A plastic scintillation dosimeter for high dose rate brachytherapy

被引:106
|
作者
Lambert, J. [1 ]
McKenzie, D. R.
Law, S.
Elsey, J.
Suchowerska, N.
机构
[1] Univ Sydney, Sch Phys, Sydney, NSW 2006, Australia
[2] Royal Prince Alfred Hosp, Dept Radiat Oncol, Camperdown, NSW 2050, Australia
[3] Univ Sydney, Opt Fibre Technol Ctr, Eveleigh, NSW 1430, Australia
[4] Bandwidth Foundry Pty Ltd, Eveleigh, NSW 1430, Australia
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2006年 / 51卷 / 21期
关键词
D O I
10.1088/0031-9155/51/21/008
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In vivo dose verification in brachytherapy requires a small insertable dosimeter with a real-time readout capability. Fibre optic scintillation dosimeters, consisting of a plastic scintillator coupled to an optical fibre, are one of the most promising dosimeters for this application. We have developed two sizes of the BrachyFOD (TM) scintillation dosimeter which have external diameters of 2.2 mm and 1 mm and have determined their important dosimetric characteristics ( depth dose relation, angular dependence, temperature dependence, energy dependence). We have shown that the background signal created by Cerenkov and fibre fluorescence does not significantly affect the performance in most clinical geometries using an Ir-192 source from an HDR brachytherapy unit. The dosimeter design enables readout at less than 0.5 s intervals. The BrachyFOD (TM) satisfies the need for a real-time in vivo brachytherapy dosimeter.
引用
收藏
页码:5505 / 5516
页数:12
相关论文
共 50 条
  • [1] A plastic scintillation dosimeter for high dose rate brachytherapy
    Lambert, J.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S32 - S32
  • [2] Optimization of a multipoint plastic scintillator dosimeter for high dose rate brachytherapy
    Rosales, Haydee M. Linares
    Duguay-Drouin, Patricia
    Archambault, Louis
    Beddar, Sam
    Beaulieu, Luc
    [J]. MEDICAL PHYSICS, 2019, 46 (05) : 2412 - 2421
  • [3] Small dimension plastic dosimeter in High Dose Rate Brachytherapy with 192Ir source
    Rego, Florbela
    Ribeiro, Tiago Ramos
    Peralta, Luis
    Jesus, Carlos
    [J]. 2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 249 - 252
  • [4] CLINICAL TRIAL OF A CUSTOMISED PLASTIC SCINTILLATION DOSIMETER FOR HDR BRACHYTHERAPY
    Lambert, J.
    Suchowerska, N.
    Jackson, M.
    Yin, Y. B.
    McKenzie, D.
    Law, S.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2008, 88 : S382 - S382
  • [5] Validation of plastic scintillation detectors for applications in low-dose-rate brachytherapy
    Therriault-Proulx, Francois
    Beaulieu, Luc
    Beddar, Sam
    [J]. BRACHYTHERAPY, 2017, 16 (04) : 903 - 909
  • [6] Optically Stimulated Luminescent Dosimeter Performance in High Dose Rate Brachytherapy
    Tien, C.
    Hiatt, J.
    Curran, B.
    Sternick, E.
    [J]. MEDICAL PHYSICS, 2012, 39 (06) : 3775 - 3775
  • [7] Feasibility study of a dosimeter for in vivo measurements of the absolute dose delivered in high dose rate brachytherapy
    Carrara, M.
    Gambarini, G.
    Borroni, M.
    Tomatis, S.
    Cerrotta, A.
    Fallai, C.
    Olmi, P.
    Zonca, G.
    [J]. NUOVO CIMENTO DELLA SOCIETA ITALIANA DI FISICA B-BASIC TOPICS IN PHYSICS, 2010, 125 (5-6): : 667 - 673
  • [8] A Novel Continuous Dosimeter for In-Vivo Absolute Dose Measurements in High Dose Rate Brachytherapy
    Carrara, M.
    Gambarini, G.
    Gambini, I.
    Tomatis, S.
    Fallai, C.
    Olmi, P.
    Zonca, G.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 75 (03): : S577 - S578
  • [9] Scintillation-Based Optical Fiber Dosimeters for High Dose Rate and Low Dose Rate Brachytherapy
    Van Erps, J.
    Gierej, Agnieszka
    Baghdasaryan, Tigran
    Martyn, Michael
    Woulfe, Peter
    Mc Laughlin, Owen
    Prise, Kevin
    Workman, Geraldine
    O'Keeffe, Sinead
    Rochlitz, Kurt
    Verlinski, Sergey
    Giaz, Agnese
    Santoro, Romualdo
    Caccia, Massimo
    Berghmans, Francis
    [J]. OPTICAL SENSING AND DETECTION VIII, 2024, 12999
  • [10] Characterization of the Components of a Scintillation Dosimeter Prototype for Brachytherapy
    Boughaba, N. H.
    Martinez-Roig, M.
    Yahlali, N.
    Diaz, J.
    Bouzid, B.
    [J]. 1ST INTERNATIONAL CONFERENCE ON RADIATIONS AND APPLICATIONS (ICRA-2017), 2018, 1994