Coherence scanning interferometry: measurement and correction of three-dimensional transfer and point-spread characteristics

被引:42
|
作者
Mandal, Rahul [1 ]
Coupland, Jeremy [1 ]
Leach, Richard [2 ]
Mansfield, Daniel [3 ]
机构
[1] Univ Loughborough, Wolfson Sch Mech & Mfg Engn, Loughborough LE11 3TU, Leics, England
[2] Natl Phys Lab, Engn Measurement Div, Teddington TW11 0LW, Middx, England
[3] Taylor Hobson Ltd, Leicester LE4 9JQ, Leics, England
基金
英国工程与自然科学研究理事会;
关键词
PROFILOMETRY;
D O I
10.1364/AO.53.001554
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
When applied to the measurement of smooth surfaces, coherence scanning interferometry can be described by a three-dimensional linear filtering operation that is characterized either by the point-spread function in the space domain or equivalently by the transfer function (TF) in the spatial frequency domain. For an ideal, aberration-free instrument, these characteristics are defined uniquely by the numerical aperture of the objective lens and the bandwidth of the illumination source. In practice, however, physical imperfections such as those in lens aberrations, reference focus, and source alignment mean that the instrument performance is not ideal. Currently, these imperfections often go unnoticed as the instrument performance is typically only verified using rectilinear artifacts such as step heights and lateral grids. If an object of varying slope is measured, however, significant errors are often observed as the surface gradient increases. In this paper, a new method of calibration and adjustment using a silica micro-sphere as a calibration artifact is introduced. The silica microsphere was used to compute the point-spread and TF characteristics of the instrument, and the effect of these characteristics on instrument performance is discussed. Finally, a straightforward method to correct for phase and amplitude imperfections in the TF is described using a modified inverse filter. (C) 2014 Optical Society of America
引用
收藏
页码:1554 / 1563
页数:10
相关论文
共 50 条
  • [21] Three-dimensional shape measurement system based on dual oscillating mirrors with point scanning
    Li X.-D.
    Cui L.
    Zhao H.-J.
    Jiang H.-Z.
    Guangxue Jingmi Gongcheng/Optics and Precision Engineering, 2010, 18 (07): : 1648 - 1653
  • [22] Coherence retrieval via three-dimensional intensity measurement
    Zhang, Runnan
    Cai, Zewei
    Sun, Jiasong
    Zhou, Ning
    Chen, Qian
    Zuo, Chao
    AOPC 2021: OPTICAL SENSING AND IMAGING TECHNOLOGY, 2021, 12065
  • [23] Three-dimensional displacement measurement in fluids using holographic interferometry
    Lopez, JM
    Moore, AJ
    SECOND IBEROAMERICAN MEETING ON OPTICS, 1996, 2730 : 106 - 109
  • [24] Three-dimensional displacement measurement using diffractive optic interferometry
    Gilbert, JA
    Shepherd, RL
    Cole, HJ
    Ashley, PR
    OPTICAL ENGINEERING, 1997, 36 (12) : 3336 - 3342
  • [25] Measurement of a Three-dimensional Gas Temperature Field with Holographic Interferometry
    CHEN Shaohua(Yibin Teacher’s College
    HUANG Suyi
    SHI Dufang(Huazhong University of Science and Technology
    SemiconductorPhotonicsandTechnology, 1996, (01) : 72 - 78
  • [26] Three-dimensional point spread function measurements of imaging spectrometers
    Jemec, Jurij
    Pernus, Franjo
    Likar, Bostjan
    Burmen, Miran
    JOURNAL OF OPTICS, 2017, 19 (09)
  • [27] A three-dimensional point spread function for phase retrieval and deconvolution
    Liu, Xinyue
    Wang, Liang
    Wang, Jianli
    Meng, Haoran
    OPTICS EXPRESS, 2012, 20 (14): : 15392 - 15405
  • [28] Image restoration for three-dimensional fluorescence microscopy using an orthonormal basis for efficient representation of depth-variant point-spread functions
    Patwary, Nurmohammed
    Preza, Chrysanthe
    BIOMEDICAL OPTICS EXPRESS, 2015, 6 (10): : 3826 - 3841
  • [29] Efficient Cross-sectional Evaluation Method of three-dimensional NAND Flash Memory by Cooperation of Coherence Scanning Interferometry and Scanning Electron Microscopy
    Konno, Azusa
    Fujimura, Ichiro
    Mise, Hiromi
    Aiso, Toru
    Takeuchi, Syuichi
    Koyama, Kazuhiro
    Sasajima, Masahiro
    METROLOGY, INSPECTION, AND PROCESS CONTROL XXXVI, 2022, 12053
  • [30] Measurement of the three-dimensional microscope point spread function using a Shack-Hartmann wavefront sensor
    Beverage, JL
    Shack, RV
    Descour, MR