Real-time GPU implementation of a weighted filtered back-projection algorithm for stationary gantry CT reconstruction

被引:0
|
作者
Thompson, William [1 ]
Morton, Edward [1 ]
Katsevich, Alexander [2 ,3 ]
Yoon, Seongjin [2 ]
Frenkel, Michael [2 ]
机构
[1] Rapiscan Syst, 2805 Columbia St, Torrance, CA 90503 USA
[2] ITomography Corp, Texas Med Ctr Innovat Inst, 2450 Holcombe Blvd, Houston, TX 77021 USA
[3] Univ Cent Florida, Dept Math, Orlando, FL 32816 USA
关键词
image reconstruction; weight function; real-time; GPU;
D O I
10.1117/12.2534855
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We present details of a real-time implementation of a new algorithm designed to reduce streak artifacts in switched-source stationary gantry CT reconstruction. The algorithm is of the filtered back-projection type, and uses a voxel-specific weighting function to account for the non-uniform distribution of illumination angles caused by such a scanning geometry. The main challenge in developing a real-time implementation is the storage and memory bandwidth requirements imposed by the weighting function. This has been addressed by storing weights at a low precision and reduced resolution, and using interpolation to recover weights at the full resolution. Results demonstrate real-time performance of the algorithm at a realistic problem size, running on a low-cost consumer grade laptop.
引用
收藏
页数:5
相关论文
共 50 条
  • [31] Comparison of the image qualities of filtered back-projection, adaptive statistical iterative reconstruction, and model-based iterative reconstruction for CT venography at 80 kVp
    Jin Hyeok Kim
    Ki Seok Choo
    Tae Yong Moon
    Jun Woo Lee
    Ung Bae Jeon
    Tae Un Kim
    Jae Yeon Hwang
    Myeong-Ja Yun
    Dong Wook Jeong
    Soo Jin Lim
    [J]. European Radiology, 2016, 26 : 2055 - 2063
  • [32] Comparison of Iterative Model Reconstruction versus Filtered Back-Projection in Pediatric Emergency Head CT: Dose, Image Quality, and Image-Reconstruction Times
    Southard, R. N.
    Bardo, D. M. E.
    Temkit, M. H.
    Thorkelson, M. A.
    Augustyn, R. A.
    Martinot, C. A.
    [J]. AMERICAN JOURNAL OF NEURORADIOLOGY, 2019, 40 (05) : 866 - 871
  • [33] Comparison of the image qualities of filtered back-projection, adaptive statistical iterative reconstruction, and model-based iterative reconstruction for CT venography at 80 kVp
    Kim, Jin Hyeok
    Choo, Ki Seok
    Moon, Tae Yong
    Lee, Jun Woo
    Jeon, Ung Bae
    Kim, Tae Un
    Hwang, Jae Yeon
    Yun, Myeong-Ja
    Jeong, Dong Wook
    Lim, Soo Jin
    [J]. EUROPEAN RADIOLOGY, 2016, 26 (07) : 2055 - 2063
  • [34] Comparison of a Deep Learning-Based Reconstruction Algorithm with Filtered Back Projection and Iterative Reconstruction Algorithms for Pediatric Abdominopelvic CT
    Son, Wookon
    Kim, MinWoo
    Hwang, Jae-Yeon
    Kim, Yong-Woo
    Park, Chankue
    Choo, Ki Seok
    Kim, Tae Un
    Jang, Joo Yeon
    [J]. KOREAN JOURNAL OF RADIOLOGY, 2022, 23 (07) : 752 - 762
  • [35] A real-time back projection imaging algorithm for impulse surface penetrating radar
    Lei, WT
    Huang, CL
    Su, Y
    [J]. IGARSS 2005: IEEE INTERNATIONAL GEOSCIENCE AND REMOTE SENSING SYMPOSIUM, VOLS 1-8, PROCEEDINGS, 2005, : 1734 - 1737
  • [36] ?Image quality evaluation of the Precise image CT deep learning reconstruction algorithm compared to Filtered Back-projection and iDose4: a phantom study at different dose levels?
    Barca, Patrizio
    Domenichelli, Sara
    Golfieri, Rita
    Pierotti, Luisa
    Spagnoli, Lorenzo
    Tomasi, Silvia
    Strigari, Lidia
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2023, 106
  • [37] Comparison of Image Noise in Dual Energy CT Images Reconstructed Using Filtered Back-Projection, Hybrid Iterative Reconstruction, and Deep Learning Methods
    Prakash, P.
    Nett, B.
    Tang, J.
    [J]. MEDICAL PHYSICS, 2020, 47 (06) : E521 - E521
  • [38] GPU Implementation of List-mode DRAMA for Real-time OpenPET Image Reconstruction
    Kinouchi, Shoko
    Yamaya, Taiga
    Yoshida, Eiji
    Tashima, Hideaki
    Kudo, Hiroyuki
    Suga, Mikio
    [J]. 2010 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD (NSS/MIC), 2010, : 2273 - 2276
  • [39] Real-Time Back-Projection of Fetal ECG Sources in OL-JADE for the Optimization of Blind Electrodes Positioning
    Pani, Danilo
    Argiolas, Stefania
    Raffo, Luigi
    [J]. COMPUTING IN CARDIOLOGY 2010, VOL 37, 2010, 37 : 289 - 292
  • [40] A Real-Time, GPU-Based Implementation of Aperture Domain Model Image REconstruction
    Khan, Christopher
    Dei, Kazuyuki
    Schlunk, Siegfried
    Ozgun, Kathryn
    Byram, Brett
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2021, 68 (06) : 2101 - 2116