Comparison BIPM.RI(I)-K6 of the standards for absorbed dose to water of the KRISS, Korea and the BIPM in accelerator photon beams

被引:1
|
作者
Kessler, C. [1 ]
Burns, D. T. [1 ]
Picard, S. [1 ]
Roger, P. [1 ]
Kim, I. J. [2 ]
Kim, B. C. [2 ]
Kim, Y. H. [2 ]
Kim, J. H. [2 ]
Chung, J. P. [2 ]
Yi, C. Y. [2 ]
机构
[1] Bur Int Poids & Mesures, Pavillon Breteuil, F-92312 Sevres, France
[2] Korea Res Inst Stand & Sci, Daejeon 34113, South Korea
关键词
Photoelectrons - Particle beams;
D O I
10.1088/0026-1394/57/1A/06001
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A key comparison has been made between the absorbed dose to water standards of the KRISS, Korea and the BIPM in accelerator photon beams. The results show the standards to be in agreement within the expanded uncertainty (k = 2) of the comparison of 12 parts in 103. The results are analysed and presented in terms of degrees of equivalence, suitable for entry in the BIPM key comparison database. Main text To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/. The final report has been peer-reviewed and approved for publication by the CCRI, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).
引用
收藏
页数:2
相关论文
共 50 条
  • [21] Key comparison BIPM.RI(I)-K4 of the absorbed dose to water standards of the GUM, Poland and the BIPM in 60Co gamma radiation
    Kessler, C.
    Burns, D.
    Knyziak, A.
    Szymko, M.
    Derlacinski, M.
    [J]. METROLOGIA, 2021, 58 (1A)
  • [22] Key comparison BIPM.RI(I)-K4 of the absorbed dose to water standards of the PTB, Germany and the BIPM in 60Co gamma radiation
    Kessler, C.
    Burns, D. T.
    Kapsch, R. -P.
    Krauss, A.
    [J]. METROLOGIA, 2016, 53
  • [23] Key comparison BIPM.RI(I)-K4 of the absorbed dose to water standards of the NIM, China and the BIPM in 60Co gamma radiation
    Kessler, C.
    Burns, D.
    Jin, J.
    Zhang, X.
    Wang, Z.P.
    Liu, F.B.
    Huang, J.
    Qi, Y.P.
    Yang, X.Y.
    Wang, K.
    [J]. Metrologia, 2022, 59 (1 A)
  • [24] Key comparison BIPM.RI(I)-K4 of the absorbed dose to water standards of the BFKH, Hungary, and the BIPM in 60Co gamma radiation
    Kessler, C.
    Burns, D.
    Finta, V.
    [J]. Metrologia, 2023, 60 (1 A)
  • [26] Key comparison BIPM.RI(I)-K4 of the absorbed dose to water standards of the METAS, Switzerland and the BIPM in 60Co gamma radiation
    Kessler, C.
    Burns, D. T.
    Voros, S.
    Hofstetter-Boillat, B.
    [J]. METROLOGIA, 2015, 52
  • [27] Key comparison BIPM.RI(I)-K4 of the absorbed dose to water standards of the BEV, Austria and the BIPM in 60Co gamma radiation
    Kessler, C.
    Burns, D.
    Steurer, A.
    Tiefenboeck, W.
    Merza, V.
    [J]. METROLOGIA, 2021, 58 (1A)
  • [28] Key comparison BIPM.RI(I)-K7 of the air-kerma standards of the KRISS, Republic of Korea, and the BIPM in mammography X-rays
    Kessler, C.
    Burns, D.
    Roger, P.
    Yi, Chul-Young
    Kim, Yun Ho
    [J]. Metrologia, 2022, 59 (1 A)
  • [29] Key comparison BIPM.RI(I)-K1 of the air-kerma standards of the KRISS, Republic of Korea, and the BIPM in 60Co gamma radiation
    Kessler, C.
    Burns, D.
    Yi, Chul-Young
    Kim, Yun Ho
    Seong, Young Min
    [J]. Metrologia, 2022, 59 (1 A)
  • [30] Key comparison BIPM.RI(I)-K9 of the absorbed dose to water standards of the PTB, Germany and the BIPM in medium-energy x-rays
    Burns, D. T.
    Kessler, C.
    Bueermann, L.
    Ketelhut, S.
    [J]. METROLOGIA, 2018, 55