Noninvasive pulse contour analysis for determination of cardiac output in patients with chronic heart failure

被引:15
|
作者
Roth, Sebastian [1 ]
Fox, Henrik [1 ]
Fuchs, Uwe [2 ]
Schulz, Uwe [2 ]
Costard-Jaeckle, Angelika [2 ]
Gummert, Jan F. [2 ]
Horstkotte, Dieter [1 ]
Oldenburg, Olaf [1 ]
Bitter, Thomas [1 ]
机构
[1] Ruhr Univ Bochum, Clin Cardiol, Herz & Diabet Zentrum NRW, Georgstr 11, D-32545 Bad Oeynhausen, Germany
[2] Ruhr Univ Bochum, Clin Thorac & Cardiovasc Surg, Herz & Diabet Zentrum NRW, Georgstr 11, D-32545 Bad Oeynhausen, Germany
关键词
Heart failure; Cardiac output; Pulmonary artery catheter; Noninvasive pulse contour analysis; CNAP monitor; PULMONARY-ARTERY CATHETER; PRESSURE; METAANALYSIS; PRECISION; AGREEMENT; THERMODILUTION; ACCURACY; GERMANY; SYSTEM; DEVICE;
D O I
10.1007/s00392-017-1198-7
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Determination of cardiac output (CO) is essential in diagnosis and management of heart failure (HF). The gold standard to obtain CO is invasive assessment via thermodilution (TD). Noninvasive pulse contour analysis (NPCA) is supposed as a new method of CO determination. However, a validation of this method in HF is pending and performed in the present study. Methods Patients with chronic-stable HF and reduced left ventricular ejection fraction (LVEF = 45%; HF-REF) underwent right heart catheterization including TD. NPCA using the CNAP Monitor (V5.2.14, CNSystems Medizintechnik AG) was performed simultaneously. Three standardized TD measurements were compared with simultaneous auto-calibrated NPCA CO measurements. Results In total, 84 consecutive HF-REF patients were enrolled prospectively in this study. In 4 patients (5%), TD was not successful and for 22 patients (26%, 18 with left ventricular assist device), no NPCA signal could be obtained. For the remaining 58 patients, Bland-Altman analysis revealed a mean bias of + 1.92 L/min (limits of agreement +/- 2.28 L/min, percentage error 47.4%) for CO. With decreasing cardiac index, as determined by the gold standard of TD, there was an increasing gap between CO values obtained by TD and NPCA (r = -0.75, p < 0.001), resulting in a systematic overestimation of CO in more severe HF. TD-CI classified 52 (90%) patients to have a reduced CI (< 2.5 L/min/m(2)), while NPCA documented a reduced CI in 18 patients (31%) only. Conclusions In HF-REF patients, auto-calibrated NPCA systematically overestimates CO with decrease in cardiac function. Therefore, to date, NPCA cannot be recommended in this cohort.
引用
收藏
页码:395 / 404
页数:10
相关论文
共 50 条
  • [41] Continuous cardiac output measurement by arterial pulse contour analysis: Accuracy under changes in cardiac output?
    Reuter, DA
    Ehrmann, Z
    Felbinger, TW
    Kilger, E
    Goedje, O
    Meier-Hellmann, AD
    Pfeiffer, UJ
    Goetz, AE
    CRITICAL CARE MEDICINE, 2000, 28 (12) : A71 - A71
  • [42] The accuracy of cardiac output monitoring by pulse contour analysis in intensive care
    Pittman, JAL
    Sum-Ping, JS
    Sherwood, MW
    El-Moalem, H
    Mark, JB
    INTENSIVE CARE MEDICINE, 2002, 28 : S78 - S78
  • [43] Usefulness of the Diagnosis "Decreased Cardiac Output (00029)" in Patients With Chronic Heart Failure
    Rojas Sanchez, Lyda Zoraya
    Hernandez Vargas, Juliana Alexandra
    Trujillo Caceres, Silvia Juliana
    Roa Diaz, Zayne Milena
    Jurado Arenales, Adriana Milena
    Toloza Perez, Yesith Guillermo
    INTERNATIONAL JOURNAL OF NURSING KNOWLEDGE, 2017, 28 (04) : 192 - 198
  • [44] Non-invasive measurement of cardiac output in patients with chronic heart failure
    Leslie, SJ
    McKee, S
    Newby, DE
    Webb, DJ
    Denvir, MA
    BLOOD PRESSURE MONITORING, 2004, 9 (05) : 277 - 280
  • [45] Accuracy of continuous cardiac output monitoring by pulse-contour-analysis in haemodynamic unstable patients
    Goedje, O
    Friedl, R
    Hannekum, A
    INTENSIVE CARE MEDICINE, 2001, 27 : S258 - S258
  • [46] Assessment of cardiac output measurement in dogs by transpulmonary pulse contour analysis
    Shih, Andre
    Maisenbacher, Herbert W.
    Bandt, Carsten
    Ricco, Carolina
    Bailey, James
    Rivera, Jess
    Estrada, Amara
    JOURNAL OF VETERINARY EMERGENCY AND CRITICAL CARE, 2011, 21 (04) : 321 - 327
  • [47] CONTINUOUS MEASUREMENT OF CARDIAC-OUTPUT USING PULSE CONTOUR ANALYSIS
    IRLBECK, M
    FORST, H
    BRIEGEL, J
    HALLER, M
    PETER, K
    ANAESTHESIST, 1995, 44 (07): : 493 - 500
  • [48] ESTIMATION OF CARDIAC OUTPUT BY NINE DIFFERENT PULSE CONTOUR ALGORITHMS IN CARDIAC SURGERY PATIENTS
    Broch, O.
    Carstens, A.
    Gruenewald, M.
    Albrecht, M.
    Renner, J.
    ANESTHESIA AND ANALGESIA, 2014, 118 : S57 - S57
  • [49] Monitoring cardiac output in beating heart coronary artery bypass graft surgery: use of pulse contour cardiac output
    TML Chan
    R Bhagrath
    EMC Ashley
    Critical Care, 5 (Suppl 6):
  • [50] CARDIAC OUTPUT ANALYSIS WITH MULTIBEAT ANALYSIS OF ARTERIAL WAVEFORM IN HEART FAILURE PATIENTS
    Ji, Yoon
    Grigsby, Joel
    Agashe, Harsha
    Proch, Katy
    Flynn, Brigid
    CRITICAL CARE MEDICINE, 2023, 51 (01) : 83 - 83