Method for correction of rotation errors in Micro-CT System

被引:11
|
作者
Zhao, Jintao [1 ]
Hu, Xiaodong [1 ,2 ]
Zou, Jing [1 ,2 ]
Zhao, Gengyan [1 ]
Lv, Hanyu [1 ]
Xu, Linyan [1 ]
Xu, Ying [1 ]
Hu, Xiaotang [1 ]
机构
[1] Tianjin Univ, State Key Lab Precis Measuring Technol & Instrume, Tianjin 300072, Peoples R China
[2] BCMIIS, Beijing 100089, Peoples R China
基金
中国国家自然科学基金;
关键词
Micro-CT; Rotation errors; Precision machined cylinder; Resolution; Capacitive sensor; CONE-BEAM CT; COMPUTED-TOMOGRAPHY; CALIBRATION; RESOLUTION; MISALIGNMENT;
D O I
10.1016/j.nima.2016.01.051
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
In Micro-CT (Computed Tomography) system, a series of projection data of sample are collected by the detector as the precision stage rotates step by step. However, the accuracy of projection images is limited by rotation errors during the acquisition process. Therefore, evaluating the performance of precision rotary stage and developing corresponding compensation method are necessary in Micro-CT system. In this paper, a metered system is designed which is composed of four precision capacitive sensors, a precision machined steel cylinder and four flexible hinges. Based on the metered system, a method to calibrate and correct the errors when the precision stage turns is proposed. Firstly, the theoretical analysis is proposed and the imperfect situations are considered. And then, the method has been applied to correct experimental data taken from a microscope type of Micro-CT system. Successful results are shown through evaluating MTF (Modulation Transfer Function) of Micro-CT system. Lastly, a sample of tungsten wire is scanned and the reconstructed images are compared before and after using the calibrated method. (C) 2016 Elsevier B.V. All rights reserved.
引用
收藏
页码:149 / 159
页数:11
相关论文
共 50 条
  • [1] Posture correction of dual-energy micro-CT system
    Zhang, Qi
    Zou, Jing
    Zhao, Jintao
    International Journal of Optomechatronics, 2024, 18 (01):
  • [2] Beam hardening correction and quantitative micro-CT
    Prevrhal, S
    DEVELOPMENTS IN X-RAY TOMOGRAPHY IV, 2004, 5535 : 152 - 161
  • [3] The effect of scatter correction on dual energy micro-CT
    Clark, D.
    Johnston, S. M.
    Johnson, G. A.
    Badea, C. T.
    MEDICAL IMAGING 2013: PHYSICS OF MEDICAL IMAGING, 2013, 8668
  • [4] Volumetric micro-CT system for in vivo microscopy
    Badea, CT
    Hedlund, LW
    Wheeler, CT
    Mai, W
    Johnson, GA
    2004 2ND IEEE INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING: MACRO TO NANO, VOLS 1 and 2, 2004, : 1377 - 1380
  • [5] Performance Evaluation of a Prototype Micro-CT System
    Cho, Seungryong
    Bian, Junguo
    Pelizzari, Charles A.
    Souris, Jeffrey S.
    Chen, Chin-Tu
    Pan, Xiaochuan
    2006 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOL 1-6, 2006, : 3127 - 3129
  • [6] A micro-CT system for little animal imaging
    Zhong, AJ
    Ye, C
    Luo, QM
    PHOTONICS AND IMAGING IN BIOLOGY AND MEDICINE, 2003, 5254 : 330 - 335
  • [7] Measurement of spatial resolution of the micro-CT system
    Rong Jun-Yan
    Fu Guo-Tao
    Wei Cun-Feng
    Sun Cui-Li
    Wei Long
    CHINESE PHYSICS C, 2010, 34 (03) : 412 - 416
  • [8] Measurement of spatial resolution of the micro-CT system
    戎军艳
    付国涛
    魏存峰
    孙翠丽
    魏龙
    Chinese Physics C, 2010, (03) : 412 - 416
  • [9] Correction of micro-CT image geometric artefacts based on marker
    Tao, Huanjie
    Lu, Xiaobo
    IET IMAGE PROCESSING, 2018, 12 (06) : 888 - 895
  • [10] A Flat-Field Correction Method for Photon-Counting-Detector-based Micro-CT
    Park, So E.
    Kim, Jae G.
    Hegazy, M. A. A.
    Cho, Min H.
    Lee, Soo Y.
    MEDICAL IMAGING 2014: PHYSICS OF MEDICAL IMAGING, 2014, 9033