Consistency of Faraday cup and ionization chamber dosimetry of proton fields and the role of nuclear interactions

被引:3
|
作者
Wulff, Joerg [1 ,2 ,3 ]
Paul, Anne [1 ,2 ,3 ,4 ]
Baecker, Claus Maximilian [1 ,2 ,3 ]
Baumann, Kilian-Simon [5 ,6 ,7 ]
Esser, Johannes Niklas [1 ,2 ,3 ]
Koska, Benjamin [1 ,2 ,3 ]
Timmermann, Beate [1 ,2 ,3 ,4 ,8 ,9 ]
Verbeek, Nico Gerd [1 ,2 ,3 ]
Baeumer, Christian [1 ,2 ,3 ,8 ,10 ,11 ]
机构
[1] West German Proton Therapy Ctr Essen WPE, Essen, Germany
[2] Univ Hosp Essen, Essen, Germany
[3] West German Canc Ctr WTZ, Essen, Germany
[4] Heinrich Heine Univ Dusseldorf, Dusseldorf, Germany
[5] Marburg Univ Hosp, Dept Radiotherapy & Radiat Oncol, Marburg, Germany
[6] Marburg Ion Beam Therapy Ctr MIT, Marburg, Germany
[7] Univ Appl Sci, Inst Med Phys & Radiat Protect, Giessen, Germany
[8] German Canc Consortium DKTK, Essen, Germany
[9] Univ Hosp Essen, Dept Particle Therapy, Essen, Germany
[10] Tech Univ Dortmund, Dept Phys, Dortmund, Germany
[11] West German Proton Therapy Ctr Essen WPE gGmbH, Muhlenbach 1, D-45147 Essen, Germany
基金
欧盟地平线“2020”;
关键词
Faraday cup; proton therapy; reference dosimetry; BEAM MONITOR CALIBRATION; DOSE-AREA PRODUCT; THERAPY FACILITY; MONTE-CARLO; RADIATION; GEANT4; MODELS; VALIDATION; EMISSION;
D O I
10.1002/mp.16819
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
BackgroundA Faraday cup (FC) facilitates a quite clean measurement of the proton fluence emerging from clinical spot-scanning nozzles with narrow pencil-beams. The utilization of FCs appears to be an attractive option for high dose rate delivery modes and the source models of Monte-Carlo (MC) dose engines. However, previous studies revealed discrepancies of 3%-6% between reference dosimetry with ionization chambers (ICs) and FC-based dosimetry. This has prevented the widespread use of FCs for dosimetry in proton therapy.PurposeThe current study aims at bridging the gap between FC dosimetry and IC dosimetry of proton fields delivered with spot-scanning treatment heads. Particularly, a novel method to evaluate FC measurements is introduced.MethodsA consistency check is formulated, which makes use of the energy balance and the reciprocity theorem. The measurement data comprise central-axis depth distributions of the absorbed dose of quasi-monochromatic fields with a width of about 28.5 cm and FC measurements of the reciprocal fields with a single spot. These data are complemented by a look-up of energy-range tables, the average Q-value of transmutations, and the escape energy carried away by neutrons and photons. The latter data are computed by MC simulations, which in turn are validated with measurements of the distal dose tail and neutron out-of-field doses. For comparison, the conventional approach of FC evaluation is performed, which computes absorbed dose from the product of fluence and stopping power. The results from the FC measurements are compared with the standard dosimetry protocols and improved reference dosimetry methods.ResultsThe deviation between the conventional FC-based dosimetry and the IC-based one according to standard dosimetry protocols was -4.7 (+/- 3.3)% for a 100 MeV field and -3.6 (+/- 3.5)% for 200 MeV, thereby agreeing within the reported uncertainties. The deviations could be reduced to -4.0 (+/- 2.9)% and -3.0 (+/- 3.1)% by adopting state-of-the-art reference dosimetry methods. The alternative approach using the energy balance gave deviations of only -1.9% (100 MeV) and -2.6% (200 MeV) using state-of-the-art dosimetry. The standard uncertainty of this novel approach was estimated to be about 2%.ConclusionsAn alternative concept has been established to determine the absorbed dose of monoenergetic proton fields with an FC. It eliminates the strong dependence of the conventional FC-based approach on the MC simulation of the stopping-power and of the secondary ions, which according to the study at hand is the major contributor to the underestimation of the absorbed dose. Some contributions to the uncertainty of the novel approach could potentially be reduced in future studies. This would allow for accurate consistency tests of conventional dosimetry procedures.
