Automatic Masking for Robust 3D-2D Image Registration in Image-Guided Spine Surgery

被引:5
|
作者
Ketcha, M. D. [1 ]
De Silva, T. [1 ]
Uneri, A. [2 ]
Kleinszig, G. [4 ]
Vogt, S. [4 ]
Wolinsky, J. -P. [3 ]
Siewerdsen, J. H. [1 ,2 ]
机构
[1] Johns Hopkins Univ, Dept Biomed Engn, Baltimore, MD 21218 USA
[2] Johns Hopkins Univ, Dept Comp Sci, Baltimore, MD 21218 USA
[3] Johns Hopkins Univ, Dept Neurosurg, Baltimore, MD USA
[4] Siemens Healthcare XP Div, Erlangen, Germany
关键词
D O I
10.1117/12.2216913
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
During spinal neurosurgery, patient-specific information, planning, and annotation such as vertebral labels can be mapped from preoperative 3D CT to intraoperative 2D radiographs via image-based 3D-2D registration. Such registration has been shown to provide a potentially valuable means of decision support in target localization as well as quality assurance of the surgical product. However, robust registration can be challenged by mismatch in image content between the preoperative CT and intraoperative radiographs, arising, for example, from anatomical deformation or the presence of surgical tools within the radiograph. In this work, we develop and evaluate methods for automatically mitigating the effect of content mismatch by leveraging the surgical planning data to assign greater weight to anatomical regions known to be reliable for registration and vital to the surgical task while removing problematic regions that are highly deformable or often occluded by surgical tools. We investigated two approaches to assigning variable weight (i.e., "masking") to image content and/or the similarity metric: (1) masking the preoperative 3D CT ("volumetric masking"); and (2) masking within the 2D similarity metric calculation ("projection masking"). The accuracy of registration was evaluated in terms of projection distance error (PDE) in 61 cases selected from an IRB-approved clinical study. The best performing of the masking techniques was found to reduce the rate of gross failure (PDE > 20 mm) from 11.48% to 5.57% in this challenging retrospective data set. These approaches provided robustness to content mismatch and eliminated distinct failure modes of registration. Such improvement was gained without additional workflow and has motivated incorporation of the masking methods within a system under development for prospective clinical studies.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Multi-stage 3D-2D registration for correction of anatomical deformation in image-guided spine surgery
    Ketcha, M. D.
    De Silva, T.
    Uneri, A.
    Jacobson, M. W.
    Goerres, J.
    Kleinszig, G.
    Vogt, S.
    Wolinsky, J-P
    Siewerdsen, J. H.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (11): : 4604 - 4622
  • [2] 3D-2D Image Registration for Image-guided Surgery by Combining Normalized Cross Correlation and Laplacian Correlation
    Wang, Yang
    Xiao, Ye
    Xu, Congcong
    Yang, Bo
    Zheng, Wenfeng
    Liu, Shan
    [J]. BASIC & CLINICAL PHARMACOLOGY & TOXICOLOGY, 2020, 127 : 11 - 12
  • [3] 3D-2D image registration in the presence of soft-tissue deformation in image-guided transbronchial interventions
    Vijayan, R.
    Sheth, N.
    Mekki, L.
    Lu, A.
    Uneri, A.
    Sisniega, A.
    Magaraggia, J.
    Kleinszig, G.
    Vogt, S.
    Thiboutot, J.
    Lee, H.
    Yarmus, L.
    Siewerdsen, J. H.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2023, 68 (01):
  • [4] Assessment of spline-based 2D-3D registration for image-guided spine surgery
    Zhang, Xuan
    Zheng, Guoyan
    Langlotz, Frank
    Nolte, Lutz-Peter
    [J]. MINIMALLY INVASIVE THERAPY & ALLIED TECHNOLOGIES, 2006, 15 (03) : 193 - 199
  • [5] Multi-Body 3D-2D Registration for Image-Guided Reduction of Pelvic Dislocation in Orthopaedic Trauma Surgery
    Han, R.
    Uneri, A.
    De Silva, T.
    Ketcha, M.
    Sheth, N.
    Kleinszig, G.
    Vogt, S.
    Osgood, G.
    Siewerdsen, J.
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E298 - E298
  • [6] 3D-2D registration in endovascular image-guided surgery: evaluation of state-of-the-art methods on cerebral angiograms
    Mitrovic, Uros
    Likar, Bostjan
    Pernus, Franjo
    Spiclin, Ziga
    [J]. INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2018, 13 (02) : 193 - 202
  • [7] Simultaneous 3D-2D image registration and C-arm calibration: Application to endovascular image-guided interventions
    Mitrovic, Uros
    Pernus, Franjo
    Likar, Bostjan
    Spiclin, Ziga
    [J]. MEDICAL PHYSICS, 2015, 42 (11) : 6433 - 6447
  • [8] Multi-body 3D-2D registration for image-guided reduction of pelvic dislocation in orthopaedic trauma surgery
    Han, R.
    Uneri, A.
    Ketcha, M.
    Vijayan, R.
    Sheth, N.
    Wu, P.
    Vagdargi, P.
    Vogt, S.
    Kleinszig, G.
    Osgood, G. M.
    Siewerdsen, J. H.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (13):
  • [9] A fast, accurate, and automatic 2D-3D image registration for image-guided cranial radiosurgery
    Fu, Dongshan
    Kuduvalli, Gopinath
    [J]. MEDICAL PHYSICS, 2008, 35 (05) : 2180 - 2194
  • [10] Elastic 3D-2D Image Registration
    Striewski, Paul
    Wirth, Benedikt
    [J]. JOURNAL OF MATHEMATICAL IMAGING AND VISION, 2022, 64 (05) : 443 - 462