Disposable screen-printed biosensor for transcutaneous oxygen measurement

被引:4
|
作者
Lam, YZ [1 ]
Atkinson, J [1 ]
机构
[1] Univ Southampton, Sch Engn Sci, Thick Film Unit, Southampton SO17 1BJ, Hants, England
关键词
transcutaneous oxygen; screen-printing; amperometric;
D O I
10.1088/0957-0233/13/12/336
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
A disposable transcutaneous oxygen sensor has been designed and fabricated in-house employing screen-printing methods so as to achieve cost-effective and reliable production. The oxygen sensing module of the device is based on the amperometric Clarke cell principle. A three-electrode configuration consisting of miniature gold working and counter-electrodes and a silver/silver chloride reference electrode is screen-printed onto alumina substrate resulting in approximately 15 mum thickness layers. A platinum heater element is integrated into the design to make transcutaneous measurements, typically at 44 degreesC. The performances of different screen-printed membrane materials have been evaluated in both hydrated and dry test conditions using cyclic voltammographs and static tests where the devices are subjected to different oxygen levels. All measurements are made by a fully automated computer-controlled gas rig.
引用
收藏
页码:2074 / 2081
页数:8
相关论文
共 50 条
  • [1] A disposable reagentless screen-printed amperometric biosensor for the measurement of alcohol in beverages
    Sprules, SD
    Hartley, IC
    Wedge, R
    Hart, JP
    Pittson, R
    [J]. ANALYTICA CHIMICA ACTA, 1996, 329 (03) : 215 - 221
  • [2] A screen-printed disposable biosensor for selective determination of putrescine
    Henao-Escobar, Wilder
    Dominguez-Renedo, Olga
    Asuncion Alonso-Lomillo, Maria
    Julia Arcos-Martinez, Maria
    [J]. MICROCHIMICA ACTA, 2013, 180 (7-8) : 687 - 693
  • [3] A screen-printed disposable biosensor for selective determination of putrescine
    Wilder Henao-Escobar
    Olga Domínguez-Renedo
    María Asunción Alonso-Lomillo
    María Julia Arcos-Martínez
    [J]. Microchimica Acta, 2013, 180 : 687 - 693
  • [4] FABRICATION AND CHARACTERIZATION OF A SCREEN-PRINTED, DISPOSABLE, AMPEROMETRIC CHOLESTEROL BIOSENSOR
    GILMARTIN, MAT
    HART, JP
    [J]. ANALYST, 1994, 119 (11) : 2331 - 2336
  • [5] Studies towards a disposable screen-printed amperometric biosensor for progesterone
    Hart, JP
    Pemberton, RM
    Luxton, R
    Wedge, R
    [J]. BIOSENSORS & BIOELECTRONICS, 1997, 12 (11): : 1113 - 1121
  • [6] Screen-printed transcutaneous oxygen sensor employing polymer electrolytes
    Y. Z. Lam
    J. K. Atkinson
    [J]. Medical and Biological Engineering and Computing, 2003, 41 : 456 - 463
  • [7] Screen-printed transcutaneous oxygen sensor employing polymer electrolytes
    Lam, YZ
    Atkinson, JK
    [J]. MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2003, 41 (04) : 456 - 463
  • [8] Screen-printed biosensor for allergens
    Sarkar, P
    Ghosh, D
    Bhattacharya, D
    Kataky, R
    Setford, SJ
    White, SF
    Turner, APF
    [J]. JOURNAL OF CHEMICAL TECHNOLOGY AND BIOTECHNOLOGY, 2005, 80 (12) : 1389 - 1394
  • [9] The Study of Urea Nitrogen Biosensor Based on the Disposable Screen-Printed Carbon Electrode
    Chen, Jie
    Chen, Yan-Ping
    Chen, Ping
    [J]. INTERNATIONAL JOURNAL OF ELECTROCHEMICAL SCIENCE, 2015, 10 (09): : 7794 - 7802
  • [10] DISPOSABLE SCREEN-PRINTED ELECTRODES FOR MONITORING HYDRAZINES
    WANG, J