Validation of a Monte Carlo simulation of the Inveon PET scanner using GATE

被引:8
|
作者
Lu, Lijun
Zhang, Houjin
Bian, Zhaoying
Ma, Jianhua [1 ]
Feng, Qiangjin
Chen, Wufan [1 ]
机构
[1] Southern Med Univ, Sch Biomed Engn, Guangzhou 510515, Guangdong, Peoples R China
基金
中国国家自然科学基金;
关键词
PET; PET simulation; Inveon scanner; GATE; SPECT; PLATFORM; MODEL;
D O I
10.1016/j.nima.2016.04.059
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The purpose of this study is to validate the application of GATE (Geant4 Application for Tomographic Emission) Monte Carlo simulation toolkit in order to model the performance characteristics of Siemens Inveon small animal PET system. The simulation results were validated against experimental/published data in accordance with the NEMA NU-4 2008 protocol for standardized evaluation of spatial resolution, sensitivity, scatter fraction (SF) and noise equivalent counting rate (NECR) of a preclinical PET system. An agreement of less than 18% was obtained between the radial, tangential and axial spatial resolutions of the simulated and experimental results. The simulated peak NECR of mouse-size phantom agreed with the experimental result, while for the rat-size phantom simulated value was higher than experimental result. The simulated and experimental SFs of mouse- and rat-size phantom both reached an agreement of less than 2%. It has been shown the feasibility of our GATE model to accurately simulate, within certain limits, all major performance characteristics of Inveon PET system. (C) 2016 Published by Elsevier B.V.
引用
收藏
页码:170 / 175
页数:6
相关论文
共 50 条
  • [41] Validation of GATE Monte Carlo simulations of the GE Advance/Discovery LS PET scanners
    Schmidtlein, CR
    Kirov, AS
    Nehmeh, SA
    Erdi, YE
    Humm, JL
    Amols, HI
    Bidaut, LM
    Ganin, A
    Stearns, CW
    McDaniel, DL
    Hamacher, KA
    [J]. MEDICAL PHYSICS, 2006, 33 (01) : 198 - 208
  • [42] Investigation of time-of-flight benefits in an LYSO-based PET/CT scanner: A Monte Carlo study using GATE
    Geramifar, P.
    Ay, M. R.
    Zafarghandi, M. Shamsaie
    Sarkar, S.
    Loudos, G.
    Rahmim, A.
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2011, 641 (01): : 121 - 127
  • [43] Monte Carlo Simulation Study of Several Camera Designs for the PET Component of a Dedicated Breast PET/CT Scanner
    Bowen, S. L.
    Yang, Y.
    Wu, Y.
    Cherry, S. R.
    Boone, J. M.
    Moses, W. W.
    Badawi, R. D.
    [J]. 2006 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOL 1-6, 2006, : 3920 - 3924
  • [44] Validation of a GATE simulation model for the siemens PET/CT biograph™ 6 scanner, using NEMA 2001 standards
    Gonias, P.
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2006, 33 : S137 - S137
  • [45] Monte Carlo simulation study of several camera designs for the PET component of a dedicated breast PET/CT scanner
    Bowen, S. L.
    Cherry, S. R.
    Boone, J. M.
    Badawi, R. D.
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2006, 33 : S117 - S117
  • [46] Validation and evaluation of a GATE model for MAMMI PET scanner
    Emami, Azadeh
    Ghafarian, Pardis
    Ghadiri, Hossein
    Geramifar, Parham
    Ay, Mohammad Reza
    [J]. IRANIAN JOURNAL OF NUCLEAR MEDICINE, 2020, 28 (01): : 33 - 38
  • [47] PeneloPET, a Monte Carlo PET simulation toolkit based on PENELOPE: Features and Validation
    Espana, S.
    Herraiz, J. L.
    Vicente, E.
    Vaquero, J. J.
    Desco, M.
    Udas, J. M.
    [J]. 2006 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOL 1-6, 2006, : 2597 - 2601
  • [48] PeneloPET, a Monte Carlo PET simulation tool based on PENELOPE: features and validation
    Espana, S.
    Herraiz, J. L.
    Vicente, E.
    Vaquero, J. J.
    Desco, M.
    Udias, J. M.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (06): : 1723 - 1742
  • [49] Monte Carlo simulation of two Siemens Biograph PET/CT system using GATE: Image quality performance
    Saaidi, Rahal
    Zeghari, Abdelkrim
    El Moursli, Rajaa Cherkaoui
    [J]. RADIATION PHYSICS AND CHEMISTRY, 2024, 218
  • [50] Monte Carlo simulation of Ray-Scan 64 PET system and performance evaluation using GATE toolkit
    Li, Suying
    Zhang, Qiushi
    Vuletic, Ivan
    Xie, Zhaoheng
    Yang, Kun
    Ren, Qiushi
    [J]. JOURNAL OF INSTRUMENTATION, 2017, 12