Optimizing dual energy cone beam CT protocols for preclinical imaging and radiation research

被引:15
|
作者
Schyns, Lotte E. J. R. [1 ]
Almeida, Isabel P. [1 ]
van Hoof, Stefan J. [1 ]
Descamps, Benedicte [2 ]
Vanhove, Christian [2 ]
Landry, Guillaume [3 ]
Granton, Patrick V. [1 ]
Verhaegen, Frank [1 ,4 ]
机构
[1] Maastricht Univ, GROW Sch Oncol & Dev Biol, Dept Radiat Oncol MAASTRO, Med Ctr, Maastricht, Netherlands
[2] iMinds Med IT IBiTech MEDISIP INFIN, De Pintelaan 185, Ghent, Belgium
[3] Ludwig Maximilians Univ Munchen LMU Munich, Fac Phys, Dept Med Phys, Munich, Germany
[4] McGill Univ, Dept Oncol, Med Phys Unit, Montreal, PQ, Canada
来源
BRITISH JOURNAL OF RADIOLOGY | 2017年 / 90卷 / 1069期
关键词
CARLO DOSE CALCULATIONS; X-RAY TARGETS; TISSUE SEGMENTATION; COMPUTED-TOMOGRAPHY; ATOMIC NUMBERS; BRACHYTHERAPY; RADIOTHERAPY; SIMULATION; SPECTRA; ANIMALS;
D O I
10.1259/bjr.20160480
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: The aim of this work was to investigate whether quantitative dual-energy CT (DECT) imaging is feasible for small animal irradiators with an integrated cone-beam CT (CBCT) system. Methods: The optimal imaging protocols were determined by analyzing different energy combinations and dose levels. The influence of beam hardening effects and the performance of a beam hardening correction (BHC) were investigated. In addition, two systems from different manufacturers were compared in terms of errors in the extracted effective atomic numbers (Z(eff)) and relative electron densities (rho(e)) for phantom inserts with known elemental compositions and relative electron densities. Results: The optimal energy combination was determined to be 50 and 90kVp. For this combination, Z(eff) and r rho(e) can be extracted with a mean error of 0.11 and 0.010, respectively, at a dose level of 60cGy. Conclusion: Quantitative DECT imaging is feasible for small animal irradiators with an integrated CBCT system. To obtain the best results, optimizing the imaging protocols is required. Well-separated X-ray spectra and a sufficient dose level should be used to minimize the error and noise for Z(eff) and rho(e). When no BHC is applied in the image reconstruction, the size of the calibration phantom should match the size of the imaged object to limit the influence of beam hardening effects. No significant differences in Z(eff) and rho(e) errors are observed between the two systems from different manufacturers. Advances in knowledge: This is the first study that investigates quantitative DECT imaging for small animal irradiators with an integrated CBCT system.
引用
收藏
页数:10
相关论文
共 50 条
  • [1] A generalizable new figure of merit for dose optimization in dual energy cone beam CT scanning protocols
    Li, Chenguang
    Zhou, Li
    Deng, Jun
    Wu, Hao
    Wang, Ruoxi
    Wang, Fei
    Yao, Kaining
    Chen, Chen
    Niu, Tianye
    Zhang, Yibao
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2023, 68 (18):
  • [2] Dose assessment of cone beam CT imaging protocols as part of SPECT/CT examinations
    Tonkopi, E.
    Ross, A. A.
    [J]. MEDICAL PHYSICS, 2014, 41 (08) : 7 - 8
  • [3] Evaluation of Imaging Dose for Modulation Layer Based Dual Energy Cone-Beam CT
    Ju, Eunbin
    Ahn, SoHyun
    Cho, Samju
    Keum, Ki Chang
    Lee, Rena
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3387 - 3387
  • [4] Dual Energy CT Imaging in Cone-Beam Micro-CT for Improved Attenuation Coefficient Measurement
    唐智伟
    胡广书
    [J]. Tsinghua Science and Technology, 2011, 16 (04) : 352 - 357
  • [5] Enabling Conventional Cone Beam CT with the Capability of Dual Energy Imaging Using a Simple Add-On Beam Modifier
    Vinke, R.
    Takao, S.
    Umegaki, K.
    Shirato, H.
    Peng, H.
    Xing, L.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3457 - 3457
  • [6] Small animal photon counting cone-beam CT on a preclinical radiation research platform to improve radiation dose calculation accuracy
    Hu, Xiaoyu
    Zhong, Yuncheng
    Lai, Youfang
    Shen, Chenyang
    Yang, Kai
    Jia, Xun
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2022, 67 (19):
  • [7] Dual-energy cone-beam Micro-CT for animal imaging: Preliminary study
    Cho, Seungryong
    Sidky, Emil
    Bian, Junguo
    Pan, Xiaochuan
    [J]. MEDICAL IMAGING 2007: PHYSICS OF MEDICAL IMAGING, PTS 1-3, 2007, 6510
  • [8] Dual-Energy CT Imaging in Diagnostic Imaging and Radiation Therapy
    Molloi, S.
    Li, B.
    Yin, F.
    Chen, H.
    [J]. MEDICAL PHYSICS, 2014, 41 (06) : 486 - 487
  • [9] Protocols for Dual Tracer PET/SPECT Preclinical Imaging
    Blower, Julia E.
    Bordoloi, Jayanta K.
    Rigby, Alex
    Farleigh, Matthew
    Kim, Jana
    O'Brien, Hugh
    Jackson, Jonathan
    Poyiatzis, Constantinos
    Bezer, James
    Sunassee, Kavitha
    Blower, Philip J.
    Livieratos, Lefteris
    [J]. FRONTIERS IN PHYSICS, 2020, 8
  • [10] Single Scan dual energy Cone Beam CT Using a Rotating Filter
    Fang, Chengyijue
    Xu, Gongming
    Zhu, Lei
    [J]. MEDICAL IMAGING 2020: PHYSICS OF MEDICAL IMAGING, 2020, 11312