RSC: Dosimetry in high-dose-rate brachytherapy with a radiofluorogenic gel dosimeter

被引:1
|
作者
Inoue, K. [1 ]
Watanabe, Y. [1 ,2 ]
Maeyama, T. [3 ]
Mizukami, S. [2 ]
Hayashi, S. [4 ]
Terazaki, T. [5 ]
Muraishi, H. [1 ,2 ]
Gomi, T. [1 ,2 ]
Shimono, T. [1 ,2 ]
机构
[1] Kitasato Univ, Grad Sch Med Sci, Sagamihara, Kanagawa, Japan
[2] Kitasato Univ, Sch Allied Hlth Sci, Sagamihara, Kanagawa, Japan
[3] Kitasato Univ, Sch Sci, Sagamihara, Kanagawa, Japan
[4] Hiroshima Int Univ, Fac Hlth Sci, Hiroshima, Japan
[5] Yokohama City Minato Red Cross Hosp, Yokohama, Kanagawa, Japan
关键词
RADIATION; CODE;
D O I
10.1088/1742-6596/2167/1/012032
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
A nanoclay-based radio-fluorogenic gel (NC-RFG) was used to verify the source position and dose distribution in high-dose-rate (HDR) brachytherapy. The dose response confirmed linearity up to 60 Gy. The source position could be detected with an accuracy of <= 0.3 mm, and the dose distribution near the Ir-192 source showed good agreement with the Monte Carlo simulation. NC-RFG can be expected to be a quality assurance tool suitable for the evaluating the dose distribution in HDR brachytherapy.
引用
收藏
页数:4
相关论文
共 50 条
  • [1] A Verification of High-Dose-Rate Brachytherapy Dose Distributions for Prostate Cancer with a VIPET Polymer Gel Dosimeter
    Eguchi, K.
    Watanabe, Y.
    Mizukami, S.
    Maeyama, T.
    Hayashi, S.
    Gomi, T.
    MEDICAL PHYSICS, 2018, 45 (06) : E238 - E238
  • [2] Dosimetry modeling for focal high-dose-rate prostate brachytherapy
    Mason, Josh
    Al-Qaisieh, Bashar
    Bownes, Peter
    Thwaites, David
    Henry, Ann
    BRACHYTHERAPY, 2014, 13 (06) : 611 - 617
  • [3] HIGH-DOSE-RATE BRACHYTHERAPY
    SPEISER, BL
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 27 : 116 - 116
  • [4] Clinical application of the Fricke-glucomannan gel dosimeter for high-dose-rate 192Ir brachytherapy
    Noda, S.
    Suzuki, Y.
    Hoshino, Y.
    Furukawa, S.
    Katoh, H.
    Kurotaki, K.
    Nakano, T.
    PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (14): : 3985 - 3993
  • [5] High-Dose-Rate Prostate Brachytherapy Consistently Results in High Quality Dosimetry
    White, Evan C.
    Kamrava, Mitchell R.
    Demarco, John
    Park, Sang-June
    Wang, Pin-Chieh
    Kayode, Oluwatosin
    Steinberg, Michael L.
    Demanes, D. Jeffrey
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2013, 85 (02): : 543 - 548
  • [6] Verification of dose distribution in high-dose-rate brachytherapy using a nanoclay-based radio-fluorogenic gel dosimeter
    Watanabe, Yusuke
    Maeyama, Takuya
    Mochizuki, Anri
    Mizukami, Shinya
    Hayashi, Shin-ichiro
    Terazaki, Tsuyoshi
    Muraishi, Hiroshi
    Takei, Hideyuki
    Gomi, Tsutomu
    Shimono, Tetsunori
    PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (17):
  • [7] High-dose-rate brachytherapy and pulsed brachytherapy
    Mazeron, JJ
    EUROCANCER 2002, 2002, : 57 - 58
  • [8] Urethral dosimetry and toxicity with high-dose-rate interstitial brachytherapy for vaginal cancer
    Rajagopalan, Malolan S.
    Kannan, Neeta
    Kim, Hayeon
    Houser, Christopher J.
    Beriwal, Sushil
    BRACHYTHERAPY, 2013, 12 (03) : 248 - 253
  • [9] Development of a system for on-site dosimetry audit of high-dose-rate brachytherapy
    Gwon, Hyeok-Jun
    Lee, Wonho
    Choi, Sang-Hyoun
    Cho, Gyu-Seok
    Lee, Soon Sung
    Hwang, Sun-Boong
    Kim, Kum Bae
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2025, : 909 - 917
  • [10] Urethral Dosimetry in Image Guided Tandem and Ovoid High-Dose-Rate Brachytherapy
    Hanna, J.
    Yaparpalvi, R.
    Mehta, K.
    Tome, W.
    MEDICAL PHYSICS, 2018, 45 (06) : E246 - E246