Quantitative analysis of age-dependent backscatter factor (BSF) in x-ray examinations with a Monte Carlo method using XCAT phantoms

被引:0
|
作者
Park, Su-Jin [1 ]
Kim, Jaehyuk [1 ]
Huh, Yoonsuk [1 ]
Kim, Jung-Ha [1 ]
Kim, Jung-Min [1 ]
Park, Hyesuk [1 ]
Jo, JaeMoon [1 ]
机构
[1] Samsung Elect Co Ltd, Hlth & Med Equipment Business, Suwon, South Korea
关键词
Back scatter factor; BSF; Monte Carlo simulation; XCAT phantom; age-dependent; Quantitative analysis; ESD; pediatric patient;
D O I
10.1117/12.2292857
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The international organizations, including IAEA, have established a diagnostic reference levels (DRLs) for diagnostic x-ray examinations in order to control doses. The DRLs are mostly expressed as an entrance surface dose (ESD) to quantify dose on the patients. However, there are significant uncertainties associated with the measurements of ESD, and most of them are results from back scatter factor (BSF). The accurate BSF could be determined only by Monte Carlo simulations. In addition, as the human body is much more complicate compared with a commonly used ANSI phantom, for realistic simulation the computational anthropomorphic extended cardiac-torso (XCAT) phantom has to be used in order to obtain accurate BSF. However, to our knowledge, a few BSF values have previously been reported in the literature in x-ray examinations considering patients' age with XCAT phantom with Monte Carlo methods. The aim of this study was to quantitatively analyze the BSFs using XCAT phantom with Monte Carlo simulation in various x-ray examinations considering the patient's age. For chest examinations, the BSFs were 1.21, 1.29, and 1.35 for one-year old, seven-year old, and adult patient's protocol, respectively. For adult patient, the BSFs were 1.35, 1.26, and 1.10 for chest, abdomen, and extremity examinations, respectively. In addition, the ESDs in chest examinations were 36.30 mu Gy, 52.69 mu Gy, and 111.80 mu Gy for one-, seven-year old, and adult patient's protocol, respectively. These results demonstrated that accurate BSF should be chosen considering the diagnostic examination in order to estimate the ESD properly.
引用
收藏
页数:6
相关论文
共 50 条
  • [31] Monte Carlo-based multiphysics coupling analysis method for focusing X-ray pulsar telescope
    Li L.
    Deng L.
    Mei Z.
    Lyu Z.
    Liu J.
    Zuo F.
    Li, Liansheng (liliansheng1981@163.com), 1600, Chinese Society of Astronautics (37): : 1249 - 1260
  • [32] EXPERIMENTAL-VERIFICATION OF A MONTE-CARLO METHOD FOR X-RAY MICROFLUORESCENCE ANALYSIS OF SMALL PARTICLES
    LANKOSZ, M
    HOLYNSKA, B
    PELLA, PA
    X-RAY SPECTROMETRY, 1993, 22 (01) : 54 - 57
  • [33] A Monte Carlo (MC) based individual calibration method for in vivo x-ray fluorescence analysis (XRF)
    Hansson, Marie
    Isaksson, Mats
    PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (07): : 2009 - 2019
  • [34] Unfolding X-ray spectra using a flat panel detector. Determination of the accuracy of the method with the Monte Carlo method
    Gallardo, S.
    Rodenas, J.
    Verdu, G.
    RADIATION PHYSICS AND CHEMISTRY, 2019, 155 : 233 - 238
  • [35] A standardless method of quantitative ceramic analysis using X-ray powder diffraction
    Mazumdar, S
    JOURNAL OF APPLIED CRYSTALLOGRAPHY, 1999, 32 : 381 - 386
  • [36] Evaluation of medical exposure and exposure by the public in a typical scenario of examinations using mobile X-ray equipment through the Monte Carlo simulation
    Santos, Felipe A.
    Galeano, Diego C.
    Santos, William S.
    Carvalho Junior, Alberico B.
    RADIATION MEASUREMENTS, 2016, 89 : 68 - 73
  • [37] Relation Between the Characteristic X-Ray Intensity and Incident Electron Energy Using the Monte Carlo Method and Measurements
    Gu, Runqiu
    Cheng, Jianfeng
    Lai, Wanchang
    Liao, Xianli
    Wang, Guangxi
    Zhai, Juan
    Zeng, Chenhao
    Wu, Jinfei
    Sun, Xiaochuan
    NUCLEAR TECHNOLOGY, 2022, 208 (05) : 912 - 921
  • [38] A quantitative evaluation method using processed optical images and analysis of age-dependent changes on nasolabial lines
    Jang, Sue Im
    Kim, Eun Joo
    Park, Hyeokgon
    Kim, Hyun Jung
    Suk, Jang Mi
    Kim, Beom Joon
    Lee, John Hwan
    Lee, Hae Kwang
    SKIN RESEARCH AND TECHNOLOGY, 2015, 21 (02) : 201 - 206
  • [39] Monte Carlo simulation of x-ray fluorescence analysis of gold in kidney using 99mTc radiopharmaceutical
    Mahdavi, Naser
    Shamsaei, Mojtaba
    Shafaei, Mostafa
    Rabiei, Ali
    RADIATION PHYSICS AND CHEMISTRY, 2013, 91 : 40 - 43
  • [40] Analysis of X-Ray Diffraction Imaging of Thick Tissue Samples Using GPU-Accelerated Monte Carlo
    Ferguson, K.
    Greenberg, J.
    MEDICAL PHYSICS, 2022, 49 (06) : E213 - E213