Practical implementation of quality improvement for high-dose-rate brachytherapy

被引:8
|
作者
Schubert, Leah [1 ]
Liu, Arthur [1 ]
Gan, Gregory [2 ]
Amini, Arya [1 ]
Hutchison, Rebecca [1 ]
Ernest, Janyce [1 ]
Thornton, Dale [1 ]
Stoehr, Scott [1 ]
Hinman, Bonnie [1 ]
Stuhr, Kelly [1 ]
Westerly, David [1 ]
Schefter, Tracey [1 ]
Fisher, Christine [1 ]
机构
[1] Univ Colorado, Dept Radiat Oncol, Aurora, CO USA
[2] Univ New Mexico, Dept Internal Med, Albuquerque, NM 87131 USA
关键词
TREATMENT DELIVERY; PATIENT SAFETY; CHECKLIST; GUIDELINES; MANAGEMENT; ERRORS;
D O I
10.1016/j.prro.2015.09.003
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: High-dose-rate (HDR) brachytherapy is a high-risk procedure with serious errors reported in the medical literature. Our goal was to develop a quality improvement framework for HDR brachytherapy using a multidisciplinary approach. This work describes the time, personnel, and materials involved in implementation as well as staff-reported safety benefits of quality improvement checklists. Methods and materials: Quality improvement was achieved using a department-wide multidisciplinary approach. Process mapping of the entire HDR program, from initial scheduling through follow-up, was performed. The scope of the project was narrowed to the point of treatment delivery. Two types of multidisciplinary checklists were created: a safety-timeout checklist to ensure safety-critical actions were performed before treatment initiation; and detailed procedure checklists that served as written procedures for physicians, physicists, dosimetrists, and nurses. Implementation was carried out through initial training led by various staff members, creation of visual training guides, piloting and use of checklists for all treatments, and auditing of checklist compliance. Results: Process maps of the entire HDR program were generated and used to guide subsequent changes in the treatment delivery process. A single safety-timeout checklist and the individual procedure checklists were created and used at the time of treatment delivery. The 3-month audit showed that the safety-timeout checklist was used for 100% of treatment fractions. Individual procedure checklists were used for 85% of fractions. All cross-covering physicians and physicists continued to use these checklists 100% of the time. Staff survey results indicated improvements in safety and increased benefits for cross-covering staff. Conclusions: In using a multidisciplinary approach to quality improvement, process mapping and comprehensive checklists for HDR treatment delivery have been implemented. This has resulted in improved practices that are optimal in our department. This experience can provide others with practical strategies toward implementing such changes in their own facilities. (C) 2016 American Society for Radiation Oncology. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:34 / 43
页数:10
相关论文
共 50 条
  • [21] Clinical implementation of a novel applicator in high-dose-rate brachytherapy treatment of esophageal cancer
    Buzurovic, Ivan M.
    Hansen, Jorgen L.
    Bhagwat, Mandar S.
    O'Farrell, Desmond A.
    Friesen, Scott
    Harris, Thomas C.
    Damato, Antonio L.
    Cormack, Robert A.
    Martin, Neil E.
    Devlin, Phillip M.
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2016, 8 (04) : 319 - 325
  • [22] High-dose-rate interstitial brachytherapy for gynecologic malignancies
    Beriwal, Sushil
    Bhatnagar, Ajay
    Heron, Dwight E.
    Selvaraj, Raj
    Mogus, Robert
    Kim, Hayeon
    Gerszten, Kristina
    Kelley, Joseph
    Edwards, Robert P.
    BRACHYTHERAPY, 2006, 5 (04) : 218 - 222
  • [23] High-dose-rate brachytherapy as monotherapy for prostate cancer
    Demanes, D. Jeffrey
    Ghilezan, Michel I.
    BRACHYTHERAPY, 2014, 13 (06) : 529 - 541
  • [24] Postoperative high-dose-rate brachytherapy in the prevention of keloids
    Veen, Ronald E.
    Kal, Henk B.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 69 (04): : 1205 - 1208
  • [25] High-dose-rate brachytherapy for cancer of the biliary tract
    Yokouchi, J.
    Satani, K.
    Kanesaka, N.
    Waki, H.
    Abe, K.
    Takekawa, S.
    EJC SUPPLEMENTS, 2005, 3 (02): : 208 - 208
  • [26] HIGH-DOSE-RATE BRACHYTHERAPY FOR CERVICAL-CARCINOMA
    STITT, JA
    THOMADSEN, BR
    FOWLER, JF
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1992, 24 (03): : 574 - 574
  • [27] Interstitial high-dose-rate brachytherapy in eyelid cancer
    Mareco, Virginia
    Bujor, Laurentiu
    Abrunhosa-Branquinho, Andre N.
    Ferreira, Miguel Reis
    Ribeiro, Tiago
    Vasconcelos, Ana Luisa
    Ferreira, Cidalina Reis
    Jorge, Marilia
    BRACHYTHERAPY, 2015, 14 (04) : 554 - 564
  • [28] Postoperative Management of Keloids with High-Dose-Rate Brachytherapy
    Durante, S.
    Vavassori, A.
    Franzetti, J.
    Rotondi, M.
    Comi, S.
    Cambria, R.
    Cattani, F.
    De Lorenzi, F.
    Rietjens, M.
    Lazzari, R.
    Orecchia, R.
    Fossa, B. A. Jereczek
    RADIOTHERAPY AND ONCOLOGY, 2022, 170 : S1613 - S1613
  • [29] High-dose-rate brachytherapy in lower eyelid cancer
    Martinez-Monge, Rafael
    Gomez-Iturriaga, Alfonso
    BRACHYTHERAPY, 2007, 6 (03) : 227 - 229
  • [30] FRACTIONATED HIGH-DOSE-RATE BRACHYTHERAPY FOR INTRACRANIAL GLIOMAS
    WOO, S
    BUTLER, EB
    GRANT, W
    BERNER, B
    GILDENBERG, P
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1994, 28 (01): : 247 - 249