|
作者
Hibscher, Daniel [1 ]
Oron, Yahav [1 ]
Handzel, Ophir [1 ]
Warshavsky, Anton [1 ]
Horowitz, Gilad [1 ]
Ungar, Omer J. [1 ]
机构
[1] Tel Aviv Univ, Tel Aviv Sourasky Med Ctr, Sackler Sch Med, Dept Otolaryngol Head Neck & Maxillofacial Surg, 6 Weizmann St, IL-6423906 Tel Aviv, Israel
关键词
Hearing loss; Sudden sensorineural hearing loss; Rinne test; Conductive hearing loss; Tuning fork;
D O I
10.1007/s00405-021-06608-9
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
Objective To validate a smartphone-based Rinne test employing the vibration application of mobile telephones. Study design Prospective controlled clinical study. Setting Tertiary referral medical center. Methods Twenty consecutive patients hospitalized in the otolaryngology department of a tertiary medical center due to unilateral hearing loss (HL study group), and 30 consecutive inpatients on the same ward who had no otological history (controls) were enrolled. Each participant underwent the traditional 512 Hz tuning fork-based Rinne test, as well as a smartphone-based Rinne test by means of a single uncovered smartphone with a vibration application. The test results were compared to those of formal audiometry. Results The overall agreement between the traditional Rinne test and the smartphone-based test was 98%. The Sensitivity was 85% for both tests, specificity was 90% and 93% for smartphone and tuning fork tests, respectively. The smartphone-based Rinne test could correctly discriminate between patients with an air-bone gap >= 25 dB at 512 Hz from patients with a lower or no air-bone gap at 512 Hz. The smartphone-based Rinne could not evaluate two patients with a moderately severe/severe sensorineural hearing loss due to their inability to detect the vibrations. Conclusion A smartphone-based Rinne test was validated for the detection of an air-bone gap >= 25 dB at 512 Hz in the clinical setting. The validity of patient-operated smartphone-based Rinne test awaits further study. Level of evidence 2B.
引用
收藏
页码:4767 / 4773
页数:7
相关论文
共 50 条
  • [21] The Finger Dexterity Test: Validation study of a smartphone-based manual dexterity assessment
    Van Laethem, Delphine
    Denissen, Stijn
    Costers, Lars
    Descamps, Annabel
    Baijot, Johan
    Van Remoortel, Ann
    Van Merhaegen-Wieleman, Annick
    D'hooghe, Marie B.
    D'Haeseleer, Miguel
    Smeets, Dirk
    Sima, Diana M.
    Van Schependom, Jeroen
    Nagels, Guy
    MULTIPLE SCLEROSIS JOURNAL, 2024, 30 (01) : 121 - 130
  • [22] Validation of EncephalApp, Smartphone-Based Stroop Test, for the Diagnosis of Covert Hepatic Encephalopathy
    Bajaj, Jasmohan S.
    Heuman, Douglas M.
    Sterling, Richard K.
    Sanyal, Arun J.
    Siddiqui, Muhammad
    Matherly, Scott
    Luketic, Velimir
    Stravitz, R. Todd
    Fuchs, Michael
    Thacker, Leroy R.
    Gilles, HoChong
    White, Melanie B.
    Unser, Ariel
    Hovermale, James
    Gavis, Edith
    Noble, Nicole A.
    Wade, James B.
    CLINICAL GASTROENTEROLOGY AND HEPATOLOGY, 2015, 13 (10) : 1828 - U335
  • [23] Study of the factors related to air-bone gap in enlarged vestibular aqueduct
    Wang, Rong
    Zhuang, Bo-xiang
    Guo, Weiwei
    Li, Jianan
    Lin, Chang
    Yang, Shiming
    ACTA OTO-LARYNGOLOGICA, 2024, 144 (01) : 39 - 43
  • [24] AIR-BONE GAP IN MENIERES-DISEASE AFTER EXPOSURE TO OVERPRESSURE
    DENSERT, B
    ARLINGER, S
    DENSERT, O
    AUDIOLOGY, 1987, 26 (06): : 339 - 347
  • [25] Design and Validation of a Novel Smartphone-Based Visual Acuity Test: The K-VA Test
    Karampatakis, Vasileios
    Almaliotis, Diamantis
    Talimtzi, Persefoni
    Almpanidou, Stavroula
    OPHTHALMOLOGY AND THERAPY, 2023, 12 (03) : 1657 - 1670
  • [26] Design and Validation of a Novel Smartphone-Based Visual Acuity Test: The K-VA Test
    Vasileios Karampatakis
    Diamantis Almaliotis
    Persefoni Talimtzi
    Stavroula Almpanidou
    Ophthalmology and Therapy, 2023, 12 : 1657 - 1670
  • [27] Correlation of air-bone gap and size of Enlarged Vestibular Aqueduct in children
    Van Beck, Jenna
    Chinnadurai, Sivakumar
    Morrison, Alison Kemph
    Zuniga, M. Geraldine
    Smith, Bridget
    Lohse, Christine M.
    McCaslin, Devin
    INTERNATIONAL JOURNAL OF PEDIATRIC OTORHINOLARYNGOLOGY, 2020, 132
  • [28] VERIFICATION OF THE AIR-BONE GAP USING CANCELLATION AND EVOKED-RESPONSES
    BOEZEMAN, EHJF
    KAPTEYN, TS
    FEENSTRA, L
    SNEL, AM
    AUDIOLOGY, 1985, 24 (03): : 174 - 185
  • [29] Prognostic factors influencing postoperative air-bone gap in stapes surgery
    Albera, Andrea
    Parandero, Fiorella
    Andriani, Roberto
    Albera, Roberto
    Riva, Giuseppe
    Canale, Andrea
    ACTA OTORHINOLARYNGOLOGICA ITALICA, 2022, 42 (04) : 380 - 387
  • [30] Air-bone gap and resonant frequency in large vestibular aqueduct syndrome
    Nakashima, T
    Ueda, H
    Furuhashi, A
    Sato, E
    Asahi, K
    Naganawa, S
    Beppu, R
    AMERICAN JOURNAL OF OTOLOGY, 2000, 21 (05): : 671 - 674