Alternative Placement of Bispectral Index Electrode for Monitoring Depth of Anesthesia during Neurosurgery

被引:0
|
作者
Akavipat, Phuping [1 ]
Hungsawanich, Nate [1 ]
Jansin, Rungruedee [1 ]
机构
[1] Prasat Neurol Inst, Dept Anesthesiol, Bangkok 10400, Thailand
关键词
Bland and Altman analysis; electroencephalogram; instrumentation; equipment; BIS; SENSOR; PERFORMANCE; POSITION; MONTAGE; EVENTS; VALUES; SITE;
D O I
暂无
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
In neurosurgery in particular, the recommended placement of electrodes for monitoring depth of anesthesia during surgery sometimes conflicts with the surgical site or patient positioning. Therefore, we proposed this study to evaluate the agreement and correlation of bispectral index values recorded from the usual frontal area and the alternate, post-auricular areas in neurosurgery patients. Thirtyfour patients scheduled for neurosurgery under general anesthesia were included. Bispectral index (BIS) sensors were placed at both the frontal and post-auricular areas. The anesthesia given was clinically adjusted according to the frontal (standard) BIS reading. The BIS values and impedance were recorded; Pearson's correlation and Bland-Altman plots were analyzed. The bias +/- 2SD for the electrode placement before, during, and post-anesthesia were 0 +/- 23.32, 1.5 +/- 10.69, and 2.1 +/- 13.52, while the limits of agreement were -23.3 to 23.3, -12.2 to 9.2, and -17.7 to 13.5, respectively. The correlation coefficient between frontal- and post-auricular-area electrodes was 0.74 with a p-value < 0.001. The post-auricular placement of a BIS electrode is a practical alternative to frontal lobe placement. Nevertheless, proper electrode location is important to minimize error.
引用
下载
收藏
页码:151 / 155
页数:5
相关论文
共 50 条
  • [31] MONITORING DEPTH OF ANAESTHESIA USING PRST SCORE AND BISPECTRAL INDEX
    Rahul, R.
    Sowmya, M. J.
    Rangalakshmi, S.
    Kumar, Roshan B. N.
    Karthik, G. S.
    JOURNAL OF EVOLUTION OF MEDICAL AND DENTAL SCIENCES-JEMDS, 2015, 4 (25): : 4282 - 4292
  • [32] Comparison of the Depth of Anesthesia Produced with Dexmedetomidine-Sevoflurane or Medetomidine-Sevoflurane by Using Bispectral Index Monitoring
    Saritas, Zulfikar Kadir
    Korkmaz, Musa
    Saritas, Tuba Berra
    Sivaci, Remziye Gul
    ACTA SCIENTIAE VETERINARIAE, 2014, 42
  • [33] Do we need bispectral index monitoring during total intravenous anesthesia for lumbar discectomies?
    Akcali, Didem Tuba
    Ozkose, Zerrin
    Yardim, Sahin
    TURKISH NEUROSURGERY, 2008, 18 (02) : 125 - 133
  • [34] Alternative sensor position for monitoring of depth of anesthesia
    Farias, C. G. L.
    Nunes, R. R.
    Medeiros, G.
    Fernandes, M. B. C.
    Costa, A. D. S.
    Siqueira, I. D. A.
    EUROPEAN JOURNAL OF ANAESTHESIOLOGY, 2014, 31 : 49 - 49
  • [35] Paradoxical arousal during bispectral index monitoring
    Rampil, IJ
    Cai, HM
    Embree, PB
    Greenwald, SD
    ANESTHESIOLOGY, 2000, 93 (3A) : U130 - U130
  • [36] The impact of EEG-Bispectral Index monitoring on drug usage and recovery during ambulatory anesthesia
    Badrinath, S
    Avramov, MN
    Papaioannou, ME
    Ivankovich, AD
    ANESTHESIA AND ANALGESIA, 1999, 88 (02):
  • [37] Would bispectral index monitoring optimise TCI-propofol anesthesia during spinal surgery?
    Chehab, A
    Sesay, M
    Grenier, B
    Maurette, P
    ANESTHESIOLOGY, 1999, 91 (3A) : U287 - U287
  • [38] Transcranial Doppler ultrasound, bispectral index, and electroencephalographic monitoring of entropy during sevoflurane anesthesia in children
    Abad Gurumeta, A.
    Calvo Vecino, J. M.
    Abad Motos, A.
    Oliden Gutierrez, L.
    Garcia Alvarez, R.
    Nieto Moreno, A.
    REVISTA ESPANOLA DE ANESTESIOLOGIA Y REANIMACION, 2010, 57 (10): : 630 - 638
  • [39] Monitoring the bispectral index during spinal surgery
    Barrera, E.
    Fernandez Galinski, S.
    Arbones, E.
    Pamies, C.
    Escolano, F.
    Puig, M.
    EUROPEAN JOURNAL OF ANAESTHESIOLOGY, 2004, 21 : 39 - 39
  • [40] Bispectral index monitoring of sedation during endoscopy
    Bower, AL
    Ripepi, A
    Dilger, J
    Boparai, N
    Brody, FJ
    Ponsky, JL
    GASTROINTESTINAL ENDOSCOPY, 2000, 52 (02) : 192 - 196