A DEVICE TO MEASURE OXYGEN CONSUMPTION DURING THE HYPOTHERMIC PERFUSION OF THE LIVER

被引:0
|
作者
Rodriguez, Joaquin V.
Federico, Maria Belen
Pizarro, Maria Dolores
Guibert, Edgardo E. [2 ]
Quintana, Alejandra B. [1 ]
Scandizzi, Angel L.
机构
[1] Univ Nacl Rosario, Dept Ciencias Fisiol, RA-2000 Rosario, Santa Fe, Argentina
[2] Univ Nacl Rosario, Dept Ciencias Biol, Fac Ciencias Bioquim & Farmaceut, RA-2000 Rosario, Santa Fe, Argentina
关键词
liver; cold storage; hypothermic perfusion; oxygen consumption; ISOLATED HEPATOCYTES; ORGAN PRESERVATION; MACHINE PERFUSION; COLD PRESERVATION; RAT-LIVER; TRANSPLANTATION;
D O I
暂无
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
This work deals with the construction and performance of a device designed to measure the oxygen consumption by the liver during hypothermic perfusion in the rat model. Due to its simple design and the utilization of standard materials, it could serve to determine the role of oxygenation during hypothermic perfusion of the liver. The system consists of a reservoir containing the preservation solution, a peristaltic pump and an internal oxygenator made of silicone tube. A five ports manifold connects the circulation to the liver (inflow), to a hydrostatic manometer and to two sample ports; the liver outflow and temperature sensor or gas calibration. Finally the exit port connects the circulation fluid with an oxygen electrode. The preservation solution is pumped through the liver at a constant pressure (77 +/- 15 mmH(2)O) and a perfusion flow of 0.39 - 0.49 mL.min(-1).g liver(-1). To test the system, two to four hours perfusion experiments were performed, at temperatures of 5 and 10 degrees C. Two preservation solutions were evaluated: Custodiol and Bes-Gluconate-Sucrose. The solubility of oxygen in the preservation solutions was determined, and the oxygen consumption by preserved rat livers was measured.
引用
收藏
页码:335 / 346
页数:12
相关论文
共 50 条
  • [41] Muscle Oxygen Consumption Is Regulated By Oxygen Gradient During Contraction With Hypoxic Perfusion
    Ojino, Minoru
    Takakura, Hisashi
    Jue, Thomas
    Iwase, Satoshi
    Masuda, Kazumi
    JOURNAL OF GENERAL INTERNAL MEDICINE, 2012, 27 : 388 - 388
  • [42] OXYGEN CONSUMPTION OF THE HYPOTHERMIC POTASSIUM ARRESTED HEART
    BERNE, RM
    JONES, RD
    CROSS, FS
    PROCEEDINGS OF THE SOCIETY FOR EXPERIMENTAL BIOLOGY AND MEDICINE, 1958, 99 (01): : 84 - 86
  • [43] OXYGEN CONSUMPTION AND CARDIAC OUTPUT IN THE HYPOTHERMIC DOG
    HEGNAUER, AH
    DAMATO, HE
    AMERICAN JOURNAL OF PHYSIOLOGY, 1954, 178 (01): : 138 - 142
  • [44] OXYGEN CONSUMPTION AND MECHANICAL EFFICIENCY OF THE HYPOTHERMIC HEART
    REISSMANN, KR
    VANCITTERS, RL
    JOURNAL OF APPLIED PHYSIOLOGY, 1956, 9 (03) : 427 - 430
  • [45] The effect of hypothermic perfusion with oxygenated perfluorochemical during in situ extended hepatectomy of canine liver
    Ohya, T
    Ohwada, S
    Kawashima, Y
    JOURNAL OF THE AMERICAN COLLEGE OF SURGEONS, 1996, 182 (03) : 219 - 225
  • [46] Hepatic artery/portal vein flow competition during hypothermic machine perfusion of the liver
    Hillewaert, W.
    Liu, Q.
    Vekemans, K.
    Segers, P.
    Verdonck, P.
    Pirenne, J.
    Brasil, J.
    Monbaliu, D.
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2007, 30 (08): : 710 - 710
  • [47] LIVER ADENINE-NUCLEOTIDE METABOLISM DURING HYPOTHERMIC ANOXIA AND A RECOVERY PERIOD IN PERFUSION
    PONTEGNIEISTACE, S
    LAMBOTTE, L
    JOURNAL OF SURGICAL RESEARCH, 1977, 23 (05) : 339 - 347
  • [48] Clinical assessment of liver metabolism during hypothermic oxygenated machine perfusion using microdialysis
    Patrono, Damiano
    Roggio, Dorotea
    Mazzeo, Anna Teresa
    Catalano, Giorgia
    Mazza, Elena
    Rizza, Giorgia
    Gambella, Alessandro
    Rigo, Federica
    Leone, Nicola
    Elia, Vincenzo
    Dondossola, Daniele
    Lonati, Caterina
    Fanelli, Vito
    Romagnoli, Renato
    ARTIFICIAL ORGANS, 2022, 46 (02) : 281 - 295
  • [49] Extracorporeal Hypothermic Perfusion Device for Intestinal Graft Preservation to Decrease Ischemic Injury During Transportation
    Armando Salim Muñoz-Abraham
    Roger Patrón-Lozano
    Raja R. Narayan
    Sami S. Judeeba
    Abedalrazaq Alkukhun
    Tariq I. Alfadda
    Joseph T. Belter
    David C. Mulligan
    Raffaella Morotti
    Joseph P. Zinter
    John P. Geibel
    Manuel I. Rodríguez-Dávalos
    Journal of Gastrointestinal Surgery, 2016, 20 : 313 - 321
  • [50] Extracorporeal Hypothermic Perfusion Device for Intestinal Graft Preservation to Decrease Ischemic Injury During Transportation
    Munoz-Abraham, Armando Salim
    Patron-Lozano, Roger
    Narayan, Raja R.
    Judeeba, Sami S.
    Alkukhun, Abedalrazaq
    Alfadda, Tariq I.
    Belter, Joseph T.
    Mulligan, David C.
    Morotti, Raffaella
    Zinter, Joseph P.
    Geibel, John P.
    Rodriguez-Davalos, Manuel I.
    JOURNAL OF GASTROINTESTINAL SURGERY, 2016, 20 (02) : 313 - 321