Performance and long-term consistency of five Galliform 68Ge/68Ga generators used for clinical Ga-68 preparations over a 4 year period

被引:3
|
作者
Sammartano, Antonino [1 ]
Migliari, Silvia [1 ]
Scarlattei, Maura [1 ]
Baldari, Giorgio [1 ]
Serreli, Giulio [2 ]
Lazzara, Chiara [1 ]
Garau, Ludovico [1 ]
Ghetti, Caterina [2 ]
Ruffini, Livia [1 ]
机构
[1] Azienda Osped Univ Parma, Div Nucl Med, Diagnost Dept, Via Gramsci 14, I-43126 Parma, Italy
[2] Azienda Osped Univ Parma, Diagnost Dept, Med Phys Unit, Parma, Italy
关键词
68Ga-radiopharmaceuticals; 68Ge; 68Ga generator; radiochemical purity; RADIOPHARMACEUTICALS; VALIDATION; FUTURE;
D O I
10.1097/MNM.0000000000001545
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background Gallium-68 is a positron emitter for PET applications that can be produced without cyclotron by a germanium (Ge-68) chloride/gallium (Ga-68) chloride generator. Short half-life (67.71 min) of Ga-68, matching pharmacokinetic properties of small biomolecules, facilitates isotope utilization in compounding radiopharmaceuticals for PET imaging. The increasing cost of good manufacturing practice-compliant generators has strengthened the need for radionuclide efficient use by planning specific radiopharmaceutical sessions during the week, careful maintenance of the generator and achievement of high labeling yield and radiochemical purity (RCP) of the radiolabeled molecules. Methods The aim of this study was to evaluate the annual performance of five consecutive Ge-68/Ga-68 generators used for small-scale preparations of 68Ga-radiopharmaceuticals. To assess the long-term efficiency of isotope production we measured the weekly elution yield. To assess process efficiency we measured elution yield, labeling yield and RCP of four radiopharmaceutical preparations (Ga-68-DOTATOC, Ga-68-PSMA-HBED-CC, Ga-68-PENTIXAFOR and Ga-68-DOTATATE). Results The annual mean elution yield of the generators was 74.7%, higher than that indicated by the manufacturer, and it never went below 65%. The Ge-68 level in the final products was under the detection limits in all the produced batches (mean value 0.0000048%). The RCP of radiopharmaceuticals determined by high-performance liquid chromatography was 98 +/- 0.22%. The mean yield of radiolabelling was 64.68, 68.71, 57 and 63.68% for Ga-68-DOTATOC, Ga-68-PSMA-HBED-CC, (68)GaPENTIXAFOR and Ga-68-DOTATATE. Conclusion The ability to prepare in the hospital radiopharmacy high-purity and pharmaceutically acceptable Ga-68-radiolabeled probes on a routine basis facilitates patient access to precision imaging for clinical and research aims.
引用
收藏
页码:568 / 576
页数:9
相关论文
共 50 条
  • [1] Characterization of 68Ge/68Ga ITG GmbH generators and their long-term validation
    Tworowska, I.
    Schultz, M.
    Ranganathan, D.
    Thamake, S.
    Delpassand, E.
    Zhernosekov, K.
    Marx, S.
    JOURNAL OF NUCLEAR MEDICINE, 2015, 56
  • [2] 68Ge/68Ga generators and tracer production
    Velikyan, I.
    JOURNAL OF NUCLEAR MEDICINE, 2015, 56
  • [3] Instantaneous quantitative determination of 68Ge breakthrough of 68Ge/68Ga generators
    Roesch, Frank
    Loktionova, Natalia
    JOURNAL OF NUCLEAR MEDICINE, 2012, 53
  • [4] Long-term evaluation of TiO2-based 68Ge/68Ga generators and optimized automation of [68Ga]DOTATOC radiosynthesis
    Lin, Mai
    Ranganathan, David
    Mori, Tetsuya
    Hagooly, Aviv
    Rossin, Raffaella
    Welch, Michael J.
    Lapi, Suzanne E.
    APPLIED RADIATION AND ISOTOPES, 2012, 70 (10) : 2539 - 2544
  • [5] Past, present and future of 68Ge/68Ga generators
    Roesch, F.
    APPLIED RADIATION AND ISOTOPES, 2013, 76 : 24 - 30
  • [6] Quantitative determination of 68Ge breakthrough of 68Ge/68Ga generators via TLC
    Eppard, E.
    Loktionova, N. S.
    Roesch, F.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2012, 39 : S261 - S262
  • [7] Studies on the separation of Ge and Ga with inorganic materials for the preparation of 68Ge/68Ga generators
    Brambilla, Tania P.
    Osso, Jr Joao A.
    JOURNAL OF LABELLED COMPOUNDS & RADIOPHARMACEUTICALS, 2011, 54 : S400 - S400
  • [8] Characterization of SnO2-based 68Ge/68Ga generators and 68Ga-DOTATATE preparations: radionuclide purity, radiochemical yield and long-term constancy
    Sudbrock, Ferdinand
    Fischer, Thomas
    Zimmermanns, Beate
    Guliyev, Mehrab
    Dietlein, Markus
    Drzezga, Alexander
    Schomaecker, Klaus
    EJNMMI RESEARCH, 2014, 4 : 1 - 10
  • [9] Characterization of SnO2-based 68Ge/68Ga generators and 68Ga-DOTATATE preparations: radionuclide purity, radiochemical yield and long-term constancy
    Ferdinand Sudbrock
    Thomas Fischer
    Beate Zimmermanns
    Mehrab Guliyev
    Markus Dietlein
    Alexander Drzezga
    Klaus Schomäcker
    EJNMMI Research, 4
  • [10] Long-term evaluation of TiO2-based 68Ge/68Ga generators and optimized automation of 68Ga-DOTATOC radiosynthesis
    Lin, Mai
    Ranganathan, David
    Lapi, Suzanne
    Welch, Michael
    JOURNAL OF NUCLEAR MEDICINE, 2012, 53