Establishment of regional diagnostic reference levels for digital mammography in Western Province of Sri Lanka

被引:4
|
作者
Niroshani, H. S. [1 ,2 ,3 ]
Jeyasugiththan, Jeyasingam [1 ]
Senanayake, G. [4 ]
Negishi, Toru [3 ]
机构
[1] Univ Colombo, Dept Nucl Sci, Colombo, Sri Lanka
[2] Gen Sir John Kotelawala Def Univ, Dept Radiog & Radiotherapy, Werahera, Sri Lanka
[3] Tokyo Metropolitan Univ, Grad Sch Human Hlth Sci, Dept Radiol Sci, Tokyo, Japan
[4] Gen Sir John Kotelawala Def Univ, Fac Med, Dept Clin Sci, Werahera, Sri Lanka
关键词
mammography; mean glandular dose (MGD); BREAST SCREENING-PROGRAM; DOSE SURVEY; MOLYBDENUM; RHODIUM;
D O I
10.1088/1361-6498/abd075
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The radiation dose to the breasts should be kept to a minimum as breast tissues are highly sensitive to radiation. In mammography, the mean glandular dose (MGD) is used to specify the patient dose. In this study, data on the MGD during diagnostic mammographic examinations was collected using the database from six digital mammography facilities available in the Western Province in Sri Lanka. Examinations involving breast pathology, breast implants, or compressed breast thicknesses (CBT) outside the range of 20-110 mm were excluded in this study. The mean MGD per breast was 3.50 mGy, with a mean CBT of 57 mm. The mean MGD per facility varies from 1.58 to 2.27 mGy, with overall 75th and 95th percentiles of 2.15 and 2.82 mGy, respectively. The 75th and 95th percentile MGD per image, for the average CBT of 57 +/- 12 mm, were 2.00 and 2.65 mGy respectively. The 75th percentile value of the MGD is suggested for the Western Province and it depends on the specific CBT.
引用
收藏
页码:79 / 96
页数:18
相关论文
共 50 条
  • [41] Local Diagnostic Reference Levels for Full-Field Digital Mammography and Digital Breast Tomosynthesis in a Tertiary Hospital in Malaysia
    Norsuddin, Norhashimah Mohd
    Segar, Sharveeni
    Ravintaran, Rathieswari
    Zain, Norhayati Mohd
    Karim, Muhammad Khalis Abdul
    [J]. HEALTHCARE, 2022, 10 (10)
  • [42] Reproductive phenology of Calophyllum inophyllum in Yeppoon, Australia and Meegoda Western Province, Sri Lanka
    Hathurusingha S.
    Ashwath N.
    Wijesekara K.
    Midmore D.
    [J]. Journal of Forestry Research, 2011, 22 (4) : 615 - 619
  • [43] Prevalence of Salmonella in marketed Penaeus monodon shrimps in North Western Province, Sri Lanka
    Ubeyratne, J. Kamalika H.
    Kleer, Josef
    Hildebrandt, Goetz
    Fries, Reinhard
    Khattiya, Rutch
    Padungtod, Pawin
    Baumann, Maximilian P. O.
    Zessin, Karl-Hans
    [J]. BERLINER UND MUNCHENER TIERARZTLICHE WOCHENSCHRIFT, 2008, 121 (11-12): : 418 - 421
  • [44] Including all? Perceptions of Mainstream Teachers on Inclusive Education in the Western Province of Sri Lanka
    Hettiaarachi, S.
    Ranaweera, M.
    Walisundara, D.
    Daston-Attanayake, L.
    Das, A. K.
    [J]. INTERNATIONAL JOURNAL OF SPECIAL EDUCATION, 2018, 33 (02) : 427 - 447
  • [45] Proposed diagnostic reference levels for general radiography and mammography in Japan
    Asada, Yasuki
    Kondo, Yuya
    Kobayashi, Masanao
    Kobayashi, Kenichi
    Ichikawa, Takuma
    Matsunaga, Yuta
    [J]. JOURNAL OF RADIOLOGICAL PROTECTION, 2020, 40 (03) : 867 - 876
  • [46] The establishment of computed tomography diagnostic reference levels in Portugal
    Santos, Joana
    Foley, Shane
    Paulo, Graciano
    McEntee, Mark F.
    Rainford, Louise
    [J]. RADIATION PROTECTION DOSIMETRY, 2014, 158 (03) : 307 - 317
  • [47] Establishment of diagnostic reference levels in cardiac computed tomography
    Rawashdeh, Mohammad
    Saade, Charbel
    Zaitoun, Maha
    Abdelrahman, Mostafa
    Brennan, Patrick
    Alewaidat, Haytham
    McEntee, Mark F.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (10): : 181 - 186
  • [48] Establishment of Provincial Diagnostic Reference Levels in Pediatric Imaging
    Tonkopi, E.
    O'Brien, K.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3716 - 3716
  • [49] Proposed procedure for the establishment of diagnostic reference levels in Germany
    Veit, R
    Bauer, B
    Bernhardt, HJ
    Lechel, U
    [J]. RADIATION PROTECTION DOSIMETRY, 1998, 80 (1-3) : 117 - 120
  • [50] Establishment of diagnostic reference levels for dental intraoral radiography
    Manousaridis, G.
    Koukorava, C.
    Hourdakis, C. J.
    Kamenopoulou, V.
    Yakoumakis, E.
    Tsiklakis, K.
    [J]. RADIATION PROTECTION DOSIMETRY, 2013, 156 (04) : 455 - 457