Optimal Path Planning for Robotic Insertion of Steerable Electrode Arrays in Cochlear Implant Surgery

被引:49
|
作者
Zhang, Jian [1 ]
Roland, J. Thomas, Jr. [2 ]
Manolidis, Spiros [3 ]
Simaan, Nabil [1 ]
机构
[1] Columbia Univ, Dept Mech Engn, New York, NY 10027 USA
[2] NYU, Med Ctr, Dept Otolaryngol & Neurosurg, New York, NY 10016 USA
[3] Beth Israel Hosp, Dept Head & Neck Surg, New York, NY 10003 USA
基金
美国国家科学基金会;
关键词
surgical assistance; cochlear implant path planning; underactuated robot; steerable electrode array; STRAIGHT; DESIGN;
D O I
10.1115/1.3039513
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper presents an optimal path planning method of steerable electrode arrays for robot assisted cochlear implant surgery In this paper, the authors present a novel design of steerable electrode arrays that can actively bend at the up An embedded strand in the electrode array provides an active steering degrees of freedom (DoF) This paper addresses the calibration of the steerable electrode arrays and the optimal path planning for inserting it into planar and three dimensional scala tympani models The goal of the path planning is to minimize the intracochlear forces that the electrode array applies on the walls of the scala tympam during insertion This problem is solved by designing insertion path planning algorithms that provide best fit between the shape of the electrode aunty and the curved scala tympani during insertion Optimality measures that account fittape discrepancies between the steerable electrode array and the scala tympam are used to solve for the optimal path planning of the robot Different arrangements of DoF and insertion speed force feedback (ISFF) are simulated and experimentally validated in this paper A quality of insertion metric describing the gap between the steerable electrode array and the scala tympam model is presented and as correspondence to the insertion force is shown The results of using 1DoF 2DoF and 4DoF electrode array insertion setups are compared The 1DoE insertion setup fiver nonsteerable electrode arrays The 2DoF insertion setup uses single axis insertion with steerable electrode arrays. The 4DoF insertion setup allows full control of the insertion depth and the approach angle of the electrode with respect to the cochlea while using steerable electrode arrays It is shown that living steerable electrode arrays significantly reducer the maximal insertion force (59.6% or more) and effectively prevents buckling of the electrode array The 4DoF Insertion setup further reduces the maximal electrode insertion forces The results of using ISFF for steerable electrodes show a slight decrease in the insertion forcer in contrast to a slight increase for nonsteerable electrodes These results show that further research is required in order to determine the optimal ISFF control law and its effectiveness in reducing electrode invention fortes [DOI 10.1115/1.3039513]
引用
收藏
页数:10
相关论文
共 50 条
  • [1] Path Planning and Workspace Determination for Robot-Assisted Insertion of Steerable Electrode Arrays for Cochlear Implant Surgery
    Zhang, Jian
    Wei, Wei
    Manolidis, Spiros
    Roland, J. Thomas, Jr.
    Simaan, Nabil
    MEDICAL IMAGE COMPUTING AND COMPUTER-ASSISTED INTERVENTION - MICCAI 2008, PT II, PROCEEDINGS, 2008, 5242 : 692 - 700
  • [2] Optimal path generation in scala tympani and path planning for robotic cochlear implant of perimodiolar electrode
    Wang, Zuo
    Li, Jianjun
    Wu, Yongzhen
    Zhu, Ruifeng
    Wang, Binrui
    Zhao, Kai
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2020, 234 (06) : 578 - 589
  • [3] Inroads Toward Robot-Assisted Cochlear Implant Surgery Using Steerable Electrode Arrays
    Zhang, Jian
    Wei, Wei
    Ding, Jienan
    Roland, J. Thomas, Jr.
    Manolidis, Spiros
    Simaan, Nabil
    OTOLOGY & NEUROTOLOGY, 2010, 31 (08) : 1199 - 1206
  • [4] A pilot study of robot-assisted cochlear implant surgery using steerable electrode arrays
    Zhang, Jian
    Xu, Kai
    Simaan, Nabil
    Manolidis, Spiros
    MEDICAL IMAGE COMPUTING AND COMPUTER-ASSISTED INTERVENTION - MICCAI 2006, PT 1, 2006, 4190 : 33 - 40
  • [5] Model and Parameter Identification of Friction during Robotic Insertion of Cochlear-Implant Electrode Arrays
    Zhang, Jian
    Bhattacharyya, Samrat
    Simaan, Nabil
    ICRA: 2009 IEEE INTERNATIONAL CONFERENCE ON ROBOTICS AND AUTOMATION, VOLS 1-7, 2009, : 3528 - 3533
  • [6] Preservation of Residual Hearing After Cochlear Implant Surgery With Deep Insertion Electrode Arrays
    Sierra, Cristina
    Calderon, Manuela
    Barcena, Eduardo
    Tisaire, Alexandra
    Raboso, Eduardo
    OTOLOGY & NEUROTOLOGY, 2019, 40 (04) : E373 - E380
  • [7] Devellopment of a steerable cochlear implant electrode array
    Chen, B. K.
    Kha, H. N.
    Clark, G. M.
    3RD KUALA LUMPUR INTERNATIONAL CONFERENCE ON BIOMEDICAL ENGINEERING 2006, 2007, 15 : 607 - +
  • [8] In vivo measures of cochlear length and insertion depth of nucleus cochlear implant electrode arrays
    Ketten, DR
    Vannier, MW
    Skinner, MW
    Gates, GA
    Wang, G
    Neely, JG
    ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 1998, 107 (11): : 1 - 16
  • [9] Magnetically Steered Robotic Insertion of Cochlear-Implant Electrode Arrays: System Integration and First-In-Cadaver Results
    Bruns, Trevor L.
    Riojas, Katherine E.
    Ropella, Dominick S.
    Cavilla, Matt S.
    Petruska, Andrew J.
    Freeman, Michael H.
    Labadie, Robert F.
    Abbott, Jake J.
    Webster, Robert J., III
    IEEE ROBOTICS AND AUTOMATION LETTERS, 2020, 5 (02) : 2240 - 2247
  • [10] Preoperative prediction of insertion depth of lateral wall cochlear implant electrode arrays
    Khan, Mohammad M. R.
    Labadie, Robert F.
    Noble, Jack H.
    MEDICAL IMAGING 2020: IMAGE-GUIDED PROCEDURES, ROBOTIC INTERVENTIONS, AND MODELING, 2021, 11315