摘要
Measuring the thickness of different corneal microlayers is important for the diagnosis of common corneal eye diseases such as dry eye, keratoconus, Fuchs endothelial dystrophy, and corneal graft rejection. High resolution corneal images, obtained using optical coherence tomography (OCT), made it possible to measure the thickness of different corneal microlayers in vivo. The manual segmentation of these images is subjective and time consuming. Therefore, automatic segmentation is necessary. Several methods were proposed for segmenting corneal OCT images, but none of these methods segment all the microlayer interfaces and they are not robust. In addition, the lack of a large annotated database of corneal OCT images impedes the application of machine learning methods such as deep learning which proves to be very powerful. In this paper, we present a new corneal OCT image segmentation algorithm using Randomized Hough Transform. To the best of our knowledge, we developed the first automatic segmentation method for the six corneal microlayer interfaces. The proposed method includes a robust estimate of relative distances of inner corneal interfaces with respect to outer corneal interfaces. Also, it handles properly the correct ordering and the non-intersection of corneal microlayer interfaces. The proposed method was tested on 15 corneal OCT images that were randomly selected. OCT images were manually segmented by two trained operators for comparison. Comparison with the manual segmentation shows that the proposed method has mean segmentation error of 3.77 +/- 4.25 pixels across all interfaces which corresponds to 5.66 +/- 6.38 mu m. The mean segmentation error between the two manual operators is 4.07 +/- 4.71 pixels, which corresponds to 6.11 +/- 7.07 mu m. The proposed method takes a mean time of 2.59 +/- 0.06 seconds to segment six corneal interfaces.
引用
收藏
页数:11
相关论文
共 50 条
  • [41] Automatic Measurement Method for Corneal Thickness of Optical Coherence Tomography Images
    Gao Y.
    Li Z.
    Zhang J.
    Nan N.
    Wang X.
    Wang X.
    Guangxue Xuebao/Acta Optica Sinica, 2019, 39 (03):
  • [42] Range image segmentation based on randomized Hough transform
    Ding, YH
    Ping, XJ
    Hu, M
    Wang, D
    PATTERN RECOGNITION LETTERS, 2005, 26 (13) : 2033 - 2041
  • [43] Speckle Noise Reduction of Optical Coherence Tomography Images with a Wavelet Transform
    Murakami, Tsuyoshi
    Ogawa, Koichi
    2018 IEEE 14TH INTERNATIONAL COLLOQUIUM ON SIGNAL PROCESSING & ITS APPLICATIONS (CSPA 2018), 2018, : 31 - 34
  • [44] Evaluation of Automated Multiclass Fluid Segmentation in Optical Coherence Tomography Images Using the Pegasus Fluid Segmentation Algorithms
    Terry, Louise
    Trikha, Sameer
    Bhatia, Kanwal K.
    Graham, Mark S.
    Wood, Ashley
    TRANSLATIONAL VISION SCIENCE & TECHNOLOGY, 2021, 10 (01): : 1 - 8
  • [45] Fast detection and segmentation of drusen in retinal optical coherence tomography images
    Farsiu, Sina
    Chiu, Stephanie J.
    Izatt, Joseph A.
    Toth, Cynthia A.
    OPHTHALMIC TECHNOLOGIES XVIII, 2008, 6844
  • [46] Variability of manual ciliary muscle segmentation in optical coherence tomography images
    Chang, Yu-Cherng
    Liu, Keke
    Cabot, Florence
    Yoo, Sonia H.
    Ruggeri, Marco
    Ho, Arthur
    Parel, Jean-Marie
    Manns, Fabrice
    BIOMEDICAL OPTICS EXPRESS, 2018, 9 (02): : 791 - 800
  • [47] A Study on Automated Segmentation of Retinal Layers in Optical Coherence Tomography Images
    Ngo, Lua
    Yih, Geown
    Ji, Seungbae
    Han, Jae-Ho
    2016 4TH INTERNATIONAL WINTER CONFERENCE ON BRAIN-COMPUTER INTERFACE (BCI), 2016,
  • [48] Automatic vessel lumen segmentation in optical coherence tomography (OCT) images
    Zhang, Huaizhong
    Essa, Ehab
    Xie, Xianghua
    APPLIED SOFT COMPUTING, 2020, 88
  • [49] Needle Segmentation in Volumetric Optical Coherence Tomography Images for Ophthalmic Microsurgery
    Zhou, Mingchuan
    Roodaki, Hessam
    Eslami, Abouzar
    Chen, Guang
    Huang, Kai
    Maier, Mathias
    Lohmann, Chris P.
    Knoll, Alois
    Nasseri, Mohammad Ali
    APPLIED SCIENCES-BASEL, 2017, 7 (08):
  • [50] Faster Segmentation Algorithm for Optical Coherence Tomography Images with Guaranteed Smoothness
    Xu, Lei
    Stojkovic, Branislav
    Ding, Hu
    Song, Qi
    Wu, Xiaodong
    Sonka, Milan
    Xu, Jinhui
    MACHINE LEARNING IN MEDICAL IMAGING, 2011, 7009 : 308 - +