Technical Note: Patient-weight dependence of the effective dose conversion coefficients for diagnostic x-ray imaging procedures

被引:0
|
作者
Golikov, Vladislav [1 ]
Druzhinina, Polina [1 ]
机构
[1] St Petersburg Res Inst Radiat Hyg, Mira Str 8, St Petersburg 197101, Russia
关键词
conversion coefficients; effective dose; patients; x-ray examination; PEDIATRIC BARIUM MEAL; COMPUTATIONAL PHANTOMS; ORGAN; ADULT; AGE;
D O I
10.1002/mp.14446
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose The aim of this study was to provide a method and reference data for the evaluation of the weight dependence of the conversion coefficients from air collision kerma integrated over beam area to the effective dose for patients undergoing diagnostic x-ray examinations. Method A simple physics-based analytical model was developed in order to calculate the weight dependence of the effective dose conversion coefficients based on existing values of the conversion coefficients between patient effective dose and air collision kerma integrated over beam area calculated by the Monte Carlo method on the base of computational MIRD-5-type stylized models. Results The analytical expressions that make possible taking into account a patient's weight when assessing the effective dose were presented. The following categories of diagnostic x-ray examinations are considered: CT; plain radiography; cardiac angiography; fluoroscopy. The results indicate that the weight dependence of the effective dose conversion coefficients for different categories of diagnostic x-ray examinations has the same form. Correction factor for evaluation of the new value of the effective dose conversion coefficient for a patient with weightW(pa.)when we know its value for a standard patient with weightW(st.)depends on the ratios of their weights (W-st./W-pat.) and fractions of the incident energy which is imparted to a patient. Conclusions The paper presents an approach and necessary data for the estimation of the weight dependence of the effective dose conversion coefficients for patients undergoing different diagnostic x-ray examinations.
引用
收藏
页码:5366 / 5372
页数:7
相关论文
共 50 条
  • [41] Evaluation of a new very low dose imaging protocol: feasibility and impact on X-ray dose levels in electrophysiology procedures
    Bourier, Felix
    Reents, Tilko
    Ammar-Busch, Sonia
    Buiatti, Alessandra
    Kottmaier, Marc
    Semmler, Verena
    Telishevska, Marta
    Brkic, Amir
    Grebmer, Christian
    Lennerz, Carsten
    Kolb, Christof
    Hessling, Gabriele
    Deisenhofer, Isabel
    EUROPACE, 2016, 18 (09): : 1406 - 1410
  • [42] Determination of conversion coefficients between air kerma or photon fluence and ambient dose equivalent for diagnostic X ray beams
    Nogueira, MS
    Campos, LL
    Mota, HC
    RADIATION PROTECTION DOSIMETRY, 1999, 81 (02) : 123 - 132
  • [43] Mean glandular dose coefficients (DgN) for x-ray spectra used in contemporary breast imaging systems
    Nosratieh, Anita
    Hernandez, Andrew
    Shen, Sam Z.
    Yaffe, Martin J.
    Seibert, J. Anthony
    Boone, John M.
    PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (18): : 7179 - 7190
  • [44] CAN HALF VALUE LAYERMEASUREMENTS BE USED TOGETHERWITH THE EFFECTIVE ENERGY TO OBTAIN CONVERSION COEFFICIENTS FOR X-RAY SPECTRA?
    Behnke, B.
    Hupe, O.
    RADIATION PROTECTION DOSIMETRY, 2017, 173 (04) : 277 - 285
  • [45] Patient effective dose evaluation for chest X-ray examination in three digital radiography centers
    Paydar, R.
    Takavar, A.
    Kardan, M. R.
    Babakhani, A.
    Deevband, M. R.
    Saber, S.
    IRANIAN JOURNAL OF RADIATION RESEARCH, 2012, 10 (3-4): : 139 - 143
  • [46] MODERN IMAGING PROCEDURES AND DIAGNOSTIC AND THERAPEUTIC METHODS - INFLUENCE ON STRUCTURE AND METHOD OF WORKING OF AN X-RAY DEPARTMENT
    POHLENZ, O
    COMPUTERTOMOGRAPHIE, 1981, 1 (03): : 117 - 124
  • [47] Effective dose in diagnostic radiology as a function of x-ray beam filtration for a constant exit dose and constant film density
    Behrman, RH
    Yasuda, G
    MEDICAL PHYSICS, 1998, 25 (05) : 780 - 790
  • [48] SPOT REGION OF INTEREST IMAGING: A NOVEL FUNCTIONALITY AIMED AT X-RAY DOSE REDUCTION IN NEUROINTERVENTIONAL PROCEDURES
    Borota, Ljubisa
    Patz, Andreas
    RADIATION PROTECTION DOSIMETRY, 2020, 188 (03) : 322 - 331
  • [49] PRACTICAL COMPARISON OF DIFFERENT PATIENT'S DOSE ESTIMATION METHODS DURING MEDICAL X-RAY PROCEDURES
    Jariomenko, Timur
    Medvedevas, Nikolajus
    MEDICAL PHYSICS IN THE BALTIC STATES, 2019, : 39 - 42
  • [50] ESTIMATION OF THE ORGAN AND EFFECTIVE DOSE TO PATIENTS UNDERGOING MEDICAL DIAGNOSTIC X-RAY EXAMINATIONS IN SAUDI ARABIA
    Ababneh, Zaid Q.
    Almusallam, Moayed
    Alfajri, Ibrahim
    Ababneh, Anas M.
    Aljarrah, Khaled M.
    Tries, Mark A.
    Al Anazi, Nawaf
    RADIATION PROTECTION DOSIMETRY, 2021, 194 (01) : 1 - 8