Evaluation of the i-STAT portable clinical analyzer for use in a helicopter

被引:17
|
作者
Burritt, MF [1 ]
Santrach, PJ [1 ]
Hankins, DG [1 ]
Herr, D [1 ]
Newton, NC [1 ]
机构
[1] MAYO CLIN & MAYO FDN,DIV EMERGENCY MED SERV & INTERNAL MED,ROCHESTER,MN 55905
关键词
electrolyte analysis; glucose; helicopter; point-of-care testing; single-use cartridge;
D O I
10.3109/00365519609088630
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
We evaluated the performance of the i-STAT Portable Clinical Analyzer (PCA) for use in a helicopter with the analyses performed by the flight nurses. Imprecision and initial split-sample comparative studies were performed in the Hospital Laboratory and Clinical Chemistry Laboratory. Day-to day imprecision (CV) ranged from 0.5 to 6.7% in the clinical laboratories and 0.5 to 5.3% in the helicopter with urea nitrogen and glucose determinations giving the highest values. Comparison studies between the PCA and the Kodak Ektachem E700 and the Beckman Astra 8 gave acceptable results, although urea nitrogen did show a bias in the high range (mean difference 4.91 and 3.69 mmol/L respectively). It was not considered medically significant for the anticipated patient population of the PCA in our hospitals. Comparison studies of patient specimens between the PCA in the helicopter and the clinical laboratories showed no significant differences, with the exception of glucose. Lower glucose values were obtained in the laboratory due to the time lag between specimen collection in the helicopter and analysis in the laboratory.
引用
收藏
页码:121 / 128
页数:8
相关论文
共 50 条
  • [31] Precision Evaluation of the i-STAT® Glucose Test in the i-STAT G Cartridge using the i-STAT 1 System for Venous, Arterial and Capillary Whole Blood Specimens in Clinical Settings
    Thibodeau, L.
    Aguirre, M. Gonzalez
    Hahn, S.
    Zhang, H.
    CLINICAL CHEMISTRY, 2024, 70 : I129 - I129
  • [32] Evaluation of the i-STAT portable point-of-care analyzer for determination of blood gases and acid-base status in newborn calves
    Bleul, Ulrich
    Goetz, Eva
    JOURNAL OF VETERINARY EMERGENCY AND CRITICAL CARE, 2014, 24 (05) : 519 - 528
  • [33] Evaluation of portable point of care instrument:: the i-Stat®.: Report of 7,000 analyses
    Feuillu, A
    Morel, I
    Mollard, JF
    ANNALES DE BIOLOGIE CLINIQUE, 2002, 60 (02) : 153 - 164
  • [34] A novel developmental physiology laboratory exercise using the i-STAT point-of-care clinical analyzer
    Murphy, MJ
    FASEB JOURNAL, 2003, 17 (04): : A8 - A8
  • [35] Correlation study between four i-Stat portable analyzers.
    Dinovo, EC
    Lee, KD
    Shin, WY
    CLINICAL CHEMISTRY, 1999, 45 (06) : A29 - A29
  • [36] Effect of emicizumab on activated clotting time performed on i-STAT Alinity analyzer
    Seyve, Landry
    Prigent, Jean Baptiste
    Lo Presti, Caroline
    Bedague, Damien
    Szymanski, Gautier
    Bulabois, Benedicte
    Barro, Claire
    Marlu, Raphael
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2024, 46 (06) : 1132 - 1135
  • [37] Analytical Performances of the Novel i-STAT Alinity Point-of-Care Analyzer
    Larcher, Romaric
    Lottelier, Maxence
    Badiou, Stephanie
    Dupuy, Anne-Marie
    Bargnoux, Anne-Sophie
    Cristol, Jean-Paul
    DIAGNOSTICS, 2023, 13 (02)
  • [38] Evaluation of the IRMA TRUpoint and i-STAT creatinine assays
    Nichols, James H.
    Bartholomew, Cathy
    Bonzagi, Arianna
    Garb, Jane L.
    Jin, Long
    CLINICA CHIMICA ACTA, 2007, 377 (1-2) : 201 - 205
  • [39] Improvised Method for Increasing the Temperature of an i-STAT Analyzer and Cartridge in Cold Environments
    Radler, David
    Wetschler, Matthew
    Christensen, Mark
    Lipman, Grant S.
    WILDERNESS & ENVIRONMENTAL MEDICINE, 2015, 26 (02) : 280 - 282
  • [40] Selected Blood Chemistry and Gas Reference Ranges for Broiler Breeders Using the i-STAT® Handheld Clinical Analyzer
    Martin, Michael P.
    Wineland, Michael
    Barnes, H. John
    AVIAN DISEASES, 2010, 54 (03) : 1016 - 1020