Design, Manufacturing and Tests of an Implantable Centrifugal Blood Pump

被引:0
|
作者
Bock, Eduardo [1 ,2 ,3 ]
Antunes, Pedro [1 ,4 ]
Uebelhart, Beatriz [1 ,5 ]
Leao, Tarcisio [3 ]
Fonseca, Jeison [1 ,6 ]
Cavalheiro, Andre [4 ]
Santos Filho, Diolino [4 ]
Cardoso, Jose Roberto [4 ]
Utiyama, Bruno [1 ]
Leme, Juliana [1 ]
Silva, Cibele [1 ]
Andrade, Aron [1 ]
Arruda, Celso [2 ]
机构
[1] IDPC, Institute Dante Pazzanese Cardiol, London, England
[2] Univ Estadual Campinas, Unicamp, Campinas, Brazil
[3] Federal Inst Technol, IFSP, Madras, Tamil Nadu, India
[4] Univ Sao Paulo, Escola Politecn, EPUSP, Sao Paulo, Brazil
[5] Fac Tech, FATEC, Hyderabad, Andhra Pradesh, India
[6] Inst Technol & Aeronaut, Sao Jose Dos Campos, Brazil
关键词
Artificial Organs; Blood Pumps; Ventricular Assist Devices; PIVOT BEARING SYSTEM; FLOW VISUALIZATION; SPIRAL PUMP;
D O I
暂无
中图分类号
TP18 [人工智能理论];
学科分类号
081104 ; 0812 ; 0835 ; 1405 ;
摘要
An implantable centrifugal blood pump was developed for long-term ventricular assistance in cardiac patients. In vitro tests were performed, as wear evaluation, performance tests and hemolysis tests in human blood. Numerical computational simulations were performed during design process in order to predict its best geometry. Wear evaluations helped to select the best materials for double pivot bearing system proposed to achieve longer durability. Performance tests pointed the best impeller geometry. The implantable centrifugal blood pump was compared with other blood pumps founded in literature. The proposed implantable centrifugal blood pump showed the best performance. But, its results showed a strong descendant curve in high flow. Other prototype was manufactured with a different inlet port angle to overcome this problem. The normalized index of hemolysis (NIH) measured 0.0054 mg/100L that can be considered excellent since it is close to the minimum found in literature (between 0.004 g/ 100L e 0.02 g/ 100L). The authors' expectation is that this pump will become a promising Technological Innovation for Sustainability.
引用
收藏
页码:410 / +
页数:3
相关论文
共 50 条
  • [1] THE VIENNA IMPLANTABLE CENTRIFUGAL BLOOD PUMP
    SCHIMA, H
    TRUBEL, W
    WIESELTHALER, G
    SCHMIDT, C
    MULLER, MR
    SIEGL, H
    LOSERT, U
    WOLNER, E
    [J]. ARTIFICIAL ORGANS, 1994, 18 (07) : 500 - 505
  • [2] Performances of a newly developed implantable centrifugal blood pump
    Tsukiya, T.
    Taenaka, Y.
    Tatsumi, E.
    Takano, H.
    Konishi, Y.
    Murai, G.
    Masuzawa, T.
    Okada, Y.
    [J]. ASAIO Journal, 2000, 46 (02)
  • [3] Development of an implantable centrifugal blood pump for circulatory assist
    Wakisaka, Y
    Taenaka, Y
    Chikanari, K
    Okuzono, Y
    Endo, S
    Takano, H
    [J]. ASAIO JOURNAL, 1997, 43 (05) : M608 - M614
  • [4] Performance of a newly developed implantable centrifugal blood pump
    Tsukiya, T
    Taenaka, Y
    Tatsumi, E
    Takano, H
    [J]. ASAIO JOURNAL, 2001, 47 (05) : 559 - 562
  • [5] Establishment of flow estimation for an implantable centrifugal blood pump
    Wakisaka, Y
    Okuzono, Y
    Taenaka, Y
    Chikanari, K
    Masuzawa, T
    Takano, H
    [J]. ASAIO JOURNAL, 1997, 43 (05) : M659 - M662
  • [6] Relationship of blood pressure and pump flow in an implantable centrifugal blood pump during hypertension
    Akimoto, T
    Yamazaki, K
    Litwak, P
    Litwak, KN
    Tagusari, O
    Mori, T
    Antaki, JF
    Kameneva, MV
    Watach, MJ
    Umezu, M
    Tomioka, J
    Kormos, RL
    Koyanagi, H
    Griffith, BP
    [J]. ASAIO JOURNAL, 2000, 46 (05) : 596 - 599
  • [7] Bench Tests of an Implantable Axial Blood Pump
    Konysheva E.G.
    Kudinov V.L.
    Dozorov K.N.
    Kalyanin S.A.
    Kuz'min G.S.
    [J]. Biomedical Engineering, 2011, 44 (06) : 222 - 227
  • [8] Design of a Centrifugal Blood Pump: Heart Turcica Centrifugal
    Demir, Onur
    Biyikli, Emre
    Lazoglu, Ismail
    Kucukaksu, Suha
    [J]. ARTIFICIAL ORGANS, 2011, 35 (07) : 720 - U138
  • [9] An implantable centrifugal blood pump for long term circulatory support
    Yamazaki, K
    Litwak, P
    Kormos, RL
    Mori, T
    Tagusari, O
    Antaki, JF
    Kameneva, M
    Watach, M
    Gordon, L
    Umezu, M
    Tomioka, J
    Koyanagi, H
    Griffith, BP
    [J]. ASAIO JOURNAL, 1997, 43 (05) : M686 - M691
  • [10] Development of a miniaturized implantable centrifugal blood pump-Baylor gyro KP pump
    Takano, T.
    Schulte-Eistrup, S.
    Kawahito, S.
    Maeda, T.
    Nonaka, K.
    Linneweber, J.
    Glueck, J.
    Schima, H.
    Fujisawa, A.
    Nosé, Y.
    [J]. ASAIO Journal, 2000, 46 (02)