Quantitative analysis of age-dependent backscatter factor (BSF) in x-ray examinations with a Monte Carlo method using XCAT phantoms

被引:0
|
作者
Park, Su-Jin [1 ]
Kim, Jaehyuk [1 ]
Huh, Yoonsuk [1 ]
Kim, Jung-Ha [1 ]
Kim, Jung-Min [1 ]
Park, Hyesuk [1 ]
Jo, JaeMoon [1 ]
机构
[1] Samsung Elect Co Ltd, Hlth & Med Equipment Business, Suwon, South Korea
关键词
Back scatter factor; BSF; Monte Carlo simulation; XCAT phantom; age-dependent; Quantitative analysis; ESD; pediatric patient;
D O I
10.1117/12.2292857
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The international organizations, including IAEA, have established a diagnostic reference levels (DRLs) for diagnostic x-ray examinations in order to control doses. The DRLs are mostly expressed as an entrance surface dose (ESD) to quantify dose on the patients. However, there are significant uncertainties associated with the measurements of ESD, and most of them are results from back scatter factor (BSF). The accurate BSF could be determined only by Monte Carlo simulations. In addition, as the human body is much more complicate compared with a commonly used ANSI phantom, for realistic simulation the computational anthropomorphic extended cardiac-torso (XCAT) phantom has to be used in order to obtain accurate BSF. However, to our knowledge, a few BSF values have previously been reported in the literature in x-ray examinations considering patients' age with XCAT phantom with Monte Carlo methods. The aim of this study was to quantitatively analyze the BSFs using XCAT phantom with Monte Carlo simulation in various x-ray examinations considering the patient's age. For chest examinations, the BSFs were 1.21, 1.29, and 1.35 for one-year old, seven-year old, and adult patient's protocol, respectively. For adult patient, the BSFs were 1.35, 1.26, and 1.10 for chest, abdomen, and extremity examinations, respectively. In addition, the ESDs in chest examinations were 36.30 mu Gy, 52.69 mu Gy, and 111.80 mu Gy for one-, seven-year old, and adult patient's protocol, respectively. These results demonstrated that accurate BSF should be chosen considering the diagnostic examination in order to estimate the ESD properly.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] MONTE CARLO METHOD FOR RADIOLOGICAL X-RAY EXAMINATIONS
    Fulea, D.
    Cosma, C.
    Pop, I. G.
    ROMANIAN JOURNAL OF PHYSICS, 2009, 54 (7-8): : 629 - 639
  • [2] Monte Carlo investigation of backscatter point spread function for X-ray imaging examinations
    Xiong, Zhenyu
    Vijayan, Sarath
    Rudin, Stephen
    Bednarek, Daniel R.
    MEDICAL IMAGING 2017: PHYSICS OF MEDICAL IMAGING, 2017, 10132
  • [3] Quantitative Characteristic X-Ray Analysis for Different Compound Samples by Using Monte Carlo Method
    Tekin, H. O.
    Altunsoy, E. E.
    Manici, T.
    Yilmaz, B.
    ACTA PHYSICA POLONICA A, 2017, 132 (03) : 439 - 441
  • [4] Monte Carlo simulations for X-ray breast dosimetry using homogeneous and heterogeneous phantoms
    Tucciariello, R. M.
    Barca, P.
    Del Sarto, D.
    Lamastra, R.
    Traino, A. C.
    Fantacci, M. E.
    NUOVO CIMENTO C-COLLOQUIA AND COMMUNICATIONS IN PHYSICS, 2021, 44 (4-5):
  • [5] Quantitative x-ray microanalysis of heterogeneous materials using Monte Carlo simulations
    Gauvin, Raynald
    MICROCHIMICA ACTA, 2006, 155 (1-2) : 75 - 81
  • [6] Quantitative X-Ray Microanalysis of Heterogeneous Materials Using Monte Carlo Simulations
    Raynald Gauvin
    Microchimica Acta, 2006, 155 : 75 - 81
  • [7] Monte Carlo analysis of energy dependent anisotropy of bremsstrahlung x-ray spectra
    Kakonyi, Robert
    Erdelyi, Miklos
    Szabo, Gabor
    MEDICAL PHYSICS, 2009, 36 (09) : 3897 - 3905
  • [8] OPTIMIZATION OF QUANTITATIVE X-RAY PHASE-ANALYSIS BY MONTE-CARLO METHODS
    GONZALEZ, CR
    ROQUE, R
    ACTA CRYSTALLOGRAPHICA SECTION A, 1987, 43 : 622 - 626
  • [9] A novel Monte-Carlo based method for quantitative thin film X-ray microanalysis
    Armigliato, A
    Balboni, R
    Frabboni, S
    Rosa, R
    MIKROCHIMICA ACTA, 2000, 132 (2-4) : 213 - 218
  • [10] Monte Carlo methods for the in vivo analysis of cisplatin using X-ray fluorescence
    Hugtenburg, RP
    Turner, JR
    Mannering, DM
    Robinson, BA
    APPLIED RADIATION AND ISOTOPES, 1998, 49 (5-6) : 673 - 676