We present a technique to measure the wavefront in the exit pupil of a microscope to determine the microscope's three-dimensional point spread function (PSF) experimentally. The wavefront yields the microscope PSF through a Fourier transform that models propagation of light from the exit pupil to the image plane. A Shack-Hartmann wavefront sensor is used to measure the wavefront shape by recording lateral displacements of a grid of focused spots created by a lenslet array. The displacement of each spot is related to the local wavefront slope. Thus, with appropriate sampling across the exit pupil, the entire wavefront can be reconstructed. This technique does not require the use of a sub-resolution object to obtain the three-dimensional microscope PSF. Consequently, larger, brighter fluorescent objects may be imaged, thereby reducing the requirements for detector sensitivity and leading to a three-fold increase in the axial range over which the PSF is measured. The Shack-Hartmann technique results in a description of the PSF as a continuous function whose sampling is not dependent on the size of the CCD pixels. The Shack-Hartmann sensor is not limited by the numerical aperture of the objective and can easily be calibrated to measure the PSF at any wavelength.
引用
收藏
页码:61 / 75
页数:15
相关论文
共 50 条
  • [1] Three-Dimensional Surface Profile Measurement of Microlenses Using the Shack-Hartmann Wavefront Sensor
    Li, Chenhui
    Hall, Gunnsteinn
    Zhu, Difeng
    Li, Heng
    Eliceiri, Kevin W.
    Jiang, Hongrui
    JOURNAL OF MICROELECTROMECHANICAL SYSTEMS, 2012, 21 (03) : 530 - 540
  • [2] Measurement of aberrations in microlenses using a Shack-Hartmann wavefront sensor
    Pulaski, PD
    Roller, JP
    Neal, DR
    Ratte, K
    CURRENT DEVELOPMENTS IN LENS DESIGN AND OPTICAL ENGINEERING III, 2002, 4767 : 44 - 52
  • [3] The Shack-Hartmann wavefront sensor
    Mansuripur, Masud
    Optics and Photonics News, 1999, 10 (04): : 48 - 51
  • [4] Measurement error of wavefront phase with Shack-Hartmann wavefront sensor
    Shen, Feng
    Jiang, Wenhan
    Guangxue Xuebao/Acta Optica Sinica, 2000, 20 (05): : 666 - 671
  • [5] Aerosol absorption measurement by a Shack-Hartmann wavefront sensor
    Land, Jay E.
    APPLIED OPTICS, 2023, 62 (18) : 4836 - 4847
  • [6] Transverse wind Measurement with Shack-Hartmann Wavefront Sensor
    Li, Zhenghan
    Li, Xinyang
    PROCEEDINGS OF 2017 3RD IEEE INTERNATIONAL CONFERENCE ON COMPUTER AND COMMUNICATIONS (ICCC), 2017, : 2745 - 2751
  • [7] Wavefront reconstruction for the Shack-Hartmann wavefront sensor
    Topa, DM
    OPTICAL DESIGN AND ANALYSIS SOFTWARE II, 2002, 4769 : 101 - 115
  • [8] Reconfigurable Shack-Hartmann wavefront sensor
    Rha, J
    Voelz, DG
    Giles, MK
    OPTICAL ENGINEERING, 2004, 43 (01) : 251 - 256
  • [9] Measurement of wavefront aberrations of human eyes with Shack-Hartmann wavefront sensor
    Cheng S.-Y.
    Cao Z.-L.
    Hu L.-F.
    Mu Q.-Q.
    Li P.-F.
    Xuan L.
    Guangxue Jingmi Gongcheng/Optics and Precision Engineering, 2010, 18 (05): : 1060 - 1067
  • [10] Polarized Shack-Hartmann wavefront sensor
    Yang, Yanrong
    Huang, Linhai
    Xiao, Yawei
    Gu, Naiting
    FRONTIERS IN PHYSICS, 2023, 11