Optimal path generation in scala tympani and path planning for robotic cochlear implant of perimodiolar electrode

被引:5
|
作者
Wang, Zuo [1 ]
Li, Jianjun [1 ]
Wu, Yongzhen [2 ]
Zhu, Ruifeng [1 ]
Wang, Binrui [1 ]
Zhao, Kai [1 ]
机构
[1] China Jiliang Univ, 258 Xueyuan St,Xiasha Higher Educ Pk, Hangzhou 310018, Zhejiang, Peoples R China
[2] Fudan Univ, EYE & ENT Hosp, Shanghai, Peoples R China
基金
中国国家自然科学基金;
关键词
Robotic cochlear implant; optimal path; path planning; perimodiolar electrode; scala tympani; B-SPLINE CURVE; INTERPOLATION; LENGTH; TOMOGRAPHY; INSERTION; SURFACE;
D O I
10.1177/0954411920908969
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In this study, a new idea of the optimal path generation method was proposed and a path planning strategy for robotic cochlear implant of perimodiolar electrode was designed. The centerline of scala tympani channel was taken as the optimal implant path of the perimodiolar electrode, which aimed to reduce the damage of the electrode to the cochlea during implantation. First, the three-dimensional cochlear model was reconstructed based on the micro-computed tomography images of cochlea, and it was re-segmented to obtain the cross sections of the scala tympani at different angles. Then, the image processing method was used to determine the central point of the scala tympani cross sections. The cubic B-spline interpolation method was used to fit these discrete central points to generate the optimal path. Finally, the coordinate information of the optimal path was combined with the stylet extraction state of perimodiolar electrode to conduct the path planning for robotic cochlear implant, and the result was sent to the robot for kinematic inverse solution to obtain the robot motion trajectory. The robotic cochlear implant experiment was performed with the model of scala tympani. The results showed that the maximum implant force based on path planning was 0.084 N, and the maximum implant force without path planning was 0.134 N. The optimal path generation and the path planning method effectively help to reduce the damage of the electrode to the cochlea.
引用
收藏
页码:578 / 589
页数:12
相关论文
共 50 条
  • [1] Optimal Path Planning for Robotic Insertion of Steerable Electrode Arrays in Cochlear Implant Surgery
    Zhang, Jian
    Roland, J. Thomas, Jr.
    Manolidis, Spiros
    Simaan, Nabil
    JOURNAL OF MEDICAL DEVICES-TRANSACTIONS OF THE ASME, 2009, 3 (01):
  • [2] Factors Affecting Outcomes in Cochlear Implant Recipients Implanted With a Perimodiolar Electrode Array Located in Scala Tympani
    Holden, Laura K.
    Firszt, Jill B.
    Reeder, Ruth M.
    Uchanski, Rosalie M.
    Dwyer, Noel Y.
    Holden, Timothy A.
    OTOLOGY & NEUROTOLOGY, 2016, 37 (10) : 1662 - 1668
  • [3] Cochlear Size Assessment Predicts Scala Tympani Volume and Electrode Insertion Force- Implications in Robotic Assisted Cochlear Implant Surgery
    Dhanasingh, Anandhan
    Swords, Chloe
    Bance, Manohar
    Van Rompaey, Vincent
    Van de Heyning, Paul
    FRONTIERS IN SURGERY, 2021, 8
  • [4] THE NUCLEUS MULTICHANNEL COCHLEAR IMPLANT - COMPARISON OF SCALA TYMPANI VS SCALA VESTIBULI ELECTRODE PLACEMENT IN A SINGLE PATIENT
    PIJL, S
    NOEL, F
    OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 1992, 107 (03) : 472 - 474
  • [5] Potential distribution in the scala tympani youth cochlear implant stimulation
    Kral, A
    Mortazavi, D
    Hartmann, R
    Klinke, R
    PFLUGERS ARCHIV-EUROPEAN JOURNAL OF PHYSIOLOGY, 1997, 433 (06): : P174 - P174
  • [6] The consequences of neural degeneration regarding optimal cochlear implant position in scala tympani: A model approach
    Briaire, Jeroen J.
    Frijns, Johan H. M.
    HEARING RESEARCH, 2006, 214 (1-2) : 17 - 27
  • [7] Electrode Migration in Patients with Perimodiolar Cochlear Implant Electrodes
    Mittmann, Philipp
    Rademacher, Grit
    Mutze, Sven
    Ernst, Arneborg
    Todt, Ingo
    AUDIOLOGY AND NEURO-OTOLOGY, 2015, 20 (06) : 349 - 353
  • [8] In vitro modifications of the scala tympani environment and the cochlear implant array surface
    Kontorinis, Georgios
    Scheper, Verena
    Wissel, Kirsten
    Stoever, Timo
    Lenarz, Thomas
    Paasche, Gerrit
    LARYNGOSCOPE, 2012, 122 (09): : 2057 - 2063
  • [9] Cochlear Size Assessment Predicts Scala Tympani Volume and Electrode Insertion Force- Implications in Robotic Assisted Cochlear Implant Surgery (vol 8, 723897, 2021)
    Dhanasingh, Anandhan
    Swords, Chloe
    Bance, Manohar
    Van Rompaey, Vincent
    van de Heyning, Paul
    FRONTIERS IN SURGERY, 2021, 8
  • [10] Morphometric Analysis and Linear Measurements of the Scala Tympani and Implications in Cochlear Implant Electrodes
    Fujiwara, Rance J. T.
    Ishiyama, Gail
    Lopez, Ivan A.
    Ishiyama, Akira
    OTOLOGY & NEUROTOLOGY, 2023, 44 (05) : E343 - E349