RSC: Dosimetry in high-dose-rate brachytherapy with a radiofluorogenic gel dosimeter

被引:1
|
作者
Inoue, K. [1 ]
Watanabe, Y. [1 ,2 ]
Maeyama, T. [3 ]
Mizukami, S. [2 ]
Hayashi, S. [4 ]
Terazaki, T. [5 ]
Muraishi, H. [1 ,2 ]
Gomi, T. [1 ,2 ]
Shimono, T. [1 ,2 ]
机构
[1] Kitasato Univ, Grad Sch Med Sci, Sagamihara, Kanagawa, Japan
[2] Kitasato Univ, Sch Allied Hlth Sci, Sagamihara, Kanagawa, Japan
[3] Kitasato Univ, Sch Sci, Sagamihara, Kanagawa, Japan
[4] Hiroshima Int Univ, Fac Hlth Sci, Hiroshima, Japan
[5] Yokohama City Minato Red Cross Hosp, Yokohama, Kanagawa, Japan
关键词
RADIATION; CODE;
D O I
10.1088/1742-6596/2167/1/012032
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
A nanoclay-based radio-fluorogenic gel (NC-RFG) was used to verify the source position and dose distribution in high-dose-rate (HDR) brachytherapy. The dose response confirmed linearity up to 60 Gy. The source position could be detected with an accuracy of <= 0.3 mm, and the dose distribution near the Ir-192 source showed good agreement with the Monte Carlo simulation. NC-RFG can be expected to be a quality assurance tool suitable for the evaluating the dose distribution in HDR brachytherapy.
引用
收藏
页数:4
相关论文
共 50 条
  • [31] Postoperative high-dose-rate brachytherapy in the prevention of keloids
    Veen, Ronald E.
    Kal, Henk B.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 69 (04): : 1205 - 1208
  • [32] High-dose-rate interstitial brachytherapy for gynecologic malignancies
    Beriwal, Sushil
    Bhatnagar, Ajay
    Heron, Dwight E.
    Selvaraj, Raj
    Mogus, Robert
    Kim, Hayeon
    Gerszten, Kristina
    Kelley, Joseph
    Edwards, Robert P.
    BRACHYTHERAPY, 2006, 5 (04) : 218 - 222
  • [33] High-dose-rate brachytherapy as monotherapy for prostate cancer
    Demanes, D. Jeffrey
    Ghilezan, Michel I.
    BRACHYTHERAPY, 2014, 13 (06) : 529 - 541
  • [34] High-dose-rate brachytherapy for cancer of the biliary tract
    Yokouchi, J.
    Satani, K.
    Kanesaka, N.
    Waki, H.
    Abe, K.
    Takekawa, S.
    EJC SUPPLEMENTS, 2005, 3 (02): : 208 - 208
  • [35] QUALITY ASSURANCE PROGRAM FOR HIGH-DOSE-RATE BRACHYTHERAPY
    MOWER, HW
    COSTEY, GE
    LAWSON, BJ
    WRIGHT, KA
    RADIOLOGY, 1992, 185 : 273 - 273
  • [36] High-dose-rate brachytherapy in lower eyelid cancer
    Martinez-Monge, Rafael
    Gomez-Iturriaga, Alfonso
    BRACHYTHERAPY, 2007, 6 (03) : 227 - 229
  • [37] HIGH-DOSE-RATE BRACHYTHERAPY FOR CERVICAL-CARCINOMA
    STITT, JA
    THOMADSEN, BR
    FOWLER, JF
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1992, 24 (03): : 574 - 574
  • [38] QUALITY ASSURANCE TOOL FOR HIGH-DOSE-RATE BRACHYTHERAPY
    DEWERD, LA
    JURSINIC, P
    KITCHEN, R
    THOMADSEN, BR
    MEDICAL PHYSICS, 1995, 22 (04) : 435 - 440
  • [39] Interstitial high-dose-rate brachytherapy in eyelid cancer
    Mareco, Virginia
    Bujor, Laurentiu
    Abrunhosa-Branquinho, Andre N.
    Ferreira, Miguel Reis
    Ribeiro, Tiago
    Vasconcelos, Ana Luisa
    Ferreira, Cidalina Reis
    Jorge, Marilia
    BRACHYTHERAPY, 2015, 14 (04) : 554 - 564
  • [40] FRACTIONATED HIGH-DOSE-RATE BRACHYTHERAPY FOR INTRACRANIAL GLIOMAS
    WOO, S
    BUTLER, EB
    GRANT, W
    BERNER, B
    GILDENBERG, P
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1994, 28 (01): : 247 - 249