Monte Carlo simulation of the neutron spectral fluence and dose equivalent for use in shielding a proton therapy vault

被引:16
|
作者
Zheng, Yuanshui [1 ]
Newhauser, Wayne [2 ]
Klein, Eric [1 ]
Low, Daniel [1 ]
机构
[1] Washington Univ, Sch Med, Dept Radiat Oncol, St Louis, MO 63110 USA
[2] Univ Texas MD Anderson Canc Ctr, Dept Radiat Phys, Unit 94, Houston, TX 77030 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2009年 / 54卷 / 22期
关键词
INTRANUCLEAR-CASCADE CALCULATION; STRAY RADIATION; SECONDARY NEUTRON; RADIOTHERAPY; EXPOSURES; FACILITY; DESIGN; CANCER; FETUS;
D O I
10.1088/0031-9155/54/22/013
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Neutron production is of principal concern when designing proton therapy vault shielding. Conventionally, neutron calculations are based on analytical methods, which do not accurately consider beam shaping components and nozzle shielding. The goal of this study was to calculate, using Monte Carlo modeling, the neutron spectral fluence and neutron dose equivalent generated by a realistic proton therapy nozzle and evaluate how these data could be used in shielding calculations. We modeled a contemporary passive scattering proton therapy nozzle in detail with the MCNPX simulation code. The neutron spectral fluence and dose equivalent at various locations in the treatment room were calculated and compared to those obtained from a thick iron target bombarded by parallel proton beams, the simplified geometry on which analytical methods are based. The neutron spectral fluence distributions were similar for both methods, with deeply penetrating high-energy neutrons (E > 10 MeV) being most prevalent along the beam central axis, and low-energy neutrons predominating the neutron spectral fluence in the lateral region. However, unlike the inverse square falloff used in conventional analytical methods, this study shows that the neutron dose equivalent per therapeutic dose in the treatment room decreased with distance approximately following a power law, with an exponent of about -1.63 in the lateral region and -1.73 in the downstream region. Based on the simulated data according to the detailed nozzle modeling, we developed an empirical equation to estimate the neutron dose equivalent at any location and distance in the treatment vault, e. g. for cases in which detailed Monte Carlo modeling is not feasible. We applied the simulated neutron spectral fluence and dose equivalent to a shielding calculation as an example.
引用
收藏
页码:6943 / 6957
页数:15
相关论文
共 50 条
  • [1] Monte Carlo simulations of neutron spectral fluence, radiation weighting factor and ambient dose equivalent for a passively scattered proton therapy unit
    Zheng, Yuanshui
    Fontenot, Jonas
    Taddei, Phil
    Mirkovic, Dragan
    Newhauser, Wayne
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (01): : 187 - 201
  • [2] Monte Carlo Study of Neutron Dose Equivalent for a Compact Proton Therapy Unit
    Zheng, Y.
    Klein, E.
    Chen, K.
    Liu, Y.
    [J]. MEDICAL PHYSICS, 2010, 37 (06)
  • [3] Monte Carlo simulation of neutron dose equivalent by photoneutron production inside the primary barriers of a radiotherapy vault
    Choi, Chang Heon
    Park, So-Yeon
    Park, Jong Min
    Chun, Minsoo
    Kim, Jung-in
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2018, 48 : 1 - 5
  • [4] Monte Carlo study of neutron dose equivalent during passive scattering proton therapy
    Zheng, Yuanshui
    Newhauser, Wayne
    Fontenot, Jonas
    Taddei, Phil
    Mohan, Radhe
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (15): : 4481 - 4496
  • [5] Monte Carlo Simulations of Neutron Ambient Dose Equivalent in a Novel Proton Therapy Facility Design
    Titt, Uwe
    Pera, Enzo
    Gillin, Michael T.
    [J]. INTERNATIONAL JOURNAL OF PARTICLE THERAPY, 2020, 6 (04) : 29 - 37
  • [6] Monte Carlo simulation of secondary neutron dose for scanning proton therapy using FLUKA
    Lee, Chaeyeong
    Lee, Sangmin
    Lee, Seung-Jae
    Song, Hankyeol
    Kim, Dae-Hyun
    Cho, Sungkoo
    Jo, Kwanghyun
    Han, Youngyih
    Chung, Yong Hyun
    Kim, Jin Sung
    [J]. PLOS ONE, 2017, 12 (10):
  • [7] Dose equivalent consideration from neutron contamination in modified radiotherapy vault: a Monte Carlo study
    Saadatmand, Pegah
    Mahdavi, Seied Rabi
    Chegeni, Nahid
    Karimi, Amir Hossein
    [J]. BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2024, 10 (04):
  • [8] Neutron Shielding Assessment for a Compact Proton Therapy Vault
    Prusator, M.
    Ahmad, S.
    Chen, Y.
    [J]. MEDICAL PHYSICS, 2014, 41 (06) : 113 - +
  • [9] MONTE CARLO SIMULATION FOR EVALUATION OF DOSE DISTRIBUTION IN PROTON THERAPY
    Sardari, Dariush
    Salimi, Ehsan
    [J]. ROMANIAN REPORTS IN PHYSICS, 2014, 66 (01) : 148 - 156
  • [10] Systematic Analysis of the Neutron Equivalent Dose in a Mevion S250 Proton Therapy Machine Using Geant4 Monte Carlo Simulation
    Tatari, N.
    Baradaran-Ghahfarokhi, M.
    Roeder, B.
    Zhao, T.
    Khan, R.
    Sobotka, L.
    Darafsheh, A.
    [J]. MEDICAL PHYSICS, 2021, 48 (06)