Practical implementation of quality improvement for high-dose-rate brachytherapy

被引:8
|
作者
Schubert, Leah [1 ]
Liu, Arthur [1 ]
Gan, Gregory [2 ]
Amini, Arya [1 ]
Hutchison, Rebecca [1 ]
Ernest, Janyce [1 ]
Thornton, Dale [1 ]
Stoehr, Scott [1 ]
Hinman, Bonnie [1 ]
Stuhr, Kelly [1 ]
Westerly, David [1 ]
Schefter, Tracey [1 ]
Fisher, Christine [1 ]
机构
[1] Univ Colorado, Dept Radiat Oncol, Aurora, CO USA
[2] Univ New Mexico, Dept Internal Med, Albuquerque, NM 87131 USA
关键词
TREATMENT DELIVERY; PATIENT SAFETY; CHECKLIST; GUIDELINES; MANAGEMENT; ERRORS;
D O I
10.1016/j.prro.2015.09.003
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: High-dose-rate (HDR) brachytherapy is a high-risk procedure with serious errors reported in the medical literature. Our goal was to develop a quality improvement framework for HDR brachytherapy using a multidisciplinary approach. This work describes the time, personnel, and materials involved in implementation as well as staff-reported safety benefits of quality improvement checklists. Methods and materials: Quality improvement was achieved using a department-wide multidisciplinary approach. Process mapping of the entire HDR program, from initial scheduling through follow-up, was performed. The scope of the project was narrowed to the point of treatment delivery. Two types of multidisciplinary checklists were created: a safety-timeout checklist to ensure safety-critical actions were performed before treatment initiation; and detailed procedure checklists that served as written procedures for physicians, physicists, dosimetrists, and nurses. Implementation was carried out through initial training led by various staff members, creation of visual training guides, piloting and use of checklists for all treatments, and auditing of checklist compliance. Results: Process maps of the entire HDR program were generated and used to guide subsequent changes in the treatment delivery process. A single safety-timeout checklist and the individual procedure checklists were created and used at the time of treatment delivery. The 3-month audit showed that the safety-timeout checklist was used for 100% of treatment fractions. Individual procedure checklists were used for 85% of fractions. All cross-covering physicians and physicists continued to use these checklists 100% of the time. Staff survey results indicated improvements in safety and increased benefits for cross-covering staff. Conclusions: In using a multidisciplinary approach to quality improvement, process mapping and comprehensive checklists for HDR treatment delivery have been implemented. This has resulted in improved practices that are optimal in our department. This experience can provide others with practical strategies toward implementing such changes in their own facilities. (C) 2016 American Society for Radiation Oncology. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:34 / 43
页数:10
相关论文
共 50 条
  • [31] High-dose-rate brachytherapy in uterine cervical carcinoma
    Patel, FD
    Rai, B
    Mallick, I
    Sharma, SC
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 62 (01): : 125 - 130
  • [32] Developing a High-Dose-Rate Prostate Brachytherapy Program
    Waring, Jayne
    Gosselin, Tracy
    CLINICAL JOURNAL OF ONCOLOGY NURSING, 2010, 14 (02) : 199 - 205
  • [33] Biologic treatment planning for high-dose-rate brachytherapy
    Manning, MA
    Zwicker, RD
    Arthur, DW
    Arnfield, M
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 49 (03): : 839 - 845
  • [34] High-dose-rate brachytherapy as monotherapy in prostate cancer
    Rebecca Kelsey
    Nature Reviews Urology, 2015, 12 (7) : 359 - 359
  • [35] Focal-Boost Dose Escalation in High-Dose-Rate Brachytherapy
    Guthier, C.
    Harris, T.
    King, M.
    Buzurovic, I.
    Cormack, R.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [37] A stochastic frontier analysis for enhanced treatment quality of high-dose-rate brachytherapy plans
    Edimo, Paul
    Kroshko, Angelika
    Beaulieu, Luc
    Archambault, Louis
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (06):
  • [38] A dose verification method for high-dose-rate brachytherapy treatment plans
    Kumar, Rajesh
    Sharma, S. D.
    Vijaykumar, C.
    Deshpande, Sudesh
    Sharma, P. K.
    Vandana, S.
    Philomena, A.
    Chilkulwar, Ravi H.
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2008, 4 (04) : 173 - 177
  • [39] High-dose-rate interstitial brachytherapy in oral cancer Its impact on quality of life
    Bajwa, Harjot Kaur
    Singareddy, Rohith
    Alluri, Krishnam Raju
    BRACHYTHERAPY, 2016, 15 (03) : 381 - 386
  • [40] A quality assurance device for measuring afterloader performance and transit dose for nasobiliary high-dose-rate brachytherapy
    Deufel, Christopher L.
    Mullins, John P.
    Zakhary, Mark J.
    BRACHYTHERAPY, 2018, 17 (04) : 726 - 731