A pixelated Faraday cup for proton beam diagnostics

被引:3
|
作者
Papa, A. [1 ,2 ,4 ]
Bertschi, S. [3 ]
Kunzi, L. [2 ]
Signorelli, G. [4 ]
机构
[1] Univ Pisa, Dipartimento Fis Enrico Fermi, Largo B Pontecorvo 3, I-56127 Pisa, Italy
[2] Paul Scherrer Inst, CH-5232 Villigen, Switzerland
[3] Swiss Fed Inst Technol, Ramistr 101, CH-8092 Zurich, Switzerland
[4] INFN, Sez Pisa, Largo B Pontecorvo 3, I-56127 Pisa, Italy
关键词
Proton beam; Diagnostic tools; Faraday cup; CALIBRATION;
D O I
10.1016/j.nima.2018.10.021
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Diagnostics tools are largely used to tune and set up beams in a vast range of applications. They include particle and nuclear physics, medical physics up to material science and biology, just to mention a few. The requirements on the detectors used to monitor the beam condition accordingly with the specific case. We present here a pixelated Faraday cup we built and used for beam profile monitoring with the peculiar characteristic of simultaneously measuring the proton beam position, shape and intensity. The detector covers a large range of currents from a hundred mu A down to a few nA perfectly matching the range of the proton beam intensity delivered by a 1 MeV Cockcroft-Walton accelerator. The pixel size allows for a beam position determination better than 1 mm.
引用
收藏
页码:25 / 26
页数:2
相关论文
共 50 条
  • [31] MOVABLE FARADAY CUP FOR HIGH-INTENSITY ELECTRON-BEAM PULSES
    LUCHES, A
    NASSISI, V
    PERRONE, MR
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1985, 56 (05): : 758 - 759
  • [32] An improved strip-line fast Faraday cup for beam bunch measurements
    Mathew, Jose V.
    Bajaj, Aman
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2020, 91 (11):
  • [33] Transferring electron beam welding parameters using the enhanced modified faraday cup
    Lawrence Livermore National Laboratory, Livermore, CA
    不详
    Weld J (Miami Fla), 2007, 12 (388-s-398-s):
  • [34] Faraday cup with nanosecond response and adjustable impedance for fast electron beam characterization
    Hu, Jing
    Rovey, Joshua L.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2011, 82 (07):
  • [35] Transferring electron beam welding parameters using the enhanced modified faraday cup
    Palmer, T. A.
    Elmer, J. W.
    Nicklas, K. D.
    Mustaleski, T.
    WELDING JOURNAL, 2007, 86 (12) : 388S - 398S
  • [36] Development of a Faraday cup fast ion loss detector for keV beam ions
    Lazerson, Samuel A.
    Ellis, Robert
    Freeman, Chris
    Ilagan, Jessica
    Wang, Tianyao
    Shao, Lin
    Allen, Nicole
    Gates, David
    Neilson, Hutch
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2019, 90 (09):
  • [37] Faraday cup measurements of a laser-induced plasma for a laser-proton acceleration
    Park, Seong Hee
    Jeong, Young Uk
    Lee, Kitae
    Kim, Dong Hyun
    Cha, Yong-Ho
    Yoo, Byeong Duk
    Choi, Ilwoo
    Yu, Tae-Jun
    Lee, Yeung-Lak
    Hong, Kyung-Han
    Ko, Do-Kyeong
    Lee, Jongmin
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2006, 49 (01) : 342 - 346
  • [38] Consistency of Faraday cup and ionization chamber dosimetry of proton fields and the role of nuclear interactions
    Wulff, Joerg
    Paul, Anne
    Baecker, Claus Maximilian
    Baumann, Kilian-Simon
    Esser, Johannes Niklas
    Koska, Benjamin
    Timmermann, Beate
    Verbeek, Nico Gerd
    Baeumer, Christian
    MEDICAL PHYSICS, 2024, 51 (03) : 2277 - 2292
  • [39] Screens for the 70 GeV proton beam diagnostics
    Borovkov, S.D.
    Grishenkov, S.A.
    Konevskij, V.S.
    Krivonosov, E.V.
    Litvinov, L.A.
    Novikov, V.P.
    Serga, E.V.
    Kharlamov, A.V.
    Pribory i Tekhnika Eksperimenta, 1991, (04): : 33 - 36
  • [40] FARADAY-CUP MEASUREMENT OF PROTON TOTAL REACTION CROSS SECTIONS AT 100 MEV
    KIRKBY, P
    LINK, WT
    CANADIAN JOURNAL OF PHYSICS, 1966, 44 (08) : 1847 - &