引用
收藏
页码:2277 / 2292
页数:16
相关论文
共 49 条
  • [31] Set-up and calibration of a triple ionization chamber system for dosimetry in mixed neutron/photon fields
    Becker, J.
    Brunckhorst, E.
    Roca, A.
    Stecher-Rasmussen, F.
    Moss, R.
    Boettger, R.
    Schmidt, R.
    PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (13): : 3715 - 3727
  • [32] OPTIMIZATION OF ELECTRET IONIZATION-CHAMBER FOR DOSIMETRY IN MIXED NEUTRON-GAMMA RADIATION-FIELDS
    DORSCHEL, B
    PRETZSCH, G
    RADIATION PROTECTION DOSIMETRY, 1984, 9 (01) : 43 - 47
  • [33] Verification of Stereotactic Radiation Therapy Plan with Small Fields by Using Portal Dosimetry and A26 Ionization Chamber
    Zhu, D.
    Huang, Z.
    Feng, Y.
    MEDICAL PHYSICS, 2018, 45 (06) : E331 - E331
  • [34] Correction factors for ionization chamber dosimetry in CyberKnife: Machine-specific, plan-class, and clinical fields
    Gago-Arias, Araceli
    Antolin, Elena
    Fayos-Ferrer, Francisco
    Simon, Roco
    Gonzalez-Castano, Diego M.
    Palmans, Hugo
    Sharpe, Peter
    Gomez, Faustino
    Pardo-Montero, Juan
    MEDICAL PHYSICS, 2013, 40 (01)
  • [35] Proton dosimetry in a magnetic field: Measurement and calculation of magnetic field correction factors for a plane-parallel ionization chamber
    Gebauer, Benjamin
    Baumann, Kilian-Simon
    Fuchs, Hermann
    Georg, Dietmar
    Oborn, Brad M.
    Looe, Hui-Khee
    Luehr, Armin
    MEDICAL PHYSICS, 2024, 51 (03) : 2293 - 2305
  • [36] Denoising proton reference dosimetry spectrum using a large area ionization chamber-physical basis and type A uncertainty
    Tan, Hong Qi
    Lew, Kah Seng
    Koh, Calvin Wei Yang
    Lee, Kang Hao
    Chua, Clifford Ghee Ann
    Wibawa, Andrew
    Master, Zubin
    Lee, James Cheow Lei
    Park, Sung Yong
    PHYSICS IN MEDICINE AND BIOLOGY, 2025, 70 (02):
  • [37] Proton beam dosimetry in the presence of magnetic fields using Farmer-type ionization chambers of different radii
    Marot, Mathieu
    Surla, Sonja
    Burke, Elisa
    Brons, Stephan
    Runz, Armin
    Greilich, Steffen
    Karger, Christian P.
    Jaekel, Oliver
    Burigo, Lucas N.
    MEDICAL PHYSICS, 2023, 50 (07) : 4590 - 4599
  • [38] Ionization chamber dosimetry of small photon fields:: a Monte Carlo study on stopping-power ratios for radiosurgery and IMRT beams
    Sánchez-Doblado, F
    Andreo, P
    Capote, R
    Leal, A
    Perucha, M
    Arráns, R
    Núñez, L
    Mainegra, E
    Lagares, JI
    Carrasco, E
    PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (14): : 2081 - 2099
  • [39] Dosimetry of a Low-energy Proton Beam Using a Plane-parallel Ionization Chamber in the 45-MeV Proton Beam Test Irradiation Facility
    Yoon, Sang-Pil
    Hong, In-Seok
    Park, Bum-sik
    Cho, Yong-Sub
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2011, 59 (02) : 572 - 576
  • [40] Comments on 'Ionization chamber dosimetry of small photon fields: a Monte Carlo study on stopping-power ratios for radiosurgery and IMRT beams'
    Seuntjens, J
    Verhaegen, F
    PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (21): : L43 - L45