Quantitative Ultrasound Imaging for Monitoring In Situ High-Intensity Focused Ultrasound Exposure

被引:17
|
作者
Ghoshal, Goutam [1 ,2 ]
Kemmerer, Jeremy P. [2 ]
Karunakaran, Chandra [2 ]
Abuhabsah, Rami [2 ]
Miller, Rita J. [2 ]
Sarwate, Sandhya [2 ]
Oelze, Michael L. [2 ]
机构
[1] Acoust MedSyst Inc, Savoy, IL USA
[2] Univ Illinois, Dept Elect & Comp Engn, Urbana, IL 61801 USA
关键词
quantitative ultrasound imaging; HIFU; effective scatterer diameter; effective scatterer concentration; parametric imaging; NONINVASIVE TEMPERATURE ESTIMATION; TISSUE TEMPERATURE; PROPAGATION SPEED; VIVO; ATTENUATION; BACKSCATTER; DEPENDENCE; FEASIBILITY; THERAPY; PHANTOM;
D O I
10.1177/0161734614524179
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Quantitative ultrasound (QUS) imaging is hypothesized to map temperature elevations induced in tissue with high spatial and temporal resolution. To test this hypothesis, QUS techniques were examined to monitor high-intensity focused ultrasound (HIFU) exposure of tissue. In situ experiments were conducted on mammary adenocarcinoma tumors grown in rats and lesions were formed using a HIFU system. A thermocouple was inserted into the tumor to provide estimates of temperature at one location. Backscattered time-domain waveforms from the tissue during exposure were recorded using a clinical ultrasonic imaging system. Backscatter coefficients were estimated using a reference phantom technique. Two parameters were estimated from the backscatter coefficient (effective scatterer diameter (ESD) and effective acoustic concentration (EAC). The changes in the average parameters in the regions corresponding to the HIFU focus over time were correlated to the temperature readings from the thermocouple. The changes in the EAC parameter were consistently correlated to temperature during both heating and cooling of the tumors. The changes in the ESD did not have a consistent trend with temperature. The mean ESD and EAC before exposure were 120 +/- 16 mm and 32 +/- 3 db/cm(3), respectively, and changed to 144 +/- 9 mm and 51 +/- 7 db/cm(3), respectively, just before the last HIFU pulse was delivered to the tissue. After the tissue cooled down to 37 degrees C, the mean ESD and EAC were 126 +/- 8 mm and 35 +/- 4 db/cm(3), respectively. Peak temperature in the range of 50-60 degrees C was recorded by a thermocouple placed just behind the tumor. These results suggest that QUS techniques have the potential to be used for non-invasive monitoring of HIFU exposure.
引用
收藏
页码:239 / 255
页数:17
相关论文
共 50 条
  • [1] Quantitative Ultrasound for Monitoring High-Intensity Focused Ultrasound Treatment In Vivo
    Ghoshal, Goutam
    Kemmerer, Jeremy P.
    Karunakaran, Chandra
    Miller, Rita J.
    Oelze, Michael L.
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2016, 63 (09) : 1234 - 1242
  • [2] Quantitative MR temperature monitoring of high-intensity focused ultrasound therapy
    Bohris, C
    Schreiber, WG
    Jenne, J
    Simiantonakis, I
    Rastert, R
    Zabel, HJ
    Huber, P
    Bader, R
    Brix, G
    MAGNETIC RESONANCE IMAGING, 1999, 17 (04) : 603 - 610
  • [3] Thermoacoustic Lensing in Ultrasound Imaging of Nonechogenic Tissue During High-intensity Focused Ultrasound Exposure
    Jahns, Matthew
    MacDougall, Dan
    Adamson, Robert B. A.
    ULTRASONIC IMAGING, 2018, 40 (03) : 143 - 157
  • [4] Integration of photoacoustic imaging and high-intensity focused ultrasound
    Cui, Huizhong
    Staley, Jacob
    Yang, Xinmai
    JOURNAL OF BIOMEDICAL OPTICS, 2010, 15 (02)
  • [5] Strategy of High Efficiency and Refined High-intensity Focused Ultrasound and Ultrasound Monitoring Imaging of Thermal Lesion and Cavitation
    Wan, Mingxi
    Zhang, Siyuan
    Lu, Mingzhu
    Hu, Hong
    Jing, Bowen
    Liu, Runna
    Zhong, Hui
    PROCEEDINGS FROM THE 14TH INTERNATIONAL SYMPOSIUM ON THERAPEUTIC ULTRASOUND, 2017, 1821
  • [6] High-intensity focused ultrasound beam path visualization using ultrasound imaging
    Song, Jae Hee
    Chang, Jin Ho
    Yoo, Yang Mo
    JOURNAL OF THE ACOUSTICAL SOCIETY OF KOREA, 2020, 39 (01): : 16 - 23
  • [7] Analysis of tissue displacement induced by high-intensity focused ultrasound exposure for coagulation monitoring
    Yamamoto, Mariko
    Yoshizawa, Shin
    JAPANESE JOURNAL OF APPLIED PHYSICS, 2021, 60 (04)
  • [8] Modeling for Quantitative Analysis of Nakagami Imaging in Accurate Detection and Monitoring of Therapeutic Lesions by High-Intensity Focused Ultrasound
    Han, Meng
    Song, Weidong
    Zhang, Fengshou
    Li, Zhenwei
    ULTRASOUND IN MEDICINE AND BIOLOGY, 2023, 49 (07): : 1575 - 1585
  • [9] Multiple ultrasonic parametric imaging for the detection and monitoring of high-intensity focused ultrasound ablation
    Yang, Guang
    Liu, Jing
    Yang, Beiru
    Guo, Junfeng
    Wu, Chenxiaoyue
    Zhang, Bo
    Zhang, Siyuan
    ULTRASONICS, 2024, 139
  • [10] High-intensity focused ultrasound and cryotherapy
    Murat, F. J.
    CRITICAL REVIEWS IN ONCOLOGY HEMATOLOGY, 2011, 78 : S11 - S11