Error detection during VMAT delivery using EPID-based 3D transit dosimetry

被引:39
|
作者
Mijnheer, B. [1 ]
Jomehzadeh, A. [1 ,3 ]
Gonzalez, P. [1 ]
Olaciregui-Ruiz, I [1 ]
Rozendaal, R. [1 ]
Shokrani, P. [2 ]
Spreeuw, H. [1 ]
Tielenburg, R. [1 ]
Mans, A. [1 ]
机构
[1] Netherlands Canc Inst, Dept Radiat Oncol, Plesmanlaan 121, NL-1066 CX Amsterdam, Netherlands
[2] Isfahan Univ Med Sci, Dept Med Phys, Esfahan, Iran
[3] Kerman Univ Med Sci, Dept Med Phys, Kerman, Iran
关键词
EPID dosimetry; VMAT; Error detection; Quality assurance; MODULATED ARC THERAPY; VIVO PORTAL DOSIMETRY; CLINICAL-EXPERIENCE; RADIATION-THERAPY; QUALITY-ASSURANCE; VERIFICATION; ALGORITHM; RAPIDARC; SYSTEMS; IMRT;
D O I
10.1016/j.ejmp.2018.10.005
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To investigate the effectiveness of an EPID-based 3D transit dosimetry system in detecting deliberately introduced errors during VMAT delivery. Methods: An Alderson phantom was irradiated using four VMAT treatment plans (one prostate, two head-and-neck and one lung case) in which delivery, thickness and setup errors were introduced. EPID measurements were performed to reconstruct 3D dose distributions of "error" plans, which were compared with "no-error" plans using the mean gamma (gamma(mean)), near-maximum gamma (gamma 1%) and the difference in isocenter dose (Delta D-isoc) as metrics. Results: Out of a total of 42 serious errors, the number of errors detected was 33 (79%), and 27 out of 30 (90%) if setup errors are not included. The system was able to pick up errors of 5 mm movement of a leaf bank, a wrong collimator rotation angle and a wrong photon beam energy. A change in phantom thickness of 1 cm was detected for all cases, while only for the head-and-neck plans a 2 cm horizontal and vertical shift of the phantom were alerted. A single leaf error of 5 mm could be detected for the lung plan only. Conclusion: Although performed for a limited number of cases and error types, this study shows that EPID-based 3D transit dosimetry is able to detect a number of serious errors in dose delivery, leaf bank position and patient thickness during VMAT delivery. Errors in patient setup and single leaf position can only be detected in specific cases.
引用
收藏
页码:137 / 145
页数:9
相关论文
共 50 条
  • [1] 3D EPID-based in vivo dosimetry for IMRT and VMAT
    Mijnheer, B.
    Olaciregui-Ruiz, I.
    Rozendaal, R.
    Sonke, J-J
    Spreeuw, H.
    Tielenburg, R.
    van Herk, M.
    Vijlbrief, R.
    Mans, A.
    [J]. 7TH INTERNATIONAL CONFERENCE ON 3D RADIATION DOSIMETRY (IC3DDOSE), 2013, 444
  • [2] A novel approach to EPID-based 3D volumetric dosimetry for IMRT and VMAT QA
    Alhazmi, Abdulaziz
    Gianoli, Chiara
    Neppl, Sebastian
    Martins, Juliana
    Veloza, Stella
    Podesta, Mark
    Verhaegen, Frank
    Reiner, Michael
    Belka, Claus
    Parodi, Katia
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (11):
  • [3] An EPID-based 3D patient dose verification method for SBRT-VMAT delivery
    McCowan, P.
    vanUytven, E.
    vanBeek, T.
    McCurdy, B.
    [J]. MEDICAL PHYSICS, 2014, 41 (08) : 2 - 2
  • [4] Sensing ability of EPID-based in vivo dosimetry for VMAT
    Tanooka, M.
    Tarutani, K.
    Doi, H.
    Suzuki, H.
    Takada, Y.
    Fujiwara, M.
    Toda, Y.
    Fujimoto, H.
    Miyashita, M.
    Okumura, A.
    Kagawa, K.
    Kamikonya, N.
    Yamakado, K.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S939 - S940
  • [5] On the Sensitivity of An EPID-Based 3D Dose Verification System to Detect Delivery Errors in VMAT Treatments
    Gonzalez, P.
    Olaciregui-Ruiz, I.
    Mijnheer, B.
    Mans, A.
    Rozendaal, R.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3633 - 3633
  • [6] EPID-based 3D dosimetry for pre-treatment FFF VMAT stereotactic body radiotherapy plan verification using dosimetry Check™
    Chendi, Agnese
    Botti, Andrea
    Orlandi, Matteo
    Sghedoni, Roberto
    Iori, Mauro
    Cagni, Elisabetta
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2021, 81 : 227 - 236
  • [7] An assessment of a 3D EPID-based dosimetry system using conventional two- and three-dimensional detectors for VMAT
    Stevens, S.
    Dvorak, P.
    Spevacek, V.
    Pilarova, K.
    Bray-Parry, M.
    Gesner, J.
    Richmond, A.
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2018, 45 : 25 - 34
  • [8] Sensitivity analysis of EPID-based 3D dose reconstruction for VMAT QA
    Alhazmi, A.
    Gianoli, C.
    Neppl, S.
    Martins, J.
    Veloza, S.
    Podesta, M.
    Verhaegen, F.
    Reiner, M.
    Belka, C.
    Parodi, K.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S960 - S961
  • [9] EPID-based 3D in vivo dosimetry in head and neck radiotherapy
    Biltekin, F.
    Yazici, G.
    Ozyigit, G.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2021, 161 : S1354 - S1355
  • [10] Real-Time EPID-Based Detection Metric For VMAT Delivery Errors
    Passarge, M.
    Fix, M. K.
    Manser, P.
    Stampanoni, M. F. M.
    Siebers, J. V.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3713 - 3713