Determination of hydrogen peroxide by micro-flow injection-chemiluminescence using a coupled flow cell reactor chemiluminometer

被引:0
|
作者
Nozaki, O
Kawamoto, H
机构
[1] Kinki Univ, Sch Med, Dept Clin Pathol, Osaka 5898511, Japan
[2] Tottori Univ, Coll Med Technol, Dept Med Technol, Tottori 6830826, Japan
关键词
flow cell reactor; luminol; micro-flow injection-chemiluminescence; hydrogen peroxide; immobilized horseradish peroxidase;
D O I
10.1002/1522-7243(200005/06)15:3<137::AID-BIO576>3.0.CO;2-J
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A novel flow cell reactor was developed for micro-flow injection determination of hydrogen peroxide (H2O2) using horseradish peroxide (HRP)-catalysed luminol chemiluminescence. The newly developed how cell reactor for a chemiluminometer allowed mixing of the chemiluminescent reagents in front of a photomultiplier for maximum detection of the emitted light. The rapid mixing allowed a decrease in the flow rate of the pump to 0.1-0.01 mL/min, resulting in increased sensitivity of detection of light. The how cell reactor was made by packing HRP-immobilized gels into a flow cell (Teflon tube; 6 cm x 0.98 mm i.d.) located in the cell holder of a chemiluminometer (flow-through type). The HRP-immobilized gels were made by immobilizing HRP onto the Chitopearl gel by the periodate method. H2O2 specimens (50 mu L) were injected into a stream of water delivered at a how rate of 0.1 mL/min and mixed with a luminol solution (0.56 mmol/L in Tricine buffer, pH 9.2) delivered at 0.1 mL/min in the flow cell reactor. Within-run reproducibility of the assay of H2O2 was 2.4% (4.85 mu mol/L; flow rate 0.1 mL/min, injection interval 10 min). The reproducibility of the H2O2 assay was influenced by the flow rates and the injection intervals of the H2O2 specimens. As the flow rates decreased, both the light intensity and the light duration increased. Optimal light intensity was obtained at a luminol concentration of 3-8 mmol/L, but 0.56 mmol/L was sufficient for assay of H2O2 in clinical specimens. At a luminol concentration of 0.56 mmol/L, the regression equation of the standard curve for H2O2 (0-9.7 mu mol/L) was Y = 27.5 X-2 + 394 X + 58.9 (Y = light intensity; X = concentration of H2O2) and the detection limit of H2O2 was 0.2 mu mol/L. This method was used to assay glucose (2.7-16.7 mmol/L) based on a glucose oxidase (20 U/mL, pH 7.4) reaction. The standard curve for glucose was Y = 167 X-2 - 351 X + 1484 (Y = light intensity; X = glucose). The within-run reproducibility for an aqueous glucose standard (2.7 mmol/L) and a control serum (glucose, 5 mmol/L) was 4.48% (n = 5) and 5.70% (n = 9), respectively. Copyright (C) 2000 John Wiley & Sons, Ltd.
引用
收藏
页码:137 / 142
页数:6
相关论文
共 50 条
  • [1] Determination of hydrogen peroxide by flow injection-chemiluminescence using a novel flow-cell reactor in a chemiluminescence detector
    Kawamoto, H
    Nozaki, O
    BUNSEKI KAGAKU, 1999, 48 (05) : 471 - 476
  • [2] Determination of diquat by flow injection-chemiluminescence
    Lopez-Paz, J. L.
    Catala-Icardo, M.
    Anton-Garrido, B.
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2009, 394 (04) : 1073 - 1079
  • [3] Flow injection-chemiluminescence determination of acyclovir
    Long, Xingyu
    Chen, Funan
    LUMINESCENCE, 2012, 27 (06) : 478 - 481
  • [4] Determination of biothiols by flow injection-chemiluminescence method
    Zhang, Chao
    Li, Junkai
    Zhou, Ziqi
    Gong, Yige
    Yun, Keming
    Niu, Weifen
    Chinese Journal of Analysis Laboratory, 2021, 40 (07): : 865 - 868
  • [5] Flow injection-chemiluminescence determination of amidopyrine in the antondin injection
    He, YH
    Nie, F
    Lü, JR
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2001, 29 (03) : 296 - 298
  • [6] Flow injection-chemiluminescence determination of oxytetracycline hydrochloride
    Nie, F
    Wang, ZY
    He, YH
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2000, 28 (12) : 1516 - 1518
  • [7] Flow injection-chemiluminescence determination of strychnine nitrate
    Li, Junkai
    Zhou, Ziqi
    Gong, Yige
    Jia, Juan
    Yun, Keming
    Niu, Weifen
    Chinese Journal of Analysis Laboratory, 2021, 40 (08) : 890 - 892
  • [8] Flow injection-chemiluminescence determination of soybean isoflavone
    Dan, Yang
    Jing, Cang
    Hao Zaibin
    Qiu Lijuan
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2006, 34 (08) : 1113 - 1115
  • [9] Flow injection-chemiluminescence determination of puerarin in pharmaceutical preparations
    Zhang, QL
    Myint, A
    Liu, LJ
    Ge, XW
    Cui, H
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2004, 36 (03) : 587 - 592
  • [10] Flow injection-chemiluminescence determination of sulfadiazine in compound naristillae
    Liu, Haiyan
    Ren, Juanjuan
    Hao, Yuhong
    He, Pingang
    Fang, Yuzhi
    TALANTA, 2007, 72 (03) : 1036 - 1041