Consistency of Faraday cup and ionization chamber dosimetry of proton fields and the role of nuclear interactions

被引:3
|
作者
Wulff, Joerg [1 ,2 ,3 ]
Paul, Anne [1 ,2 ,3 ,4 ]
Baecker, Claus Maximilian [1 ,2 ,3 ]
Baumann, Kilian-Simon [5 ,6 ,7 ]
Esser, Johannes Niklas [1 ,2 ,3 ]
Koska, Benjamin [1 ,2 ,3 ]
Timmermann, Beate [1 ,2 ,3 ,4 ,8 ,9 ]
Verbeek, Nico Gerd [1 ,2 ,3 ]
Baeumer, Christian [1 ,2 ,3 ,8 ,10 ,11 ]
机构
[1] West German Proton Therapy Ctr Essen WPE, Essen, Germany
[2] Univ Hosp Essen, Essen, Germany
[3] West German Canc Ctr WTZ, Essen, Germany
[4] Heinrich Heine Univ Dusseldorf, Dusseldorf, Germany
[5] Marburg Univ Hosp, Dept Radiotherapy & Radiat Oncol, Marburg, Germany
[6] Marburg Ion Beam Therapy Ctr MIT, Marburg, Germany
[7] Univ Appl Sci, Inst Med Phys & Radiat Protect, Giessen, Germany
[8] German Canc Consortium DKTK, Essen, Germany
[9] Univ Hosp Essen, Dept Particle Therapy, Essen, Germany
[10] Tech Univ Dortmund, Dept Phys, Dortmund, Germany
[11] West German Proton Therapy Ctr Essen WPE gGmbH, Muhlenbach 1, D-45147 Essen, Germany
基金
欧盟地平线“2020”;
关键词
Faraday cup; proton therapy; reference dosimetry; BEAM MONITOR CALIBRATION; DOSE-AREA PRODUCT; THERAPY FACILITY; MONTE-CARLO; RADIATION; GEANT4; MODELS; VALIDATION; EMISSION;
D O I
10.1002/mp.16819
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
BackgroundA Faraday cup (FC) facilitates a quite clean measurement of the proton fluence emerging from clinical spot-scanning nozzles with narrow pencil-beams. The utilization of FCs appears to be an attractive option for high dose rate delivery modes and the source models of Monte-Carlo (MC) dose engines. However, previous studies revealed discrepancies of 3%-6% between reference dosimetry with ionization chambers (ICs) and FC-based dosimetry. This has prevented the widespread use of FCs for dosimetry in proton therapy.PurposeThe current study aims at bridging the gap between FC dosimetry and IC dosimetry of proton fields delivered with spot-scanning treatment heads. Particularly, a novel method to evaluate FC measurements is introduced.MethodsA consistency check is formulated, which makes use of the energy balance and the reciprocity theorem. The measurement data comprise central-axis depth distributions of the absorbed dose of quasi-monochromatic fields with a width of about 28.5 cm and FC measurements of the reciprocal fields with a single spot. These data are complemented by a look-up of energy-range tables, the average Q-value of transmutations, and the escape energy carried away by neutrons and photons. The latter data are computed by MC simulations, which in turn are validated with measurements of the distal dose tail and neutron out-of-field doses. For comparison, the conventional approach of FC evaluation is performed, which computes absorbed dose from the product of fluence and stopping power. The results from the FC measurements are compared with the standard dosimetry protocols and improved reference dosimetry methods.ResultsThe deviation between the conventional FC-based dosimetry and the IC-based one according to standard dosimetry protocols was -4.7 (+/- 3.3)% for a 100 MeV field and -3.6 (+/- 3.5)% for 200 MeV, thereby agreeing within the reported uncertainties. The deviations could be reduced to -4.0 (+/- 2.9)% and -3.0 (+/- 3.1)% by adopting state-of-the-art reference dosimetry methods. The alternative approach using the energy balance gave deviations of only -1.9% (100 MeV) and -2.6% (200 MeV) using state-of-the-art dosimetry. The standard uncertainty of this novel approach was estimated to be about 2%.ConclusionsAn alternative concept has been established to determine the absorbed dose of monoenergetic proton fields with an FC. It eliminates the strong dependence of the conventional FC-based approach on the MC simulation of the stopping-power and of the secondary ions, which according to the study at hand is the major contributor to the underestimation of the absorbed dose. Some contributions to the uncertainty of the novel approach could potentially be reduced in future studies. This would allow for accurate consistency tests of conventional dosimetry procedures.
