Pullout strength of pedicle screw is an important index used by surgeon and engineers for design and development of the implant. Increasing pullout strength is difficult since it is dependent on several factors such as density of bone, insertion technique etc. Current study uses a validated 2 D FE model of pullout strength experiment to study the effect of number of threads, inner and outer diameter and thread geometry on holding power of pedicle screw in a synthetic osteoporotic bone model. It was found that outer diameter has highest effect on pullout strength of pedicle screw. Coarser thread provides better holding power in an osteoporotic bone. This study can be used for developing screws with improved holding power in osteoporotic bone.
引用
收藏
页码:382 / 387
页数:6
相关论文
共 50 条
  • [11] Comparative Analysis of Effect of Density, Insertion Angle and Reinsertion on Pull-Out Strength of Single and Two Pedicle Screw Constructs Using Synthetic Bone Model
    Krishnan, Venkatesh
    Varghese, Vicky
    Kumar, Gurunathan Saravana
    ASIAN SPINE JOURNAL, 2016, 10 (03) : 414 - 421
  • [12] In vitro biomechanical study of pedicle screw pull-out strength based on different screw path preparation techniques
    Moldavsky, Mark
    Salloum, Kanaan
    Bucklen, Brandon
    Khalil, Saif
    Mehta, Jwalant S.
    INDIAN JOURNAL OF ORTHOPAEDICS, 2016, 50 (02) : 177 - 182
  • [13] A novel pedicle screw design to maximize screw-bone interface strength using finite element analysis and design of experiment techniques
    Prajapati, Arvind Kumar
    Ramavarma, Harikrishna Varma Parimanathukovilakom
    Kumar, Gurunathan Saravana
    Muraleedharan, Chirathody Vayalappil
    Divakar, Ganesh
    ASIAN SPINE JOURNAL, 2024,
  • [14] In vitro biomechanical study of pedicle screw pull-out strength based on different screw path preparation techniques
    Mark Moldavsky
    Kanaan Salloum
    Brandon Bucklen
    Saif Khalil
    Jwalant S. Mehta
    Indian Journal of Orthopaedics, 2016, 50 : 177 - 182
  • [15] Shape Memory Alloy Expandable Pedicle Screw to Enhance Fixation in Osteoporotic Bone: Primary Design and Finite Element Simulation
    Tabesh, Majid
    Goel, Vijay
    Elahinia, Mohammad H.
    JOURNAL OF MEDICAL DEVICES-TRANSACTIONS OF THE ASME, 2012, 6 (03):
  • [16] Trabecular bone strength with virtual osteoporotic simulation and finite element analysis.
    Ferreri, S
    Mittra, E
    Gruber, B
    Oin, Y
    JOURNAL OF BONE AND MINERAL RESEARCH, 2004, 19 : S224 - S224
  • [17] Comparison of pedicle screw pull-out forces using two vertebral bone material mappings: a patient-specific finite element modeling study
    Rouyin, Alireza
    Salmani, Mohammad Javad
    Arjmand, Navid
    2023 30TH NATIONAL AND 8TH INTERNATIONAL IRANIAN CONFERENCE ON BIOMEDICAL ENGINEERING, ICBME, 2023, : 33 - 38
  • [18] A Study on Self-tapping Design of Cortical Screw for Orthopaedic Application by Using Finite Element Analysis on Pull-out Strength and Bending Strength
    Thongnual, Phunyawat
    Khantachawana, Anak
    Tawonsawatruk, Tulyapruek
    2022 9TH INTERNATIONAL CONFERENCE ON BIOMEDICAL AND BIOINFORMATICS ENGINEERING, ICBBE 2022, 2022, : 209 - 214
  • [19] Optimization of Pedicle Screw Parameters for Enhancing Implant Stability Based on Finite Element Analysis
    Li, Qiang
    Zhao, Hu
    Yang, Jinshuai
    Song, Shihong
    Liu, Xuyan
    WORLD NEUROSURGERY, 2024, 183 : E345 - E354
  • [20] Quantitative dual-energy CT for phantomless evaluation of cancellous bone mineral density of the vertebral pedicle: correlation with pedicle screw pull-out strength
    Wichmann, Julian L.
    Booz, Christian
    Wesarg, Stefan
    Bauer, Ralf W.
    Kerl, J. Matthias
    Fischer, Sebastian
    Lehnert, Thomas
    Vogl, Thomas J.
    Khan, M. Fawad
    Kafchitsas, Konstantinos
    EUROPEAN RADIOLOGY, 2015, 25 (06) : 1714 - 1720