Technical Note: Patient-weight dependence of the effective dose conversion coefficients for diagnostic x-ray imaging procedures

被引:0
|
作者
Golikov, Vladislav [1 ]
Druzhinina, Polina [1 ]
机构
[1] St Petersburg Res Inst Radiat Hyg, Mira Str 8, St Petersburg 197101, Russia
关键词
conversion coefficients; effective dose; patients; x-ray examination; PEDIATRIC BARIUM MEAL; COMPUTATIONAL PHANTOMS; ORGAN; ADULT; AGE;
D O I
10.1002/mp.14446
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose The aim of this study was to provide a method and reference data for the evaluation of the weight dependence of the conversion coefficients from air collision kerma integrated over beam area to the effective dose for patients undergoing diagnostic x-ray examinations. Method A simple physics-based analytical model was developed in order to calculate the weight dependence of the effective dose conversion coefficients based on existing values of the conversion coefficients between patient effective dose and air collision kerma integrated over beam area calculated by the Monte Carlo method on the base of computational MIRD-5-type stylized models. Results The analytical expressions that make possible taking into account a patient's weight when assessing the effective dose were presented. The following categories of diagnostic x-ray examinations are considered: CT; plain radiography; cardiac angiography; fluoroscopy. The results indicate that the weight dependence of the effective dose conversion coefficients for different categories of diagnostic x-ray examinations has the same form. Correction factor for evaluation of the new value of the effective dose conversion coefficient for a patient with weightW(pa.)when we know its value for a standard patient with weightW(st.)depends on the ratios of their weights (W-st./W-pat.) and fractions of the incident energy which is imparted to a patient. Conclusions The paper presents an approach and necessary data for the estimation of the weight dependence of the effective dose conversion coefficients for patients undergoing different diagnostic x-ray examinations.
引用
收藏
页码:5366 / 5372
页数:7
相关论文
共 50 条
  • [21] COLLECTIVE EFFECTIVE DOSE IN EUROPE FROM X-RAY AND NUCLEAR MEDICINE PROCEDURES
    Bly, R.
    Jahnen, A.
    Jarvinen, H.
    Olerud, H.
    Vassileva, J.
    Vogiatzi, S.
    RADIATION PROTECTION DOSIMETRY, 2015, 165 (1-4) : 129 - 132
  • [22] Patient Effective Dose for Ultra-Low-Dose CT Hip Imaging with Tin Filtration Vs X-Ray
    Chu, R.
    Tsai, A.
    Callahan, M. J.
    Kim, D. S.
    Beaulieu, D. R.
    Chong, S. H.
    Zhang, D.
    MEDICAL PHYSICS, 2024, 51 (10) : 7775 - 7776
  • [23] OBSERVATIONS ON THE DOSE FROM DENTAL X-RAY PROCEDURES WITH A NOTE ON RADIOGRAPHY OF THE NASAL BONES
    ARDRAN, GM
    CROOKS, HE
    BRITISH JOURNAL OF RADIOLOGY, 1959, 32 (381): : 572 - 583
  • [24] Absorbed Dose to the Conceptus and Patient Size in X-Ray Projection Imaging
    Magill, D.
    Huda, W.
    He, W.
    Tipnis, S.
    Mah, E.
    Frey, G.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [25] Evaluation of dose area product vs. patient dose in diagnostic X-ray units
    Kisielewicz, K.
    Truszkiewicz, A.
    Wach, S.
    Wasilewska-Radwanska, M.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2011, 27 (02): : 117 - 120
  • [26] Impact of Very Low X-Ray Dose Levels on Imaging Quality in Electrophysiology Procedures
    Bourier, Felix
    Brost, Alexander
    Deisenhofer, Isabel
    Kolb, Christoph
    Schunkert, Heribert
    CIRCULATION, 2013, 128 (22)
  • [27] Effective dose as a limiting quantity for the evaluation of primary barriers for diagnostic x-ray facilities
    Okunade, AA
    HEALTH PHYSICS, 2005, 89 (05): : S100 - S116
  • [28] Technique and Gender Specific Conversion Coefficients for Estimation of Effective Dose from Kerma Area Product During X-Ray Radiography of Chest
    Dimov, Asen
    Tsanev, Ivan
    Penev, Dimitar
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING 2018, VOL 1, 2019, 68 (01): : 719 - 723
  • [29] Technical Note: Water-equivalent thickness of Superflab bolus material at diagnostic x-ray energies
    Blunt, Kayla F.
    Marshall, Emily L.
    Guo, Chao
    Lu, Zheng Feng
    Reiser, Ingrid
    MEDICAL PHYSICS, 2023, 50 (02) : 1237 - 1241