Validation of the A&D UM-101 professional hybrid device for office blood pressure measurement according to the International Protocol

被引:30
|
作者
Stergiou, George S. [1 ]
Giovas, Periklis P. [1 ]
Gkinos, Charilaos P. [1 ]
Tzamouranis, Dimitris G. [1 ]
机构
[1] Third Univ, Sotiria Hosp, Dept Med, Hypertens Ctr, GR-11527 Athens, Greece
关键词
accuracy; European Society of Hypertension; hybrid sphygmomanometer; International Protocol; office blood pressure; professional device; validation;
D O I
10.1097/MBP.0b013e3282c9acb0
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Objective Assessment of the accuracy of the A&D UM-101 mercury-free professional device for auscultatory blood pressure (BP) measurement according to the European Society of Hypertension International Protocol. Further to auscultation, the device has a button to mark readings during deflation. Methods Fifteen adults were included in phase 1 and another 18 in phase 2. Simultaneous BP measurements were taken by two observers (connected mercury sphygmomanometers) four times, sequentially with three measurements using the tested device (two connected tested devices, one used with and the other without the mark button). Results In phase 1, the device produced 44/45/45 measurements within 5/10/15 mmHg, respectively, for systolic BP (SBP) and 39/43/45 for diastolic (DBP). In phase 2.1, 87/97/99 measurements within 5/10/15 mmHg, respectively, for SBP, and 91/97/99 for DBP (using the mark button 65/93/98 for SBP and 76/96/99 for DBP). In phase 2.2, 29 participants had at least two of their SBP differences within 5 mmHg and none had any differences within 5 mmHg, whereas 32 and none, respectively, for DBP (with mark 24/4 participants for SBP; 29/1 for DBP). Mean SBP differences were -1.5 +/- 3.5 mmHg and DBP -1.3 +/- 3.0 (with mark -3.6 +/- 4.2 and -2.8 +/- 3.7). The difference in SBP measured by the tested device with versus without using the mark button was 3.0 +/- 3.3 mmHg (P<0.001) and DBP 1.9 +/- 2.5 mmHg (P<0.001). Conclusion The device comfortably passed the validation protocol requirements. Using the mark button, the device, however, failed to meet the validation criteria. Therefore, it is recommended for clinical use without using the mark button.
引用
收藏
页码:37 / 42
页数:6
相关论文
共 50 条
  • [41] Validation of the visocor HM40 wrist blood pressure measuring device according to the International Protocol
    Dorigatti, Francesca
    Bonso, Elisa
    Saladini, Francesca
    Palatini, Paolo
    BLOOD PRESSURE MONITORING, 2009, 14 (02) : 83 - 86
  • [42] Validation of the A&D UA-1020 upper-arm blood pressure monitor for home blood pressure monitoring according to the British Hypertension Society Protocol
    Zeng, Wei-Fang
    Kang, Yuan-Yuan
    Liu, Ming
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2013, 18 (03) : 177 - 181
  • [43] Validation of the A&D UM-212BLE monitor according to ISO 81060-2, 2018: a device with clinically important programmability
    Alpert, Bruce S.
    BLOOD PRESSURE MONITORING, 2023, 28 (02) : 113 - 115
  • [44] Validation of the Kingyield BP101H blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol
    Zeng, Wei-Fang
    Huang, Qi-Fang
    Sheng, Chang-Sheng
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2010, 15 (06) : 329 - 331
  • [45] Validation of the professional device for blood pressure measurement Microlife WatchBP Office in adults and children according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization standard
    Kollias, Anastasios
    Ntineri, Angeliki
    Kyriakoulis, Konstantinos G.
    Stambolliu, Emelina
    Lagou, Styliani
    Boubouchairopoulou, Nadia
    Stergiou, George S.
    BLOOD PRESSURE MONITORING, 2018, 23 (02) : 112 - 114
  • [46] Home blood pressure measurement: validation of the Braun BP 2550 (UG) monitor according to the ESH International Protocol
    Nolly, H
    Romero, M
    Nolly, A
    Osso, P
    Reinoso, O
    Nolly, M
    BLOOD PRESSURE MONITORING, 2004, 9 (01) : 53 - 58
  • [47] Validation of the G-care electronics SP-8000 monitor for blood pressure measurement according to the international validation protocol
    Topouchian, J.
    El Feghali, R.
    El Assaad, H.
    Pannier, B.
    Asmar, R.
    JOURNAL OF HYPERTENSION, 2007, 25 : S172 - S172
  • [48] Validation of A&D TM-2430 upper-arm blood pressure monitor for ambulatory blood pressure monitoring in children and adolescents, according to the British Hypertension Society protocol
    Yip, Gabriel Wai-Kwok
    So, Hung-Kwan
    Li, Albert Martin
    Tomlinson, Brian
    Wong, Sik-Nin
    Sung, Rita Yn-Tz
    BLOOD PRESSURE MONITORING, 2012, 17 (02) : 76 - 79
  • [49] Validation of the omron M6 upper arm blood pressure measuring device according to the international protocol in adults
    Altunkan, Sekip
    Iliman, Nevzat
    Kayaturk, Nur
    Altunkan, Erkan
    JOURNAL OF HYPERTENSION, 2006, 24 : 213 - 213
  • [50] Validation of the Thermor BIOS BD215 device for home blood pressure measurement according to the European Society of Hypertension International Protocol revision 2010
    Benetti, Elisabetta
    Fania, Claudio
    Hernandez, Veronica Marquez
    Palatini, Paolo
    BLOOD PRESSURE MONITORING, 2014, 19 (03) : 176 - 179