共 50 条
Comparison BIPM.RI(I)-K6 of the standards for absorbed dose to water of the KRISS, Korea and the BIPM in accelerator photon beams
被引:1
|作者:
Kessler, C.
[1
]
Burns, D. T.
[1
]
Picard, S.
[1
]
Roger, P.
[1
]
Kim, I. J.
[2
]
Kim, B. C.
[2
]
Kim, Y. H.
[2
]
Kim, J. H.
[2
]
Chung, J. P.
[2
]
Yi, C. Y.
[2
]
机构:
[1] Bur Int Poids & Mesures, Pavillon Breteuil, F-92312 Sevres, France
[2] Korea Res Inst Stand & Sci, Daejeon 34113, South Korea
来源:
关键词:
Photoelectrons - Particle beams;
D O I:
10.1088/0026-1394/57/1A/06001
中图分类号:
TH7 [仪器、仪表];
学科分类号:
0804 ;
080401 ;
081102 ;
摘要:
A key comparison has been made between the absorbed dose to water standards of the KRISS, Korea and the BIPM in accelerator photon beams. The results show the standards to be in agreement within the expanded uncertainty (k = 2) of the comparison of 12 parts in 103. The results are analysed and presented in terms of degrees of equivalence, suitable for entry in the BIPM key comparison database. Main text To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/. The final report has been peer-reviewed and approved for publication by the CCRI, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).