A Comparison of the Deflecting-Tip Bronchial Blocker With a Wire-Guided Blocker or Left-Sided Double-Lumen Tube

被引:33
|
作者
Dumans-Nizard, Virginie [1 ]
Liu, Ngai [1 ]
Laloe, Pierre-Antoine [1 ]
Fischler, Marc [1 ]
机构
[1] Hop Foch, Dept Anesthesiol, F-92151 Suresnes, France
关键词
one-lung ventilation; double-lumen endotracheal tube; bronchial blocker; TORQUE CONTROL BLOCKER; ONE-LUNG VENTILATION; ENDOTRACHEAL-TUBE; CARDIAC-SURGERY;
D O I
10.1053/j.jvca.2009.02.002
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Objective: To compare a new bronchial blocker, the Cohen blocker, with the Arndt blocker and a left double-lumen tube (DLT). Design: A prospective, randomized, controlled trial. Setting: University hospital. Participants: Forty-eight patients undergoing lung surgery. Intervention: Intubation with 1 of the 3 devices. Comparisons among groups included (1) time for initial positioning, (2) degree of lung collapse at pleura opening, and (3) number of intraoperative fiberoptic examinations. Measurements and Main Results: Positioning of the Cohen blocker (256 [166-341] seconds; median [interquartile range]) took no longer compared with the Arndt blocker (253 [184-305] seconds), and there was a trend toward difference between the 2 blockers and the DLT (137 [102-199] seconds) (p = 0.07). The time to place the Cohen blocker was longer in cases of left bronchus occlusion compared with a right one (340 [300-450] v 170 [124-259] seconds, p = 0.02); they were similar in the Arndt group. The degree of lung collapse was different among groups (p = 0.05), but the difference between any pair did not reach statistical significance. The number of patients who required at least 1 additional FOB examination was not statistically different (50% of patients in each blocker group v 19% in the DLT group). Conclusions: There was a trend toward a difference between times to place a bronchial blocker and the DLT. The Cohen blocker is more difficult to position in the left main bronchus than in the right one. (C) 2009 Elsevier Inc. All rights reserved
引用
收藏
页码:501 / 505
页数:5
相关论文
共 50 条
  • [21] The relationship between tracheal width and left bronchial width: Implications for left-sided double-lumen tube selection
    Brodsky, JB
    Malott, K
    Angst, M
    Fitzmaurice, BG
    Kee, SP
    Logan, L
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2001, 15 (02) : 216 - 217
  • [22] Use of the modified left-sided double-lumen tube in thoracic surgeries
    Ge, Jianling
    Zhou, Guoxia
    Peng, Lichao
    INTERNATIONAL JOURNAL OF CLINICAL AND EXPERIMENTAL MEDICINE, 2017, 10 (03): : 5222 - 5227
  • [23] A randomised comparison of the efficacy of a Coopdech bronchial blocker and a double-lumen endotracheal tube for minimally invasive esophagectomy
    Zhang, Tian-Hua
    Liu, Xiao-Qing
    Cao, Long-Hui
    Fu, Jian-Hua
    Lin, Wen-Qian
    TRANSLATIONAL CANCER RESEARCH, 2020, 9 (08) : 4686 - 4692
  • [24] Mainstem bronchial diameterfor the left-sided Broncho-Cath® double-lumen tube:: an in vitro study
    Bahk, JH
    Ryu, HG
    Park, JH
    CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 2005, 52 (03): : 341 - 342
  • [25] Is there a better right-sided tube for one-lung ventilation? A comparison of the right-sided double-lumen tube with the single-lumen tube with right-sided enclosed bronchial blocker
    Campos, JH
    Massa, FC
    ANESTHESIA AND ANALGESIA, 1998, 86 (04): : 696 - 700
  • [26] The use of a bronchial blocker to rescue an ill-fitting double-lumen endotracheal tube
    Nino, M
    Body, SC
    Hartigan, PM
    ANESTHESIA AND ANALGESIA, 2000, 91 (06): : 1370 - 1371
  • [27] Fluoroscopy, a useful alternative for confirming accurate placement of double-lumen tube or bronchial blocker
    Date, Aparna
    Wagh, Harshal D.
    Shah, Milin
    SAUDI JOURNAL OF ANAESTHESIA, 2021, 15 (04) : 473 - 474
  • [28] Unplanned impossible tracheal intubation with double-lumen tube: a case of bronchial blocker creation
    Righetti, Roberto
    Zani, Gianluca
    Diamanti, Marina
    Scaramuzza, Aliza
    Papetti, Matilde
    Graziani, Elia
    Terenzoni, Massimo
    Baccarini, Federico Domenico
    Fusari, Maurizio
    EUROPEAN RESPIRATORY JOURNAL, 2017, 50
  • [29] Double lumen tube and bronchial blocker in thoracic surgery
    Palau Marti, Cristian
    Rubio-Haro, Ruben
    Granell Gil, Manuel
    de Andres Ibanez, Jose
    ANESTHESIA AND ANALGESIA, 2021, 133 (3S_SUPPL): : 340 - 341
  • [30] Predicting Optimal Insertion Depth of a Left-sided Double-Lumen Endobronchial Tube
    Lin, Wei-Lin
    Cherng, Chen-Hwan
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2016, 30 (04) : 942 - 946