A DEVICE TO MEASURE OXYGEN CONSUMPTION DURING THE HYPOTHERMIC PERFUSION OF THE LIVER

被引:0
|
作者
Rodriguez, Joaquin V.
Federico, Maria Belen
Pizarro, Maria Dolores
Guibert, Edgardo E. [2 ]
Quintana, Alejandra B. [1 ]
Scandizzi, Angel L.
机构
[1] Univ Nacl Rosario, Dept Ciencias Fisiol, RA-2000 Rosario, Santa Fe, Argentina
[2] Univ Nacl Rosario, Dept Ciencias Biol, Fac Ciencias Bioquim & Farmaceut, RA-2000 Rosario, Santa Fe, Argentina
关键词
liver; cold storage; hypothermic perfusion; oxygen consumption; ISOLATED HEPATOCYTES; ORGAN PRESERVATION; MACHINE PERFUSION; COLD PRESERVATION; RAT-LIVER; TRANSPLANTATION;
D O I
暂无
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
This work deals with the construction and performance of a device designed to measure the oxygen consumption by the liver during hypothermic perfusion in the rat model. Due to its simple design and the utilization of standard materials, it could serve to determine the role of oxygenation during hypothermic perfusion of the liver. The system consists of a reservoir containing the preservation solution, a peristaltic pump and an internal oxygenator made of silicone tube. A five ports manifold connects the circulation to the liver (inflow), to a hydrostatic manometer and to two sample ports; the liver outflow and temperature sensor or gas calibration. Finally the exit port connects the circulation fluid with an oxygen electrode. The preservation solution is pumped through the liver at a constant pressure (77 +/- 15 mmH(2)O) and a perfusion flow of 0.39 - 0.49 mL.min(-1).g liver(-1). To test the system, two to four hours perfusion experiments were performed, at temperatures of 5 and 10 degrees C. Two preservation solutions were evaluated: Custodiol and Bes-Gluconate-Sucrose. The solubility of oxygen in the preservation solutions was determined, and the oxygen consumption by preserved rat livers was measured.
引用
收藏
页码:335 / 346
页数:12
相关论文
共 50 条
  • [31] Hypothermic oxygenated perfusion for a steatotic liver graft
    Lai, Quirino
    Ruberto, Franco
    Melandro, Fabio
    Laureiro, Zoe Larghi
    Rossi, Massimo
    Mennini, Gianluca
    HEPATOBILIARY & PANCREATIC DISEASES INTERNATIONAL, 2020, 19 (01) : 88 - 90
  • [32] TISSUE HYPOXIA DURING HYPOTHERMIC PERFUSION
    WOODS, M
    HARKEN, AH
    FEDERATION PROCEEDINGS, 1977, 36 (03) : 417 - 417
  • [33] Hypothermic machine perfusion for liver graft preservation
    Amin, Arpit
    Panayotova, Guergana
    Guarrera, James V.
    CURRENT OPINION IN ORGAN TRANSPLANTATION, 2022, 27 (02) : 98 - 105
  • [34] PROTECTION OF THE BRAIN DURING HYPOTHERMIC PERFUSION
    MOHRI, H
    SADAHIRO, M
    AKIMOTO, H
    HANEDA, K
    TABAYASHI, K
    OHMI, M
    ANNALS OF THORACIC SURGERY, 1993, 56 (06): : 1493 - 1496
  • [35] FIBRINOLYSIS DURING HYPOTHERMIC RENAL PERFUSION
    RISBERG, B
    BLOHME, I
    MICROVASCULAR RESEARCH, 1976, 11 (03) : 424 - 424
  • [36] PANCURONIUM DOES NOT DECREASE OXYGEN-CONSUMPTION DURING HYPOTHERMIC OR NORMOTHERMIC BYPASS
    WERLHOF, V
    SESSLER, DI
    ANESTHESIOLOGY, 1995, 83 (3A) : A911 - A911
  • [37] No synergistic effect of carbon monoxide and oxygen during hypothermic liver preservation
    Koetting, M.
    Dombrowski, F.
    Minor, T.
    BRITISH JOURNAL OF SURGERY, 2010, 97 : S94 - S94
  • [38] Intraoperative Oxygen Consumption During Liver Transplantation
    Shibata, M.
    Matsusaki, T.
    Kaku, R.
    Umeda, Y.
    Yagi, T.
    Morimatsu, H.
    TRANSPLANTATION PROCEEDINGS, 2015, 47 (10) : 2902 - 2906
  • [39] Muscle Oxygen Consumption Is Regulated By Oxygen Gradient During Contraction With Hypoxic Perfusion
    Ojino, Minoru
    Takakura, Hisashi
    Jue, Thomas
    Iwase, Satoshi
    Masuda, Kazumi
    MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 2011, 43 (05): : 388 - 388
  • [40] Liver preservation with HTK:: salutary effect of hypothermic aerobiosis by either gaseous oxygen or machine perfusion
    Minor, T
    Olschewski, P
    Tolba, RH
    Akbar, S
    Kocálková, M
    Dombrowski, F
    CLINICAL TRANSPLANTATION, 2002, 16 (03) : 206 - 211