Choosing a preamplifier for tuning fork signal detection in scanning force microscopy

被引:16
|
作者
Jahncke, CL [1 ]
Brandt, O
Fellows, KE
Hallen, HD
机构
[1] St Lawrence Univ, Dept Phys, Canton, NY 13617 USA
[2] N Carolina State Univ, Dept Phys, Raleigh, NC 27695 USA
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2004年 / 75卷 / 08期
关键词
D O I
10.1063/1.1777386
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
In scanning probe microscopy, it is critical to maintain small probe sample separations for high resolution imaging. Quartz crystal tuning forks are typically used for detecting shear forces in near-field scanning optical microscopy and normal forces in other atomic force-related microscopies. In this article we compare several tuning fork based detection schemes to determine which solution gives the best signal to noise ratio. The high impedance and low signals produced by the tuning fork necessitate care in selection of an appropriate preamplifier. We find that a carefully guarded voltage preamplifier sensing a mechanically driven tuning fork performs the best, but an electrically driven fork with a current preamplifier offers simpler construction with only 25% lower signal to noise ratio on average. (C) 2004 American Institute of Physics.