Image quality and dose assessment in digital breast tomosynthesis: A Monte Carlo study

被引:16
|
作者
Baptista, M. [1 ]
Di Maria, S. [1 ]
Oliveira, N. [2 ]
Matela, N. [2 ]
Janeiro, L. [2 ]
Almeida, P. [2 ]
Vaz, P. [1 ]
机构
[1] Univ Lisbon, Inst Super Tecn, Ctr Ciencias & Tecnol Nucl, P-2695066 Bobadela Lrs, Portugal
[2] Univ Lisbon, Fac Ciencias, Inst Biofis & Engn Biomed, P-1749016 Lisbon, Portugal
关键词
Digital breast tomosynthesis; Image quality; Signal difference to noise ratio; Mean glandular dose; Monte Carlo simulations; TO-NOISE RATIO; MAMMOGRAPHY; SIMULATION; OPTIMIZATION; POPULATION; DETECTORS; CONTRAST; SYSTEM;
D O I
10.1016/j.radphyschem.2013.12.036
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
Mammography is considered a standard technique for the early detection of breast cancer. However, its sensitivity is limited essentially due to the issue of the overlapping breast tissue. This limitation can be partially overcome, with a relatively new technique, called digital breast tomosynthesis (DBT). For this technique, optimization of acquisition parameters which maximize image quality, whilst complying with the ALARA principle, continues to be an area of considerable research. The aim of this work was to study the best quantum energies that optimize the image quality with the lowest achievable dose in DBT and compare these results with the digital mammography (DM) ones. Monte Carlo simulations were performed using the state-of-the-art computer program MCNPX 2.7.0 in order to generate several 2D cranio-caudal (CC) projections obtained during an acquisition of a standard DBT examination. Moreover, glandular absorbed doses and photon flux calculations, for each projection image, were performed. A homogeneous breast computational phantom with 50%/50% glandular/adipose tissue composition was used and two compressed breast thicknesses were evaluated: 4 cm and 8 cm. The simulated projection images were afterwards reconstructed with an algebraic reconstruction tool and the signal difference to noise ratio (SDNR) was calculated in order to evaluate the image quality in DBT and DM. Finally, a thorough comparison between the results obtained in terms of SDNR and dose assessment in DBT and DM was performed. (C) 2014 Elsevier Ltd. All rights reserved.
引用
收藏
页码:158 / 162
页数:5
相关论文
共 50 条
  • [41] Effects of Reduced Compression in Digital Breast Tomosynthesis on Pain, Anxiety, and Image Quality
    Suhaimi, Siti Aishah Abdullah
    Mohamed, Afifah
    Ahmad, Mahadir
    Chelliah, Kanaga Kumari
    [J]. MALAYSIAN JOURNAL OF MEDICAL SCIENCES, 2015, 22 (06): : 40 - 46
  • [42] Assessment of absorbed dose in deformed breast tissue by Monte Carlo simulation
    Ashoor, Mansour
    Khorshidi, Abdollah
    [J]. SN APPLIED SCIENCES, 2020, 2 (08):
  • [43] Assessment of absorbed dose in deformed breast tissue by Monte Carlo simulation
    Mansour Ashoor
    Abdollah Khorshidi
    [J]. SN Applied Sciences, 2020, 2
  • [44] In-plane image quality and NPWE detectability index in digital breast tomosynthesis
    Monnin, P.
    Verdun, F. R.
    Bosmans, H.
    Marshall, N. W.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (09):
  • [45] Variation in digital breast tomosynthesis image quality at differing heights above the detector
    Davidson, Rob
    Al Khalifah, Khaled
    Zhou, Abel
    [J]. JOURNAL OF MEDICAL RADIATION SCIENCES, 2022, 69 (02) : 174 - 181
  • [46] Quantitative assessment of microcalcification cluster image quality in digital breast tomosynthesis, 2-dimensional and synthetic mammography
    Andreas E. Petropoulos
    Spyros G. Skiadopoulos
    Anna N. Karahaliou
    Gerasimos A. T. Messaris
    Nikolaos S. Arikidis
    Lena I. Costaridou
    [J]. Medical & Biological Engineering & Computing, 2020, 58 : 187 - 209
  • [47] Quantitative assessment of microcalcification cluster image quality in digital breast tomosynthesis, 2-dimensional and synthetic mammography
    Petropoulos, Andreas E.
    Skiadopoulos, Spyros G.
    Karahaliou, Anna N.
    Messaris, Gerasimos A. T.
    Arikidis, Nikolaos S.
    Costaridou, Lena I.
    [J]. MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2020, 58 (01) : 187 - 209
  • [48] Evaluation of a Generative Adversarial Network to Improve Image Quality and Reduce Radiation-Dose during Digital Breast Tomosynthesis
    Gomi, Tsutomu
    Kijima, Yukie
    Kobayashi, Takayuki
    Koibuchi, Yukio
    [J]. DIAGNOSTICS, 2022, 12 (02)
  • [49] Accuracy of the independent atom approximation in digital tomosynthesis Monte Carlo simulations
    Primidis, Thomas G.
    Soloviev, Vadim Y.
    Welsch, Carsten P.
    [J]. BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2021, 7 (05):
  • [50] A Monte Carlo-based model for simulation of digital chest tomosynthesis
    Ullman, Gustaf
    Dance, David R.
    Sandborg, Michael
    Carlsson, Gudrun Alm
    Svalkvist, Angelica
    Bath, Magnus
    [J]. RADIATION PROTECTION DOSIMETRY, 2010, 139 (1-3) : 159 - 163