Error detection during VMAT delivery using EPID-based 3D transit dosimetry

被引:39
|
作者
Mijnheer, B. [1 ]
Jomehzadeh, A. [1 ,3 ]
Gonzalez, P. [1 ]
Olaciregui-Ruiz, I [1 ]
Rozendaal, R. [1 ]
Shokrani, P. [2 ]
Spreeuw, H. [1 ]
Tielenburg, R. [1 ]
Mans, A. [1 ]
机构
[1] Netherlands Canc Inst, Dept Radiat Oncol, Plesmanlaan 121, NL-1066 CX Amsterdam, Netherlands
[2] Isfahan Univ Med Sci, Dept Med Phys, Esfahan, Iran
[3] Kerman Univ Med Sci, Dept Med Phys, Kerman, Iran
关键词
EPID dosimetry; VMAT; Error detection; Quality assurance; MODULATED ARC THERAPY; VIVO PORTAL DOSIMETRY; CLINICAL-EXPERIENCE; RADIATION-THERAPY; QUALITY-ASSURANCE; VERIFICATION; ALGORITHM; RAPIDARC; SYSTEMS; IMRT;
D O I
10.1016/j.ejmp.2018.10.005
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To investigate the effectiveness of an EPID-based 3D transit dosimetry system in detecting deliberately introduced errors during VMAT delivery. Methods: An Alderson phantom was irradiated using four VMAT treatment plans (one prostate, two head-and-neck and one lung case) in which delivery, thickness and setup errors were introduced. EPID measurements were performed to reconstruct 3D dose distributions of "error" plans, which were compared with "no-error" plans using the mean gamma (gamma(mean)), near-maximum gamma (gamma 1%) and the difference in isocenter dose (Delta D-isoc) as metrics. Results: Out of a total of 42 serious errors, the number of errors detected was 33 (79%), and 27 out of 30 (90%) if setup errors are not included. The system was able to pick up errors of 5 mm movement of a leaf bank, a wrong collimator rotation angle and a wrong photon beam energy. A change in phantom thickness of 1 cm was detected for all cases, while only for the head-and-neck plans a 2 cm horizontal and vertical shift of the phantom were alerted. A single leaf error of 5 mm could be detected for the lung plan only. Conclusion: Although performed for a limited number of cases and error types, this study shows that EPID-based 3D transit dosimetry is able to detect a number of serious errors in dose delivery, leaf bank position and patient thickness during VMAT delivery. Errors in patient setup and single leaf position can only be detected in specific cases.
引用
收藏
页码:137 / 145
页数:9
相关论文
共 50 条
  • [21] Characterization of a new EPID-based system for in-vivo dosimetry in VMAT treatments
    Bresciani, S.
    Poli, M.
    Miranti, A.
    Maggio, A.
    Di Dia, A.
    Bracco, C.
    Stasi, M.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S736 - S737
  • [22] Clinical pilot study for EPID-based in vivo dosimetry using EPIgray™ for head and neck VMAT
    Halliday, Sophie D.
    Day, Rebecca A.
    Greig, Lynne
    Louwe, Robert J. W.
    [J]. PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2022, 45 (04) : 1335 - 1340
  • [23] Clinical pilot study for EPID-based in vivo dosimetry using EPIgray™ for head and neck VMAT
    Sophie D. Halliday
    Rebecca A. Day
    Lynne Greig
    Robert J. W. Louwe
    [J]. Physical and Engineering Sciences in Medicine, 2022, 45 : 1335 - 1340
  • [24] Hot-spot analysis for incident detection in real-time 3D EPID-based in vivo dosimetry
    Rozendaal, R.
    Spreeuw, H.
    Mijnheer, B. J.
    Mans, A.
    Gonzalez, P.
    Olaciregui-Ruiz, I.
    Van Herk, M.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S427 - S428
  • [25] Implementation of An Efficient Workflow for the Analysis of Alerts Observed During Large Scale EPID-Based 3D in Vivo Dosimetry
    Mijnheer, B.
    van Mourik, A.
    Ruiz, I. Olaciregui
    Mans, A.
    [J]. MEDICAL PHYSICS, 2017, 44 (06) : 2989 - 2989
  • [26] A remote EPID-based dosimetric auditing method for VMAT delivery using a digital phantom concept
    Greer, P.
    Legge, K.
    Miri, N.
    Vial, P.
    Fuangrod, T.
    Lehmann, J.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2017, 123 : S285 - S286
  • [27] Clinical results of an EPID-based in-vivo dosimetry for prostate cancer treated by VMAT
    Falco, M. D.
    Giancaterino, S.
    De Nicola, A.
    Perrotti, F.
    Menna, S.
    Fidanzio, A.
    Piermattei, A.
    Genovesi, D.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S706 - S706
  • [28] Investigating the effectiveness of monitoring relevant variations during IMRT and VMAT treatments by EPID-based 3D in vivo verification performed using planning CTs
    Li, Yinghui
    Zhu, Jinhan
    Shi, Jinping
    Chen, Lixin
    Liu, Xiaowei
    [J]. PLOS ONE, 2019, 14 (06):
  • [29] 3D EPID based dosimetry for pre-treatment verification of VMAT - methods and challenges
    Greer, P. B.
    [J]. 7TH INTERNATIONAL CONFERENCE ON 3D RADIATION DOSIMETRY (IC3DDOSE), 2013, 444
  • [30] Initial clinical experience with EPID-based in-vivo dosimetry for VMAT treatment verification
    Cilla, S.
    Meluccio, D.
    Fidanzio, A.
    Azario, L.
    Greco, F.
    Deodato, F.
    Macchia, G.
    Digesu, C.
    Valentini, V.
    Morganti, A. G.
    Piermattei, A.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S852 - S852