Design of a Small Centrifugal Blood Pump With Magnetic Bearings

被引:4
|
作者
Jahanmir, Said [2 ]
Hunsberger, Andrew Z. [1 ]
Ren, Zhaohui [1 ]
Heshmat, Hooshang [1 ]
Heshmat, Crystal [1 ]
Tomaszewski, Michael J. [1 ]
Walton, James F. [1 ]
机构
[1] Mohawk Innovat Technol Inc, Albany, NY 12205 USA
[2] MiTiHeart Corp, Gaithersburg, MD USA
关键词
Centrifugal blood pump; Magnetic bearings; Rotary; Left ventricular assist device; Thrombosis; VENTRICULAR ASSIST DEVICE; HEARTMATE-III; FLOW; SYSTEM; NONPULSATILE; COMPACT;
D O I
10.1111/j.1525-1594.2009.00883.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Design of a blood pump with a magnetically levitated rotor requires rigorous evaluation of the magnetic bearing and motor requirements and analysis of rotor dynamics and hydraulic performance with attention to hemolysis and thrombosis potential. Given the desired geometric dimensions, the required operating speed, flow in both the main and wash flow regions, and magnetic bearing performance, one of several design approaches was selected for a new prototype. Based on the estimated operating speed and clearance between the rotor and stator, the motor characteristics and dimensions were estimated. The motor stiffness values were calculated and used along with the hydraulic loading due to the fluid motion to determine the best design for the axial and radial magnetic bearings. Radial and axial stability of the left ventricular assist device prototype was verified using finite element rotor dynamic analysis. The analysis indicated that the rotor could be completely levitated and spun to the desired operating speed with low power loss and no mechanical contact. In vitro experiments with a mock loop test setup were performed to evaluate the performance of the new blood pump prototype.
引用
收藏
页码:714 / 726
页数:13
相关论文
共 50 条
  • [11] Development of the small caliber centrifugal blood pump
    Miyazoe, Y
    Sawairi, T
    Ito, K
    Yana, J
    ARTIFICIAL ORGANS, 1998, 22 (06) : 461 - 465
  • [12] A self-bearing centrifugal blood pump based on induction motor with active and passive magnetic bearings
    Wu, P
    Tseng, KJ
    PEDS 2003 : FIFTH INTERNATIONAL CONFERENCE ON POWER ELECTRONICS AND DRIVE SYSTEMS, VOLS 1 AND 2, PROCEEDINGS, 2003, : 1642 - 1646
  • [13] Design and Control of Hybrid Magnetic Bearings for Maglev Axial Flow Blood Pump
    Cheng, Shanbao
    Day, Steven W.
    2010 IEEE/ASME INTERNATIONAL CONFERENCE ON ADVANCED INTELLIGENT MECHATRONICS (AIM), 2010,
  • [14] Dynamic characteristics of a centrifugal blood pump with conical spiral groove bearings
    Nakamura, Yoichi
    Tsukamoto, Hiroshi
    International Journal of Emerging Multidisciplinary Fluid Sciences, 2009, 1 (01) : 53 - 60
  • [15] Impeller design for a miniaturized centrifugal blood pump
    Takano, T
    Schulte-Eistrup, S
    Yoshikawa, M
    Nakata, K
    Kawahito, S
    Maeda, T
    Nonaka, K
    Linneweber, J
    Glueck, J
    Fujisawa, A
    Makinouchi, K
    Yokokawa, M
    Nosé, Y
    ARTIFICIAL ORGANS, 2000, 24 (10) : 821 - 825
  • [16] Troubleshooting the ball bearings in a centrifugal pump
    McNally, William
    World Pumps, 2004, (454) : 28 - 29
  • [17] Structure Design and Flow Field Simulation of Magnetic Liquid Suspension Centrifugal Blood Pump
    Wu, Huachun
    Zhang, Zhengyuan
    Chen, Pu
    Ren, Yongwu
    ADVANCED DEVELOPMENT IN AUTOMATION, MATERIALS AND MANUFACTURING, 2014, 624 : 223 - 227
  • [18] Biocompatibility of alumina ceramic and polyethylene as materials for pivot bearings of a centrifugal blood pump
    Takami, Y
    Nakazawa, T
    Makinouchi, K
    Glueck, J
    Nose, Y
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1997, 36 (03): : 381 - 386
  • [19] Design, Manufacturing and Tests of an Implantable Centrifugal Blood Pump
    Bock, Eduardo
    Antunes, Pedro
    Uebelhart, Beatriz
    Leao, Tarcisio
    Fonseca, Jeison
    Cavalheiro, Andre
    Santos Filho, Diolino
    Cardoso, Jose Roberto
    Utiyama, Bruno
    Leme, Juliana
    Silva, Cibele
    Andrade, Aron
    Arruda, Celso
    TECHNOLOGICAL INNOVATION FOR SUSTAINABILITY, 2011, 349 : 410 - +
  • [20] Design considerations of volute geometry of a centrifugal blood pump
    Chan, WK
    Wong, YW
    Hu, W
    ARTIFICIAL ORGANS, 2005, 29 (12) : 937 - 948