Effect of early activation of cochlear implant on electrode impedance in pediatric population

被引:15
|
作者
Alhabib, Salman F. [1 ]
Abdelsamad, Yassin [2 ]
Yousef, Medhat [1 ,3 ]
Alzhrani, Farid [1 ]
Hagr, Abdulrahman [1 ]
机构
[1] King Saud Univ, Coll Med, King Abdullah Ear Specialist Ctr KAESC, Riyadh, Saudi Arabia
[2] MED El GmbH, Res Dept, Riyadh, Saudi Arabia
[3] Menoufia Univ Hosp, Fac Med, Audiol Unit, Menoufia, Egypt
关键词
One-day activation; Electrode impedance; Fibrosis; Minimal invasive cochlear implantation technique; SAFETY;
D O I
10.1016/j.ijporl.2020.110543
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
Objectives: To assess the evolution of electrode impedance after the early fitting of audio processors (activation after one-day) and classical fitting (activation after one-month) over an up-to-one year after cochlear implant (CI) surgery. Methods: A retrospective cohort study on Fifty-two CI recipients divided into two groups. The study group included 24 recipients (40 ears) who underwent early fitting, whereas the control group contained 28 recipients (40 ears) who underwent classical fitting. The electrode impedance was recorded during the surgery, switch-on session and at one, three, six, and twelve-months after the surgery. Results: In the study group, electrode impedance values obtained intraoperatively and at switch-on and one, three, six, and twelve-months were 4.89, 3.69, 6.52, 6.24, 6.05, and 5.81 K Omega, respectively, and only the switch-on and one-month values were significantly different (p < 0.0001). In the control group, electrode impedance values obtained intraoperatively and at switch-on and one, three, six, and twelve-months were 4.71, 7.19, 6.40, 6.05, and 5.73 K Omega, respectively. Thus, the electrode impedance value at switch-on was 52.65% (p < 0.001) greater than it intraoperatively. For both groups, the electrode impedance value at twelve-months was significantly higher than the respective intraoperative values (study group: 18.6% higher, P = 0.04; control group: 21.65% higher, P = 0.0001). Conclusion: Electrode impedance was significantly lower in the study group compared to the control group until one month after the surgery. However, the electrode impedance at twelve-months after the CI was similar in both groups.
引用
收藏
页数:6
相关论文
共 50 条
  • [21] High Electrode Impedance Values in Pediatric Cochlear Implant Recipients May Imply Insufficient Auditory and Language Skills Development
    Brkic, Faris F.
    Umihanic, Sekib
    Harcinovic, Alen
    Piric, Lejla
    Brkic, Fuad
    JOURNAL OF CLINICAL MEDICINE, 2020, 9 (02)
  • [22] Electrode discrimination by early-deafened cochlear implant patients
    Busby, PA
    Clark, GM
    AUDIOLOGY, 1996, 35 (01): : 8 - 22
  • [23] Early Outcomes With a Slim, Modiolar Cochlear Implant Electrode Array
    McJunkin, Jonathan L.
    Durakovic, Nedim
    Herzog, Jacques
    Buchman, Craig A.
    OTOLOGY & NEUROTOLOGY, 2018, 39 (01) : E28 - E33
  • [24] Nanostructuring of cochlear implant electrode contacts induces delayed impedance increase in vivo
    Linke, Ines
    Fadeeva, Elena
    Scheper, Verena
    Esser, Karl-Heinz
    Koch, Juergen
    Chichkov, Boris N.
    Lenarz, Thomas
    Paasche, Gerrit
    PHYSICA STATUS SOLIDI A-APPLICATIONS AND MATERIALS SCIENCE, 2015, 212 (06): : 1210 - 1215
  • [25] Effects of cochlear implant surgical technique on post-operative electrode impedance
    Gu, Ping
    Jiang, Yi
    Gao, Xue
    Huang, Shasha
    Yuan, Yongyi
    Wang, Guojian
    Li, Beicheng
    Xi, Xin
    Dai, Pu
    ACTA OTO-LARYNGOLOGICA, 2016, 136 (07) : 677 - 681
  • [26] How Cochlear Implant Position and Perilymph Composition affect Stimulation Electrode Impedance
    Sehlmeyer, Merle
    Biebighaeuser, Julian
    Klinger, Britta
    Bhaysar, Mit B.
    Maier, Hannes
    Zimmermann, Stefan
    2024 IEEE INTERNATIONAL SYMPOSIUM ON MEDICAL MEASUREMENTS AND APPLICATIONS, MEMEA 2024, 2024,
  • [27] Computing Cochlear Implant Electrode Interface Impedance based on Electric Field Imaging
    Choi, Charles T. M.
    Wu, Dong Lin
    TWENTIETH BIENNIAL IEEE CONFERENCE ON ELECTROMAGNETIC FIELD COMPUTATION (IEEE CEFC 2022), 2022,
  • [28] Migration of a cochlear implant ground electrode through the calvarium of a pediatric patient
    Mitchell, Ryan M.
    Horn, David L.
    INTERNATIONAL JOURNAL OF PEDIATRIC OTORHINOLARYNGOLOGY, 2015, 79 (02) : 278 - 280
  • [29] Postoperative Impedance-Based Estimation of Cochlear Implant Electrode Insertion Depth
    Schraivogel, Stephan
    Aebischer, Philipp
    Wagner, Franca
    Weder, Stefan
    Mantokoudis, Georgios
    Caversaccio, Marco
    Wimmer, Wilhelm
    EAR AND HEARING, 2023, 44 (06): : 1379 - 1388
  • [30] The Effect of Cochlear Implant Interval on Spoken Language Skills of Pediatric Bilateral Cochlear Implant Users
    Wenrich, Kaitlyn A.
    Davidson, Lisa S.
    Uchanski, Rosalie M.
    OTOLOGY & NEUROTOLOGY, 2019, 40 (06) : E600 - E605