Wear testing of crosslinked polyethylene: Wear rate variability and microbial contamination

被引:12
|
作者
Brandt, J. -M. [1 ,2 ,4 ]
Vecherya, A. [3 ]
Guenther, L. E. [3 ]
Koval, S. F. [4 ]
Petrak, M. J. [1 ,3 ]
Bohm, E. R. [1 ]
Wyss, U. P. [1 ,2 ]
机构
[1] Concordia Joint Replacement Grp, Winnipeg, MB R2K 2M9, Canada
[2] Univ Manitoba, Dept Mech & Mfg Engn, Winnipeg, MB R3T 2N2, Canada
[3] Orthopaed Innovat Ctr Inc, Winnipeg, MB R2K 2M9, Canada
[4] Univ Western Ontario, Dept Microbiol & Immunol, London, ON N6A 5C1, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
Polyethylene; Wear testing; Pin-on-disc testing; Hip simulator; Knee simulator; FEMORAL KNEE COMPONENTS; ACETABULAR CUPS; SIMULATOR; HIP; ARTHROPLASTY;
D O I
10.1016/j.jmbbm.2014.02.016
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The wear performance of two types of crosslinked polyethylene (Marathon (TM) and XLIC (TM), DePuy Synthes Inc., Warsaw, IN) was evaluated in a pin-on-disc wear tester, a hip wear simulator, and a knee wear simulator. Sodium azide was used as the microbial inhibitor in the calf serum-based lubricant. In the pin-on-disc wear tester, the Marathon wear rate of 5.33 +/- 0.54 mm(3)/Mc was significantly lower (p=0.002) than the wear rate of 6.43 +/- 0.60 mm(3)/Mc for XLK. Inversely, the Marathon wear rate of 15.07 +/- 1.03 mm(3)/Mc from the hip wear simulator was 2.2-times greater than the XLK wear rate of 6.71 +/- 1.03 mm(3)/Mc from the knee wear simulator. Differences in implant design, conformity, GUR type, and kinematic test conditions were suggested to account for the difference between the wear rates generated in the different types of wear testing apparati. In all wear tests, sodium azide was ineffective at inhibiting microbial growth in the lubricant. Eight different organisms were identified in the lubricant samples from the wear tests, which suggested the necessity of using an alternative, more effective microbial inhibitor. Careful sample preparation and thorough cleaning has shown to improve the consistency of the wear results. The wear rates generated in the hip and knee wear simulators closely reflected the wear behaviour of Marathon and XLK reported in published data that were tested under similar conditions. Crown Copyright (C) 2014 Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:208 / 216
页数:9
相关论文
共 50 条
  • [21] Enhanced wear performance of ultra high molecular weight polyethylene crosslinked by organosilan
    Xie, XL
    Tang, CY
    Wu, XC
    Li, RKY
    Wu, H
    CONTRIBUTIONS OF SURFACE ENGINEERING TO MODERN MANUFACTURING AND REMANUFACTURING, 2002, : 520 - 523
  • [22] In vivo wear of bilateral total hip replacements: conventional versus crosslinked polyethylene
    Heisel, C
    Silva, M
    Schmalzried, TP
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2005, 125 (08) : 555 - 557
  • [23] Highly Crosslinked Polyethylene Reduces Wear in Total Hip Arthroplasty at 5 Years
    Mutimer, Jonathan
    Devane, Peter A.
    Adams, Kathryn
    Horne, J. Geoffrey
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2010, 468 (12) : 3228 - 3233
  • [24] In vivo wear of bilateral total hip replacements: conventional versus crosslinked polyethylene
    Christian Heisel
    Mauricio Silva
    Thomas P. Schmalzried
    Archives of Orthopaedic and Trauma Surgery, 2005, 125 : 555 - 557
  • [25] The effect of cross-shear on the wear of virgin and highly-crosslinked polyethylene
    Korduba, L. A.
    Wang, A.
    WEAR, 2011, 271 (9-10) : 1220 - 1223
  • [26] Enhanced wear performance of ultra high molecular weight polyethylene crosslinked by organosilane
    C. Y. Tang
    X. L. Xie
    X. C. Wu
    R. K. Y. Li
    Y. W. Mai
    Journal of Materials Science: Materials in Medicine, 2002, 13 : 1065 - 1069
  • [27] Wear of highly crosslinked polyethylene against cobalt chrome and ceramic femoral heads
    Fisher, J.
    Jennings, L. M.
    Galvin, A. L.
    BIOCERAMICS AND ALTERNATIVE BEARINGS IN JOINT ARTHROPLASTY, 2006, : 185 - +
  • [28] Highly Crosslinked Polyethylene Improves Wear But Not Surface Damage in Retrieved Acetabular Liners
    Pang, Hee-Nee
    Naudie, Douglas D. R.
    McCalden, Richard W.
    MacDonald, Steven J.
    Teeter, Matthew G.
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2015, 473 (02) : 463 - 468
  • [29] Wear in highly crosslinked polyethylenes
    Kurtz, Steven
    Medel, Francisco J.
    Manley, Michael
    CURRENT ORTHOPAEDICS, 2008, 22 (06): : 392 - 399
  • [30] MEASUREMENT OF POLYETHYLENE WEAR
    MAXWELL, H
    SAMPATH, SA
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1993, 75 (06): : 981 - 981