引用
收藏
页码:2277 / 2292
页数:16
相关论文
共 49 条
  • [1] Proton beam dosimetry: A comparison between the Faraday cup and an ionization chamber
    Cambria, R
    Herault, J
    Brassart, N
    Silari, M
    Chauvel, P
    PHYSICS IN MEDICINE AND BIOLOGY, 1997, 42 (06): : 1185 - 1196
  • [2] Proton Beam Dosimetry: a Comparison between a Plastic Scintillator, Ionization Chamber and Faraday Cup
    Ghergherehchi, Mitra
    Afarideh, Hossein
    Ghannadi, Mohammad
    Mohammadzadeh, Ahmad
    Aslani, Golam Reza
    Boghrati, Behzad
    JOURNAL OF RADIATION RESEARCH, 2010, 51 (04) : 423 - 430
  • [3] Consistency in quality correction factors for ionization chamber dosimetry in scanned proton beam therapy
    Sorriaux, Jefferson
    Testa, Mauro
    Paganetti, Harald
    Bertrand, Damien
    Lee, John Aldo
    Palmans, Hugo
    Vynckier, Stefaan
    Sterpin, Edmond
    MEDICAL PHYSICS, 2017, 44 (09) : 4919 - 4927
  • [4] FARADAY CUP DOSIMETRY IN A PROTON THERAPY BEAM WITHOUT COLLIMATION
    GRUSELL, E
    ISACSSON, U
    MONTELIUS, A
    MEDIN, J
    PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (11): : 1831 - 1840
  • [5] Comparison of Indiana University Cyclotron Facility Faraday cup proton dosimetry with radiochromic films, a calorimeter, and a calibrated ion chamber
    Jones, AZ
    Bloch, CD
    Hall, ER
    Hashemian, R
    Klein, SB
    von Przewoski, B
    Murray, KM
    Foster, CC
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1999, 46 (06) : 1762 - 1765
  • [6] Comparison of Indiana University cyclotron facility Faraday cup proton dosimetry with radiochromic films, a calorimeter, and a calibrated ion chamber
    Indiana Univ, Bloomington, United States
    IEEE Trans Nucl Sci, 6 I (1762-1765):
  • [7] A Fano test for proton beams and the influence of nuclear interactions on ionization chamber factors
    Lourenco, A.
    Bouchard, H.
    Galer, S.
    Royle, G.
    Palmans, H.
    RADIOTHERAPY AND ONCOLOGY, 2017, 123 : S31 - S32
  • [8] Dosimetry of pulsed clinical proton beams by a small ionization chamber
    Nohtomi, A
    Sakae, T
    Tsunashima, Y
    Kohno, R
    MEDICAL PHYSICS, 2001, 28 (07) : 1431 - 1435
  • [9] Reference dosimetry in a scanned pulsed proton beam using ionisation chambers and a Faraday cup
    Lorin, Stefan
    Grusell, Erik
    Tilly, Nina
    Medin, Joakim
    Kimstrand, Peter
    Glimelius, Bengt
    PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (13): : 3519 - 3529
  • [10] k(Q) factors for ionization chamber dosimetry in clinical proton beams
    Vatnitsky, SM
    Siebers, JV
    Miller, DW
    MEDICAL PHYSICS, 1996, 23 (01) : 25 - 31